Agilent 6060B: troubleshooting and error codes
The ServiceForge knowledge base holds 41 documented symptoms for the Agilent 6060B, with 60 ranked causes, 110 diagnostic checks with test points and expected values, and 11 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Blank display during power-on selftest
Front panel display remains blank; no error message appears. May indicate RAM, ROM, internal timer failure, or microprocessor lockup.
Ranked causes
- 1. Primary microprocessor U203 RAM failure (occasional)
- 2. Primary microprocessor U203 ROM failure (occasional)
- 3. Primary microprocessor U203 internal timer failure (occasional)
- 4. Primary microprocessor U203 lockup (rare)
Measure SA_GATE line square wave frequency
- Test point
- TP201-8
- Expected
- 10Hz (RAM failure), 100Hz (ROM failure), or 1KHz (timer failure)
- Tools
- Oscilloscope
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit fault caution
Overpower circuit not responding correctly to input voltage/current; power limitation not activating or deactivating properly
Ranked causes
- 1. U7 comparator failure (occasional)
- 2. R123 summing resistor pack open or incorrect value (occasional)
- 3. RT551 temperature monitor or U327 failure (occasional)
- 4. +12V reference voltage low or absent (occasional)
Verify +12V reference voltage present
- Test point
- +12V reference input
- Expected
- +12V ±tolerance
- Tools
- DMM
CC/CV control circuit malfunction high voltage
Either CC or CV loop fails to control input power stage conduction correctly; detected during S.A. mode testing
Ranked causes
- 1. Incorrect CC_EN or CV_EN signal logic level (common)
- 2. Incorrect waveforms at test points (common)
- 3. Input power stage, current limit, or power limit circuit interference (occasional)
Enable S.A. mode and verify waveforms match Figure 3-5
- Test point
- TP201 and TP301
- Expected
- Waveforms match Figure 3-5; applicable to CC or CV mode selected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 105-108 at power-on despite successful self-test analysis
Unit displays error 105-108 during startup even after S.A. analysis shows no faults
Ranked causes
- 1. IMON adjustment out of specification (occasional)
- 2. U309 switch malfunction (occasional)
- 3. DAC or amplifier fault generating incorrect waveform (rare)
Verify U309 switch closure at low level test points
- Test point
- U309 switches (SDB0-7 data inputs and SLW signal outputs)
- Expected
- Switch closes when test point is at Low level; check IMON, -10V, and VMON test points per Table 3-3
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overall troubleshooting guidance high voltage
General introduction to troubleshooting procedures for Agilent 6060B/6063B Electronic Loads. Describes overall approach to isolate problems to functional areas, then to specific components. Mentions verification tests in Chapter 2.
Ranked causes
- 1. Problem not with load but associated circuit (common)
- 2. Component found defective (unknown)
- 3. Serial EEPROM chip U211 replaced (rare)
Run front panel verification tests
- Test point
- Load functions per Chapter 2 procedures
- Expected
- Load passes all verification tests
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DAC circuit malfunction or incorrect SLEW signal
SLEW signal generated by DAC circuits is not producing correct waveforms, affecting input power control circuit operation
Ranked causes
- 1. Incorrect SD0-7 data bus lines to readback DAC U322 (unknown)
- 2. Incorrect SDB0-7 data lines to main or transient DACs (unknown)
Enable S.A. mode by connecting jumpers
- Test point
- Test headers TP201 and TP301
- Expected
- Jumpers connected in S.A. mode positions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -102 displayed: Secondary microprocessor ROM failure or missing thermistor
Secondary microprocessor U301 internal ROM test failed, or thermistor RT551 is missing or open circuit.
Ranked causes
- 1. Secondary microprocessor U301 ROM failure (occasional)
- 2. Thermistor RT551 missing or open circuit (occasional)
Inspect thermistor RT551
- Test point
- RT551 location on main circuit board
- Expected
- Component present and electrically continuous
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect slew rate waveforms or poor slew rate response
Slew circuit waveforms incorrect when slew rates programmed from front panel; slew rate does not respond as specified
Ranked causes
- 1. Analog switch U317 malfunction in slew circuit (common)
- 2. CC/CV control circuit fault (rare)
Program slew rate #1 and verify waveform
- Test point
- Slew circuit output (Figure 3-4 bottom waveform)
- Expected
- Waveform matches slew rate 0.001 A/µs in Figure 3-4
- Tools
- Oscilloscope with delayed sweep capability
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect slew waveforms in S.A. mode
Slew circuit test waveforms incorrect or absent in self-analysis mode
Ranked causes
- 1. Incorrect SDB0-7 data inputs or SLW signal outputs from U305 (common)
- 2. Operational amplifier or switch fault (occasional)
Verify SDB0-7 data inputs and SLW signal outputs from U305
- Test point
- U305 pins (SDB0-7 inputs, SLW1-SLW4 outputs)
- Expected
- Signatures match S.A. Table 3-13 waveform specification
- Tools
- Oscilloscope, S.A. mode enabled
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input power stage malfunction high voltage
One or more of eight parallel input power stages not functioning; detected via signature analysis in S.A. mode or abnormal power output
Ranked causes
- 1. Defect in specific input power stage (one of eight stages) (common)
- 2. Failure in all input power stages simultaneously (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit active - power limiting caution
Load operating in overpower condition; power sinking capability limited by thermal or time duration threshold
Ranked causes
- 1. Overpower condition exists - EPU bit false (common)
- 2. Overpower condition exists - EPU bit true (common)
Check test points and reference voltage using Table 3-3 conditions
- Test point
- U12 amplifier, U7 comparators, R123 summing resistor pack, +12V reference
- Expected
- Values per Table 3-3
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage (OV) condition detected high voltage
OV circuit activates to limit input voltage; overrides CC and CV control loops
Ranked causes
- 1. Input voltage exceeds rated voltage; OV circuit triggered correctly (common)
- 2. OV circuit latch failure; R64 or D18 defective (occasional)
Measure inverting amplifier U12-7 output voltage as input voltage varies
- Test point
- U12-7 output
- Expected
- 0 to -10V as input varies from zero to rated input voltage
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current Monitor (IMON) circuit
IMON circuit adjustment required after input power stage component replacement
Ranked causes
- 1. Input power stage component replaced (common)
Verify IMON reading at adjustment test points
- Test point
- IMON adjustment test points (Figure 6-2)
- Expected
- 0 ±0.5 millivolts mV, ±0.5 mV
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EEPROM initialization
EEPROM chip U211 must be reinitialized with GP-IB address, model number and calibration constants when U211 or main PC board is replaced
Ranked causes
- 1. EEPROM chip U211 or main PC board replaced (occasional)
Execute command CAL 1
- Test point
- Load interface
- Expected
- Turn calibration mode on
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EEPROM initialization after chip U211 replacement
Serial EEPROM chip U211 must be reinitialized with proper constants (GP-IB address, model number, calibration parameters) after replacement, followed by full recalibration per Operating Manual Chapter 6.
Ranked causes
- 1. Serial EEPROM chip U211 replaced (rare)
Enable calibration mode
- Test point
- GP-IB command
- Expected
- CAL 1 executed
15 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Full recalibration after component replacement
Electronic Load requires annual recalibration and whenever specified control circuit or power stage components (U5, U6, U13, U306-308, U320-323, U325, U326, U329, U331) are replaced, per Operating Manual Chapter 6.
Ranked causes
- 1. Control circuit component replaced or annual calibration due (common)
Refer to Operating Manual Chapter 6 for detailed calibration procedure
- Test point
- Operating Manual reference
- Expected
- Chapter 6 calibration procedure followed in full
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: IMON circuit after input power stage component replacement
Current monitoring circuit requires adjustment after any input power stage component (Figure 6-1, sheet 6) replacement to ensure correct current measurement.
Ranked causes
- 1. Input power stage component replaced (common)
Turn load off and prepare for measurement
- Test point
- Input terminals and top cover
- Expected
- Load powered off, terminals disconnected, cover removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC mode PARD out of specification
Periodic and random deviation (PARD) current measurement exceeds allowed rms limit during CC mode test
Ranked causes
- 1. Internal current control circuit noise or instability (unknown)
Measure CC mode PARD rms current
- Test point
- Electronic Load output
- Expected
- 6060B: ≤4mA rms; 6063B: ≤1mA rms, RMS in range 20Hz to 10MHz
- Tools
- DMM, current probe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC Mode PARD reading exceeds specification caution
DMM measurement of input current noise (CC mode PARD) exceeds stated limits when monitor is set to 10V and >10A
Ranked causes
- 1. Input circuit noise source or poor filtering (unknown)
Connect test setup as shown in Figure 2-3; set power supply to 10V and >10A
- Test point
- Power supply terminals
- Expected
- 10V output with >10A capacity
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
CC Mode Power Limit circuit does not activate within timeout high voltage
When Electronic Load power limiting test is run, overpower circuit does not turn load off within three minutes or power levels do not reach expected limits
Ranked causes
- 1. Overpower circuit malfunction or protection threshold miscalibration (common)
Connect Electronic Load and power source per Figure 2-2; turn on load with no input power and run 5 minutes
- Test point
- Electronic Load front panel
- Expected
- Load operating with zero power dissipation
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CR Mode resistance out of specification
Calculated Electronic Load resistance values fall outside expected ranges when tested in CR Mode across low, middle, and high ohm ranges
Ranked causes
- 1. Resistance range calibration drift (unknown)
Set up test circuit with power supply, Electronic Load, and 0.100 ohm current monitor resistor
- Test point
- Input terminals and monitor resistor
- Expected
- Circuit connected as shown in Figure 2-1
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -101 displayed: Secondary microprocessor RAM failure
Secondary microprocessor U301 internal RAM test failed during selftest; input disabled, Err annunciator on.
Ranked causes
- 1. Secondary microprocessor U301 internal RAM failure (rare)
Power cycle and observe error code
- Test point
- Front panel display
- Expected
- ERROR -101 appears for 2 seconds, then INPUT DOWN 1, INPUT DWN
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -103 displayed: Secondary microprocessor timer failure
Secondary microprocessor U301 internal timer test failed during selftest; input disabled, Err annunciator on.
Ranked causes
- 1. Secondary microprocessor U301 internal timer failure (rare)
Power cycle and observe error code
- Test point
- Front panel display
- Expected
- ERROR -103 appears for 2 seconds, then INPUT DOWN 1, INPUT DWN
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -104 displayed: EEPROM checksum failure
EEPROM U211 checksum error detected during secondary interface selftest; input disabled, Err annunciator on.
Ranked causes
- 1. EEPROM U211 checksum error (occasional)
Run calibration to regenerate checksum
- Test point
- Front panel display
- Expected
- ERROR -104 should not reappear after recalibration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -105 displayed: Main DAC high tolerance
Main DAC circuit U320/U326 zero or full scale point is above high tolerance level; input disabled.
Ranked causes
- 1. Main DAC circuit U320/U326 zero or full scale calibration out of high tolerance (occasional)
Verify error persists
- Test point
- Front panel display
- Expected
- ERROR -105 appears at power-on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -106 displayed: Main DAC low tolerance
Main DAC circuit U320/U326 zero or full scale point is below low tolerance level; input disabled.
Ranked causes
- 1. Main DAC circuit U320/U326 zero or full scale calibration out of low tolerance (occasional)
Verify error persists
- Test point
- Front panel display
- Expected
- ERROR -106 appears at power-on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -107 displayed: Transient DAC high tolerance
Transient DAC circuit U321/U325 zero or full scale point is above high tolerance level; input disabled.
Ranked causes
- 1. Transient DAC circuit U321/U325 zero or full scale calibration out of high tolerance (occasional)
Verify error persists
- Test point
- Front panel display
- Expected
- ERROR -107 appears at power-on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -108 displayed: Transient DAC low tolerance
Transient DAC circuit U321/U325 zero or full scale point is below low tolerance level; input disabled.
Ranked causes
- 1. Transient DAC circuit U321/U325 zero or full scale calibration out of low tolerance (occasional)
Verify error persists
- Test point
- Front panel display
- Expected
- ERROR -108 appears at power-on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -4 displayed for 2 seconds
Primary microprocessor read/write test to GP-IB talker/listener chip failed; input remains disabled with Err annunciator on.
Ranked causes
- 1. Primary microprocessor U203 read/write test to GP-IB talker/listener chip U202 failed (common)
Verify error code -4 still displays
- Test point
- Front panel display
- Expected
- ERROR -330 shown when shift key + key pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR -5 displayed for 2 seconds
Primary microprocessor internal trigger test failed; input remains disabled with Err annunciator on.
Ranked causes
- 1. Primary microprocessor U203 internal trigger circuit test failed (common)
Verify error code -5 still displays
- Test point
- Front panel display
- Expected
- ERROR -330 shown when shift key + key pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERROR UNKNOWN displayed during secondary selftest
Unknown error displayed; EEPROM may need initialization. Input disabled, Err annunciator on.
Ranked causes
- 1. EEPROM U211 not initialized or corrupted (rare)
Attempt EEPROM reinitialization
- Test point
- Front panel display
- Expected
- Error clears after reinitialization
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty component on data bus identified
S.A. troubleshooting has isolated a problem to a specific faulty component on data lines
Ranked causes
- 1. Defective IC on SD0-7 or SDB0-7 bus (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel readings do not match expected power limit test values high voltage
Electronic Load front panel shows voltage and current readings outside expected ranges during power limit test
Ranked causes
- 1. Metering circuit error or display malfunction (occasional)
Observe and record Electronic Load front panel V and A readings during power limit test
- Test point
- Front panel voltage and current display
- Expected
- 6060B: 33V ±tolerance, 13-17A; 6063B: 45V, 7.2-10.9A
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent circuit malfunction high voltage
Load unable to sink current or sinkcurrent exceeds limits at specified voltage thresholds
Ranked causes
- 1. U8 amplifier failure or incorrect bias (unknown)
Measure voltage at U8-6
- Test point
- U8-6
- Expected
- -7V at low input voltage, -0.8V at 65V input
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit does not shut down load on power limit fault high voltage
Electronic Load continues to dissipate power beyond protection threshold when power limiting is exceeded
Ranked causes
- 1. Overpower circuit component failure or threshold miscalibration (common)
Monitor load for timeout period during power limit test (step d onwards)
- Test point
- Electronic Load operation and shutdown response
- Expected
- Prot annunciator activates and load shuts down within 3 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit malfunction (CC mode) high voltage
Overpower protection circuit fails to shut down load or turn off input within specified time limit during power limit test
Ranked causes
- 1. Overpower circuit component failure or sensor fault (rare)
Verify front panel Prot annunciator illuminates during overpower condition
- Test point
- Front panel indicator
- Expected
- Prot annunciator ON when input overpower detected
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient and slew circuit operation check
Verify transient operation and slew circuit rise/fall time are within specification (60-100µs at high slew rate; 1.2-2.0ms at low slew rate). If out of spec, perform transient generator or slew circuit troubleshooting.
Ranked causes
- 1. Slew rise time or fall time out of specification (unknown)
Set test setup and power supply
- Test point
- Power supply output
- Expected
- 6060B: 10V and 10A; 6063B: 10V and 15A
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator does not perform functions
Transient circuit fails to operate; may not switch load levels or produce HIGH control signal to DAC circuits
Ranked causes
- 1. Data bus or internal clock failure (common)
Verify data bus signature at U310 and U311
- Test point
- U310, U311 SD 0-7 lines
- Expected
- Valid signature per Table 3-10
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator toggle or pulse mode malfunction
Toggle or pulse mode test fails; transient circuit cannot switch modes
Ranked causes
- 1. Software or firmware error in toggle/pulse mode (occasional)
Execute toggle mode command and verify mode switch
- Test point
- Front panel TRAN mode display or output
- Expected
- Toggle mode active; output switches between two load levels
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator will not change frequency
TRAN ON mode fails to change frequency when FSEL inputs are programmed
Ranked causes
- 1. FSEL input error or logic failure (common)
Verify FSEL logic levels at U316-1, 2, 3
- Test point
- U316 pins 1, 2, 3
- Expected
- Logic levels per FSEL table corresponding to desired frequency
- Tools
- Logic probe
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trigger circuit malfunction
Trigger pulses not generated or signal lines not toggling correctly during trigger circuit operation in pulse or toggle mode
Ranked causes
- 1. Defective gate or IC in trigger circuit (common)
Verify common ground connection for logic probe measurement
- Test point
- TP201-4 to TP301-2
- Expected
- Connected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
-101 | Secondary microprocessor U301 internal RAM failure | During secondary interface selftest sequence at power-on | error |
-102 | Secondary microprocessor U301 internal ROM failure or thermistor RT551 missing/open | During secondary interface selftest sequence at power-on | error |
-103 | Secondary microprocessor U301 internal timer failure | During secondary interface selftest sequence at power-on | error |
-104 | EEPROM U211 checksum error | During secondary interface selftest sequence when EEPROM data is verified | error |
-105 | Main DAC circuit U320/U326 zero or full scale point above high tolerance | During secondary interface selftest when main DAC is calibrated and tested | error |
-106 | Main DAC circuit U320/U326 zero or full scale point below low tolerance | During secondary interface selftest when main DAC is calibrated and tested | error |
-107 | Transient DAC circuit U321/U325 zero or full scale point above high tolerance | During secondary interface selftest when transient DAC is calibrated and tested | error |
-108 | Transient DAC circuit U321/U325 zero or full scale point below low tolerance | During secondary interface selftest when transient DAC is calibrated and tested | error |
-4 | Primary microprocessor U203 read/write test to GP-IB talker/listener chip U202 failed | During primary interface selftest sequence at power-on | error |
-5 | Primary microprocessor U203 internal trigger circuit test failed | During primary interface selftest sequence at power-on, after RAM/ROM/timer pass | error |
OV (Overvoltage) | Input voltage has exceeded the rated input voltage limit | Input voltage at terminals exceeds rated specification | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.