Agilent 6050A: troubleshooting and error codes
The ServiceForge knowledge base holds 51 documented symptoms for the Agilent 6050A, with 97 ranked causes, 113 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
DAC circuits not generating SLEW signal high voltage
Analog SLEW signal not produced by combined outputs from main DAC/amplifier (U320/U326) and transient DAC/amplifier (U321/U325); signal controls input power stages
Ranked causes
- 1. Chip select IC U304 malfunction (unknown)
- 2. Status readback IC U303 malfunction (unknown)
- 3. EEPROM decoder U341 malfunction (unknown)
- 4. Transient DAC/amplifier U321/U325 failure (unknown)
- 5. Main DAC/amplifier U320/U326 failure (unknown)
Measure signature at U304 pins 6 and 16 for +5V rail
- Test point
- U304-6,16
- Expected
- 1CAU
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit fault caution
Overpower circuit not responding correctly to input voltage/current; power limitation not activating or deactivating properly
Ranked causes
- 1. U10 comparator failure (occasional)
- 2. R142 summing resistor pack open or incorrect value (occasional)
- 3. RT2 temperature monitor or U327 failure (occasional)
- 4. +12V reference voltage low or absent (occasional)
Verify +12V reference voltage present
- Test point
- +12V reference input
- Expected
- +12V ±tolerance
- Tools
- DMM
Fans not running or not changing speed high voltage
Fan #1 and/or Fan #2 do not operate or do not respond to speed control commands; fans run continuously at maximum speed
Ranked causes
- 1. Fan motor defective or fan connector loose (common)
- 2. Fan speed control circuit failure; U245 or U233 bit logic error (occasional)
- 3. Module heatsink temperature sensor failure or excessive heat condition (occasional)
Verify fan connector power and signal at fan motor
- Test point
- Fan connector pins
- Expected
- Waveform present matching U243-1 normal power-on waveform
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel display remains blank at power-on
No display output when unit powers on; may indicate RAM, ROM, or internal timer failure in primary microprocessor
Ranked causes
- 1. Primary microprocessor U203 RAM failure (occasional)
- 2. Primary microprocessor U203 ROM failure (occasional)
- 3. Primary microprocessor U203 internal timer failure (occasional)
Measure square wave on SA_GATE line
- Test point
- TP201-8
- Expected
- 10Hz, 100Hz, 1KHz, or no square wave
- Tools
- Oscilloscope or logic probe
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect slew rate waveform output
Electronic Load slew circuit produces incorrect waveform output when various slew rates programmed. Waveforms checked to verify slew circuit switches operation.
Ranked causes
- 1. SDB0-7 data inputs or SLW signal outputs faulty (occasional)
- 2. Slew circuit amplifier or switch (U346, U317) defective (occasional)
- 3. CC/CV control circuit malfunction (rare)
Verify SA waveforms on Figure 3-3
- Test point
- SA waveform output
- Expected
- Waveforms match Figure 3-3 specifications
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input power stage malfunction
One or more of four/eight/sixteen input power stages not operating correctly; error amplifier waveform absent or incorrect at stage output
Ranked causes
- 1. Specific input power stage defective (common)
- 2. Multiple stages defective or shared supply problem (occasional)
- 3. SENSE switch incorrectly set (occasional)
Enable SA mode and verify error amplifier waveform
- Test point
- Error amplifier output in each input power stage
- Expected
- Waveform as shown in Figure 3-6
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input power stage not functioning in specific module caution
One of four, eight or sixteen identical input power stages (FET Q2, monitor amplifier U6, error amplifier U5) is defective or producing incorrect waveforms
Ranked causes
- 1. Power FET (Q2) failure or open (common)
- 2. Fuse F5 or F13 open (stage Q2D) (common)
- 3. Error amplifier (U5) or monitor amplifier (U6) failure (occasional)
Verify error amplifier output waveform in SA mode
- Test point
- Error amplifier output of defective stage, refer to Power Board schematic for pin location
- Expected
- Waveform as shown in Figure 3-6
- Tools
- Oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent circuit limiting maximum input current abnormally
Load unable to sink expected maximum current; circuit limits sink capability below normal range
Ranked causes
- 1. Input voltage below 6.3 V (model dependent) (common)
- 2. Overvoltage (OV) circuit activation at high input voltage (occasional)
- 3. Delayed bias voltage (-15Vx) not present on startup (occasional)
Verify input voltage and measure at test point (per Table 3-3)
- Test point
- Test points specified in Table 3-3
- Expected
- Refer to Table 3-3 measurement conditions and readings
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent circuit malfunction caution
Maximum current limit not functioning correctly across input voltage range or power conditions
Ranked causes
- 1. Diode D53 failure (occasional)
- 2. Diode D54 failure (occasional)
- 3. Amplifier U14 or transistor Q11/Q12 failure (occasional)
Measure voltage at U14-6 with input voltage at or below 6.3V
- Test point
- U14-6
- Expected
- approximately -7 volts
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary microprocessor U301 internal ROM failure (Error -102)
Secondary microprocessor ROM selftest failure or power board disconnection; secondary processor ROM or power supply defect
Ranked causes
- 1. Power board disconnected (common)
- 2. Thermistor RT2 missing or open circuit (occasional)
- 3. U301 internal ROM defect (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signature analysis testing needed for troubleshooting caution
Microprocessor-based circuit faults isolated using signature analyzer to compress bit streams into four-character signatures for comparison against known correct values
Ranked causes
- 1. Faulty component in digital circuit producing incorrect signature (common)
- 2. Shorted pin or node to Vcc or ground (occasional)
- 3. Two pins shorted together (occasional)
Take signature at applicable test point using signature analyzer
- Test point
- Primary test points per Tables 3-4 through 3-9; secondary per Tables 3-10 through 3-14
- Expected
- Four-character signature matching reference table for applicable component (U203, U205, U206, U301)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator produces incorrect or no output frequency
Transient mode enabled but output frequency at U314-7 is wrong or absent; expected factory default is 1 kHz at 50% duty cycle
Ranked causes
- 1. Incorrect FSEL inputs or U311 output (common)
- 2. U312 or U313 output malfunction (occasional)
- 3. Transient trigger circuit failure (occasional)
Enable transient mode and verify TRANS_EN signal condition
- Test point
- TRANS_EN signal
- Expected
- Per Table 3-3 (not provided in excerpt)
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails power-on self-test high voltage
At power-on, front panel LCD displays error number or message. Unit does not complete normal initialization. Load input remains disabled.
Ranked causes
- 1. Primary circuit failure (GPIB interface or primary processor) (unknown)
- 2. Secondary circuit or module failure (unknown)
- 3. AC input or bias supply failure (unknown)
Observe power-on display
- Test point
- Front panel LCD
- Expected
- Normal turn-on self-test indication or error code
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails selftest with error 105-108 at turn-on
System displays error code 105-108 during power-on self-test sequence
Ranked causes
- 1. IMON adjustment out of specification (occasional)
- 2. Switches in U309 not operating properly (occasional)
- 3. Faulty DAC or amplifier generating incorrect waveform (rare)
Use Signature Analyzer (SA) to diagnose error 105-108
- Test point
- Test points specified in SA documentation
- Expected
- Valid signatures per SA mode table
- Tools
- Signature Analyzer
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC line trigger signal not toggling
Waveforms at U243-14 and Q211-11 are incorrect or absent when sending TRIG:SOUR LINE command
Ranked causes
- 1. Defective GPIB board IC or gate (unknown)
- 2. Defective Control board IC (unknown)
Measure waveform at U243-14 and Q211-11
- Test point
- U243-14, Q211-11
- Expected
- Correct waveform present
- Tools
- Oscilloscope
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All input power stages have circuit problems caution
Multiple or all input power stages are failing simultaneously; affects power delivery and load regulation
Ranked causes
- 1. Common supply voltage or control signal failure (common)
- 2. SENSE switch S1 not set to LCL for local sensing (occasional)
Verify SENSE switch position
- Test point
- S1 switch on rear panel
- Expected
- Set to LCL position if remote sensing not used
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DAC circuit waveforms incorrect
SA mode waveforms shown in Figure 3-3 are not correct; SLEW signal generation from main and transient DAC/amplifiers affected
Ranked causes
- 1. Faulty readback DAC U322 data bus lines SD0-7 (common)
- 2. Faulty main or transient DAC data lines SDB0-7 (common)
Verify SA mode is enabled
- Test point
- Jumpers TP201 and TP301
- Expected
- Jumpers connected in SA mode positions
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM U342 checksum error (Error -104)
EEPROM checksum failed during selftest; calibration data corrupted or EEPROM defective
Ranked causes
- 1. EEPROM calibration data corrupted (common)
- 2. EEPROM U342 data line failure (occasional)
Reprogram and reinitialize EEPROM
- Test point
- EEPROM U342
- Expected
- Checksum restored; error -104 does not reappear
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module not installed or secondary communication failure (Error -240)
No module installed in mainframe, or communication failure between primary and secondary processor; INPUT DWN message displayed
Ranked causes
- 1. No module installed in mainframe (common)
- 2. SRX or STX serial data line failure (occasional)
Verify serial data pulses on SRX and STX lines
- Test point
- U333-7 (SRX), U334-3 (STX)
- Expected
- Data pulses toggling between 0V and 5V
- Tools
- Logic probe
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit active - power limiting caution
Load operating in overpower condition; power sinking capability limited by thermal or time duration threshold
Ranked causes
- 1. Overpower condition exists - EPU bit false (common)
- 2. Overpower condition exists - EPU bit true (common)
Check test points and reference voltage using Table 3-3 conditions
- Test point
- U17 amplifier, U10 comparators, R142 summing resistor pack, +12V reference
- Expected
- Values per Table 3-3
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower detection and load limiting failure caution
Load continues to draw power beyond safe limits or fails to enter overpower protection mode when limits exceeded
Ranked causes
- 1. Comparator circuit U10 malfunction or output not reaching controller (occasional)
- 2. Amplifier U12 not summing signals correctly (rare)
Measure comparator U10 output voltage at each of four comparator stages
- Test point
- U10 outputs (per Figure 3-10)
- Expected
- Logic high/low transitions when input signals cross threshold
- Tools
- Oscilloscope, logic probe
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PROG signal not going negative after startup high voltage
Q11 remains on after startup; PROG stays high; input power FETs remain off
Ranked causes
- 1. Delayed bias voltage (-15Vx) not present or delayed excessively (occasional)
- 2. Q12 not turning off when -15Vx arrives (rare)
Measure -15Vx bias voltage after power-on delay (approximately 5-10 seconds)
- Test point
- -15Vx supply node
- Expected
- -15 V ±10% after startup delay, ±10%
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slew circuit generates incorrect waveforms
SA mode waveforms at oscilloscope are incorrect or do not match expected patterns for slew circuit function
Ranked causes
- 1. Incorrect SDB0-7 data inputs or SLW signal outputs from U305 (common)
- 2. Defective amplifier or switch in slew circuit (occasional)
Verify SA waveforms at oscilloscope
- Test point
- Slew circuit outputs per Figure 3-4
- Expected
- Waveforms matching SA mode patterns shown in Figure 3-4
- Tools
- Digital oscilloscope
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slew rate waveforms not generated in normal mode
Programmed slew rate waveforms (0.001, 0.5, 2.5 A/µs) are not appearing at oscilloscope after front panel configuration
Ranked causes
- 1. Slew circuit switches (U317) not operating correctly (common)
- 2. CC/CV control circuit fault (occasional)
Configure front panel for slew rate testing
- Test point
- Front panel controls
- Expected
- MODE=CURR, CURR=5, Tran Level=10, Freq=80, Dcycle=50, Tran on/off=on
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator does not perform any functions
Transient generator circuits (U349-U355) are non-functional; unable to switch between load levels or generate HIGH control signal to DAC circuits
Ranked causes
- 1. Data bus or internal clock failure (common)
- 2. Defective transient generator IC (U349-U355) (occasional)
Verify 1MHz clock signal at specified pins
- Test point
- U349-1, U350-1, U351-13, U352-4, U352-12
- Expected
- 1MHz clock signal present
- Tools
- Oscilloscope
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator malfunction
Transient generator circuits not switching load levels correctly or not producing HIGH control signal
Ranked causes
- 1. Data bus or internal clock failure (common)
- 2. Frequency control logic error (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator toggle or pulse mode not functioning
Toggle and pulse mode operations do not respond to mode commands; output signals absent or incorrect at mode-specific test points
Ranked causes
- 1. Mode control signal path failure (common)
- 2. Output stage IC (U349, U350, U351, U352) defect (occasional)
Verify toggle mode control signals
- Test point
- U349-3, U349-4, U350-3, U349-14, U351-9, U351-11, U351-14, U352-13
- Expected
- Toggle mode: U349-3=LO, U349-4=HI, U350-3=LO, U349-14=toggling, U351-9/11/14=LO, U352-13=toggling, U351-12=HI
- Tools
- Logic probe, oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient generator will not change frequency
Pressing TRAN ON and programming transient frequencies does not result in frequency change; unit remains at factory default or incorrect frequency
Ranked causes
- 1. FSEL input signal loss or corruption (common)
- 2. Pulse interval timing circuit failure (occasional)
Verify FSEL input logic levels
- Test point
- U351-1, U351-2, U351-3
- Expected
- Logic levels per FSEL table for programmed frequency
- Tools
- Logic probe
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails front panel verification tests high voltage
After passing self-test, unit fails verification of load functions. Some functions are not calibrated or not operating properly.
Ranked causes
- 1. Load function not calibrated (unknown)
- 2. Load function not operating properly (unknown)
Perform front panel verification procedures
- Test point
- Load control functions
- Expected
- All functions operate as specified in Operating Manual Chapter 3
Bus trigger signal not toggling
Trigger signal not generated when using TRIG command or *TRG common command via HPIB
Ranked causes
- 1. Defective bus trigger circuit on GPIB board (unknown)
Verify READ_TRIG level at power-on
- Test point
- U210-11 (READ_TRIG)
- Expected
- High
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration disable jumper configuration caution
Jumper RTP201 setting to prevent unauthorized calibration commands on primary microprocessor
Ranked causes
- 1. Unauthorized calibration performed on unit (unknown)
Verify calibration security jumper position
- Test point
- Primary test header TP201 pins 5 and 6
- Expected
- Jumper RTP201 installed between pins 5 and 6 for lockout; removed for normal calibration access
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current Monitor (IMON) circuit
Adjustment of IMON circuit after input power stage component replacement. Connect DMM between test points and adjust potentiometer for correct reading.
Ranked causes
- 1. Input power stage component replaced (common)
Measure voltage at IMON test points
- Test point
- J2-1 and J2-3
- Expected
- 0 ±0.5 millivolts mV, ±0.5 mV
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EEPROM initialization
Programming serial EEPROM chip U211 with GPIB address, model number, and factory constants after GPIB board or EEPROM replacement. Run commands in sequence then recalibrate.
Ranked causes
- 1. GPIB board assembly or EEPROM chip U211 replaced (occasional)
Program EEPROM initialization commands
- Test point
- Serial port or GPIB interface
- Expected
- Commands accepted without error
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Overpower circuit signal scaling caution
Verify signal scaling of input voltage and current via resistor network to set correct power limits
Ranked causes
- 1. Signal scaling resistor values out of tolerance or wrong part installed (occasional)
Measure input voltage signal level at U12 summing amplifier input
- Test point
- U12 pin (voltage input to amplifier)
- Expected
- Signal proportional to input voltage per schematic specification
- Tools
- Multimeter, oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive input current limiting at power-on
Current limited to lower than normal maximum during power-on even before high voltage reached
Ranked causes
- 1. SPCLR signal not releasing Q11 properly at end of power-on sequence (rare)
Monitor SPCLR signal and Q11 gate voltage during power-on sequence
- Test point
- SPCLR input and Q11 base
- Expected
- SPCLR high during initialization, low after ~100-500 ms; Q11 transitions off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External trigger signal not responding
Unit does not respond to external trigger input when TRIG:SOUR EXT is set and TRIG_IN is asserted
Ranked causes
- 1. Defective external trigger input circuit (unknown)
Verify external trigger input pulse at U207-10
- Test point
- U207-10
- Expected
- Pulses when TRIG_IN shorted to common
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GET function trigger not generating pulses
Trigger pulses not generated when using GET function for group enable trigger
Ranked causes
- 1. Defective P_TRIG output or firmware gate (unknown)
Verify P_TRIG pulse output
- Test point
- U208-12 (P_TRIG)
- Expected
- Pulses high
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INPUT DWN message repeated on display
INPUT DWN or INPUT DWN 1 message appears at power-on; secondary communication or AC power issue
Ranked causes
- 1. Line switch cycled repetitively or abnormal line conditions (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal timer trigger signal not pulsing
Timer-based trigger pulses not generated when sending TRIG:SOUR TIM command
Ranked causes
- 1. Defective timer circuit on GPIB board (unknown)
Verify timer enable signal
- Test point
- U228-12 (TRIG_EN)
- Expected
- Goes high
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load power mode selection (normal vs. extended)
Circuit selects between normal and extended power ranges via EPC_EN signal; extended mode requires external power connection
Ranked causes
- 1. External power unavailable or not connected (common)
Query EPU (Extended Power Unavailable) bit status
- Test point
- Status register bit 9
- Expected
- EPU bit false when extended power available; true when unavailable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main DAC circuit zero or full scale out of tolerance (Error -105, -106)
Main DAC U320/U326 zero point or full scale calibration exceeds tolerance limits during selftest
Ranked causes
- 1. Main DAC circuit calibration error or component failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit malfunction caution
Overpower limiting not activating when input power exceeds capability
Ranked causes
- 1. Overvoltage detection circuit failure at 75V threshold (unknown)
Monitor IPROG voltage when input reaches 75V
- Test point
- IPROG
- Expected
- transitions to more negative voltage (Low state)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpower circuit troubleshooting
Verify overpower protection circuit operation which limits load power sinking to normal (300-396 W) or extended (400-530 W) range based on heat sink temperature
Ranked causes
- 1. Extended power capability signal malfunction or missing external power (unknown)
Monitor output voltage at overpower comparator circuit test point
- Test point
- Comparator output on Figure 3-10
- Expected
- Voltage oscillates between -14V and 0V when circuit is active, -14 to 0 V
- Tools
- Oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary microprocessor read/write test failure (Error -4)
GPIB talker/listener chip communication failure detected during selftest; primary interface malfunction
Ranked causes
- 1. GPIB talker/listener chip U202 address or data line failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary trigger circuit test failure (Error -5, -6, -7)
Internal trigger, AC line trigger, or timer circuit trigger test failed during selftest; trigger logic malfunction
Ranked causes
- 1. Primary trigger circuit malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Risk of electrostatic discharge damage caution
Potential ESD damage to components during service procedures when protective covers are removed and power is applied
Ranked causes
- 1. Inadequate ESD protection measures during service (common)
Verify workstation meets ESD requirements
- Test point
- Work surface and grounding system
- Expected
- Distributed resistance 10^6–10^12 Ω/sq; wrist strap 1 MΩ resistor to ground
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary microprocessor U301 internal RAM failure (Error -101)
Secondary microprocessor internal RAM selftest failure; secondary processor memory defect
Ranked causes
- 1. U301 internal RAM defect (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary microprocessor U301 internal timer failure (Error -103)
Secondary microprocessor internal timer selftest failure; secondary processor timer circuit defect
Ranked causes
- 1. U301 internal timer defect (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SLEW readback signal absent or incorrect
Readback DAC/amplifier U322/U328 not converting SDB0-7 data to reference signal for microprocessor SLEW verification during selftest
Ranked causes
- 1. Readback DAC/amplifier U322/U328 failure (unknown)
Measure readback signal at comparator U327 output
- Test point
- U327 output
- Expected
- Signal tracking main SLEW signal during operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient DAC circuit zero or full scale out of tolerance (Error -107, -108)
Transient DAC U321/U325 zero point or full scale calibration exceeds tolerance limits during selftest
Ranked causes
- 1. Transient DAC circuit calibration error or component failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient mode trigger pulses not generated
Running the trigger test program does not produce 1.5µs TRIG pulses at expected test points
Ranked causes
- 1. Trigger circuit failure (unknown)
Execute trigger test program (lines 10-70)
- Test point
- U311-12, U314-1
- Expected
- 1.5µs TRIG pulses generated continuously
- Tools
- BASIC program, oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
-101 | Secondary microprocessor U301 internal RAM failure | RAM memory malfunction detected during secondary selftest part 1 | error |
-102 | Power board disconnected, thermistor RT2 missing/open, or U301 internal ROM failure | Power supply verification or ROM selftest failure during secondary selftest part 1 | error |
-103 | Secondary microprocessor U301 internal timer failure | Timer circuit malfunction detected during secondary selftest part 1 | error |
-104 | EEPROM U342 checksum error | EEPROM checksum verification failed during secondary selftest part 2 | warning |
-105 | Main DAC U320/U326 zero point above high tolerance level | DAC calibration selftest failure during secondary selftest part 2 | error |
105-108 | Selftest failure error code | Error detected during power-on self-test sequence | error |
-106 | Main DAC U320/U326 zero point below low tolerance level | DAC calibration selftest failure during secondary selftest part 2 | error |
-107 | Transient DAC U321/U325 zero point above high tolerance level | Transient DAC calibration selftest failure during secondary selftest part 2 | error |
-108 | Transient DAC U321/U325 zero point below low tolerance level | Transient DAC calibration selftest failure during secondary selftest part 2 | error |
-240 | Hardware error (secondary component failure) | EEPROM, MAIN DAC, or TRANSIENT DAC failure during secondary selftest part 2 | error |
-330 | Self-test fail (primary or secondary selftest failure) | Primary or secondary selftest failure during power-on sequence | error |
-4 | Primary microprocessor read/write test to GPIB talker/listener chip failed | GPIB talker/listener chip U202 address or data line communication failure | error |
-5 | Primary microprocessor internal trigger circuit test failed | Trigger signal generation or detection failure during selftest | error |
-6 | Primary microprocessor AC line trigger circuit test failed | AC line trigger detection failure during selftest | error |
-7 | Primary microprocessor timer circuit trigger test failed | Timer-based trigger generation failure during selftest | error |
Error 105-108 | Slew circuit malfunction detected during self-test | Self-test at turn-on detects error 105-108 | error |
Fan inoperative or stuck at max speed | Fan #1 or Fan #2 does not operate or fails to respond to speed control | User reports no fan operation or continuous maximum fan speed; Power-On diagnostics may flag fan failure | warning |
Fan speed does not change | Fan runs but speed command does not change fan RPM | User inputs speed command via front panel keypad; fan does not respond to speed changes | warning |
Frequency Selection (FSEL) Table mismatch | Programmed frequency does not match U311-5 output as per FSEL table | Transient mode enabled with specific FSEL pin states | error |
UNKNOWN | EEPROM not initialized or configuration data missing | EEPROM uninitialized or corrupted configuration on power-on | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI