Agilent 6030A: troubleshooting and error codes
The ServiceForge knowledge base holds 44 documented symptoms for the Agilent 6030A, with 63 ranked causes, 146 diagnostic checks with test points and expected values, and 15 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power supply malfunction or self-test failure high voltage
Unit fails power-on self-test, exhibits abnormal operation, or does not function correctly on startup
Ranked causes
- 1. Input power unavailable or power cord/breaker issue (common)
- 2. Incorrect mode switch A2S1 settings (common)
- 3. Loose or interrupted connections to supply or external devices (occasional)
- 4. Incorrect GPIB address switch A8S1 setting (occasional)
Verify input AC power availability at wall outlet
- Test point
- AC mains connection point
- Expected
- 110-120 VAC at wall outlet
- Tools
- DMM or known working device
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary interface clock or address/data signals not toggling caution
Clock signals at Y2 or U14 oscillator output absent, incorrect frequency, or amplitude; address and data lines A0-A15, D0-D7 stuck high or low; system not processing data correctly.
Ranked causes
- 1. Clock oscillator defective or oscillation not starting (rare)
- 2. Data or address line shorted to logic high or low (common)
- 3. Socketed IC defective (U6, U8, U12, U14, U16, U17, U36) (occasional)
- 4. GPIB (A8) assembly defective (rare)
Measure clock frequency at Y2 oscillator output (12 MHz)
- Test point
- C7+, C8+
- Expected
- 12MHz (see waveform Figure 3-3)
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signature analysis procedure for microprocessor circuits
Method to compress bit stream into four-character signature for each node to isolate faults in microprocessor-based power supply
Ranked causes
- 1. Incorrect signature at input pin with correct source signature (occasional)
- 2. Node shorted to Vcc or ground (occasional)
- 3. Two pins shorted together (occasional)
- 4. Faulty component or loading short circuit (occasional)
Verify unit enters SA mode correctly
- Test point
- Front panel display
- Expected
- SA SA displayed, all LEDs on
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supply +15V out of regulation caution
Output voltage not 15V or regulation poor. Downstream analog circuits fail or operate erratically.
Ranked causes
- 1. +15V load shorted to ground (common)
- 2. Zener diode A2VR3 open or degraded (occasional)
Measure reference voltage at A2U11-3 (ADJ)
- Test point
- A2U11-3 (ADJ)
- Expected
- 1.25 Vdc, ±0.05 V
- Tools
- Multimeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supply +5V out of regulation caution
Output voltage not 5V or ripple/noise excessive on +5V supply. PWM or downstream circuits fail to operate.
Ranked causes
- 1. +5V load shorted to ground (common)
- 2. PWM regulator A2U22 failure (occasional)
Measure +5V output at A2U22-7
- Test point
- A2U22-7
- Expected
- ≈ 2 to 4 Vdc sawtooth, 40 KHz, 2-4V
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current programming and readback accuracy high voltage
Verify current programming and readback functions within specification using GPIB controller; measure shunt resistor voltage to confirm output current setting and readback
Ranked causes
- 1. Current programming calibration (low current setpoint) (common)
- 2. Current programming calibration (full current setpoint) (common)
Measure shunt voltage at low setpoint
- Test point
- Across shunt resistor Rm
- Expected
- 4.75 to 5.25mV for 6030A, 4.75-5.25mV
- Tools
- DMM
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power limit high voltage
Adjustment of A2R25 and A2R26 potentiometers to set lower and upper power limit knee voltages
Ranked causes
- 1. Power limit knee calibration required (A8R55 must be within spec) (unknown)
- 2. Upper power limit knee calibration required (common)
Connect autotransformer to AC input; set DVM across input power rails
- Test point
- A1R3 rear (positive) to A1R1 rear (negative)
- Expected
- 240Vdc on input power rail during entire calibration
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC-to-DC converter FET drive failure high voltage
FET switches A4Q1-Q4 not switching properly or PWM on/off pulses not present at gate drives. Output voltage not regulating.
Ranked causes
- 1. FET gate drive pulse missing or malformed (common)
- 2. FET switches shorted or open (occasional)
Measure PWM-ON pulse at A2J7-26
- Test point
- A2J7-26
- Expected
- 1.7 µs, 20 KHz pulse
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect output voltage or overrange condition high voltage
Output voltage does not match expected value during OVP and PWM testing; overrange or incorrect CV mode response
Ranked causes
- 1. OVP comparator circuit failure or incorrect threshold (common)
- 2. PWM or clock circuit malfunction (common)
Measure bias voltages on OVP circuit
- Test point
- A2J7-7
- Expected
- 0.1V (6030A)
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maximum output voltage below specification
Output voltage cannot reach rated maximum
Ranked causes
- 1. CV circuit fault or low ac mains (common)
- 2. Transformer A1T1 primary or secondary failure (occasional)
Measure ac mains input voltage
- Test point
- Input transformer primary
- Expected
- Nominal mains (e.g., 115V or 230V per configuration)
- Tools
- Voltmeter with isolation transformer per Figure 3-5
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output from power supply high voltage
Supply produces no output voltage; PFETs or control circuits may have failed
Ranked causes
- 1. A4 PFET failure (Q1, Q2, Q3, Q4) (common)
- 2. Control circuit failure (CV or CC oneshots) (occasional)
Verify 320 Vdc rail voltage present
- Test point
- rear end of A1R3 (+), rear end of A1R1 (-)
- Expected
- 320 Vdc
- Tools
- Voltmeter with floating inputs
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output short circuit or overcurrent protection active
Supply enters current limit or shuts down; no load draws full current
Ranked causes
- 1. CC circuit failure preventing normal operation (common)
- 2. Power limit comparator failure (occasional)
Measure CC CONTROL voltage
- Test point
- A2CR11 cathode
- Expected
- Voltage indicating CC operation when overloaded
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary SA jumper setup failure
Unit does not enter SA mode with J5 jumper set, display does not show SA SA
Ranked causes
- 1. Data or address line shorted high or low (occasional)
- 2. U14 display processor faulty (occasional)
Set J5 jumper to SA mode position
- Test point
- J5 pins 3 and 4
- Expected
- Jumper pins 3 and 4 connected for SA mode
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC turn-on circuits malfunction high voltage
Issues with AC fault detection, relay closure timing, or PWM enablement during power-up sequence. Measured via AC FAULT signal and relay control nodes.
Ranked causes
- 1. AC power supply voltage out of spec (common)
Measure 5V supply at A2J7-24
- Test point
- A2J7-24
- Expected
- 5.0 Vdc
- Tools
- Multimeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supply -15V out of regulation caution
Output voltage not -15V or regulation poor. Analog circuits requiring negative supply fail.
Ranked causes
- 1. -15V load shorted to ground or regulator failed (common)
Measure regulator setpoint at A2U12-3 (ADJ)
- Test point
- A2U12-3 (ADJ)
- Expected
- - 1.25 Vdc, ±0.05 V
- Tools
- Multimeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration or component adjustment needed after repair high voltage
After component replacement, voltage levels or waveforms are out of specification and require calibration
Ranked causes
- 1. Component replacement alters circuit calibration (common)
Measure voltage at control board test fingers
- Test point
- Test fingers on control board (see Table 3-9)
- Expected
- Voltage levels per Chapter 6 schematic diagrams
- Tools
- DMM, oscilloscope per Chapter 2
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Common Mode high voltage
Set common mode calibration on A2 control board constant voltage circuit
Ranked causes
- 1. Common mode calibration adjustment needed (common)
Set control commands
- Test point
- Via GPIB
- Expected
- VSET 0; ISET 0; OUT OFF
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current full scale caution
Adjustment of A8R55 potentiometer to calibrate constant current output at full scale
Ranked causes
- 1. CC full scale calibration required (A8R29 and A2R9 must be within spec) (unknown)
Connect Rm current-monitoring shunt across output terminals
- Test point
- Power supply output terminals
- Expected
- 1mΩ (6031A); 10mΩ (6030A, 6032A); 100mΩ (6035A); 0.05% or better
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current load effect (load regulation) high voltage
Measure change in output current resulting from load resistance change from short-circuit to full-load in constant current mode; verify current stability
Ranked causes
- 1. Load regulation adjustment (common)
Verify CC mode operation at preset load resistance
- Test point
- Unit CC LED and AMPS display
- Expected
- CC LED lit, AMPS display reads approximately current setting
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current PARD ripple and noise measurement high voltage
Measure root-mean-square (RMS) output current noise in frequency range 20Hz to 20MHz in constant current operation; verify ripple and noise within specification
Ranked causes
- 1. Constant current PARD measurement setup (common)
Verify CC LED lit at full output current
- Test point
- Unit CC LED
- Expected
- CC LED illuminated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current source effect (line regulation) high voltage
Measure change in output current resulting from variation of AC input voltage from minimum to maximum; verify current stability across line voltage range
Ranked causes
- 1. Line regulation adjustment (common)
Verify CC mode at nominal line voltage with adjusted load
- Test point
- Unit CC LED
- Expected
- CC LED lit
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current zero caution
Adjustment of A8R29 potentiometer to calibrate constant current zero offset using test circuit
Ranked causes
- 1. CC zero offset out of specification (unknown)
Connect test circuit per Figure 2-3
- Test point
- Power supply output and test load
- Expected
- Test circuit connected as shown in Figure 2-3
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant voltage full scale caution
Adjustment of A8R58 potentiometer to calibrate constant voltage output at full scale for 6030/6031/6032/6035A
Ranked causes
- 1. CV full scale calibration required (unknown)
Remove external test circuits
- Test point
- Power supply output terminals
- Expected
- No external loads connected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant voltage zero caution
Adjustment of A8R40 potentiometer to calibrate constant voltage zero offset with external power supply connected
Ranked causes
- 1. CV zero offset out of specification (unknown)
Send VSET 0 command
- Test point
- Device command interface
- Expected
- Command accepted; output voltage set to 0V
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current monitor full scale caution
Adjustment of A2R9 potentiometer to calibrate current monitor output at full scale
Ranked causes
- 1. I-MON full scale calibration required (A2R8 must be within spec first) (unknown)
Connect Rm current-monitoring shunt across output terminals
- Test point
- Power supply output terminals
- Expected
- 1mΩ (6031A); 10mΩ (6030A, 6032A); 100mΩ (6035A); 0.05% or better
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current monitor zero caution
Adjustment of A2R8 potentiometer to calibrate current monitor zero offset
Ranked causes
- 1. Current monitor zero offset out of specification (unknown)
Send command to set voltage and current to zero with output off
- Test point
- Device command interface
- Expected
- VSET 0; ISET 0; OUT OFF command accepted
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Load transient recovery time measurement high voltage
Measure time for output voltage to return within specified band after step load change in constant voltage operation; displayed on oscilloscope with ac coupling
Ranked causes
- 1. Load transient recovery test setup (common)
Configure load for transient variation
- Test point
- Electronic load output
- Expected
- Current variation between 15.3 and 17Adc at 30Hz for 6030A
- Tools
- Electronic load with programmable capability
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Peak-to-peak PARD voltage measurement high voltage
Verify output voltage ripple and noise (PARD) in peak-to-peak measurement using oscilloscope differential setup with blocking capacitors and matched coaxial cables
Ranked causes
- 1. PARD measurement setup calibration (common)
Set oscilloscope impedance and bandwidth
- Test point
- Oscilloscope input
- Expected
- 50Ω impedance, 20MHz bandwidth
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Readback zero adjustment caution
Adjustment of A8R51 potentiometer to set readback zero reference voltage for different 6030/6031/6032/6035A models
Ranked causes
- 1. Readback zero offset out of specification (unknown)
Monitor controller display during A8R51 adjustment
- Test point
- Controller display readout
- Expected
- 0 and 50mV (6030A); 0 and 5mV (6031A); 0 and 15mV (6032A); 0 and 125mV (6035A)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote readback full scale caution
Adjustment of A8R61 potentiometer to calibrate remote readback output to match actual power supply voltage
Ranked causes
- 1. Remote readback full scale offset out of specification (unknown)
Run VOUT readback program on controller
- Test point
- Controller display
- Expected
- Program executing; VOUT values displayed continuously
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote Readback Zero high voltage
Perform remote readback zero calibration on A8 GPIB board voltage monitor buffer
Ranked causes
- 1. Remote readback zero adjustment required (common)
Connect external supply per diagram
- Test point
- Power supply inputs
- Expected
- Connected as shown in Figure 2-2
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Secondary interface voltage and current DACs caution
Verification and calibration of voltage control DAC (CV DAC), constant current DAC (CC DAC), and readback DAC circuits in secondary interface via front panel controls and measurement at specified nodes.
Ranked causes
- 1. CV DAC output out of specification (unknown)
Measure CV DAC output at U69-6 with voltage set to zero
- Test point
- U69-6
- Expected
- 0V
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Voltage monitor full scale caution
Adjustment of A8R75 potentiometer to calibrate voltage monitor output at full scale
Ranked causes
- 1. Voltage monitor full scale calibration required (unknown)
Attach DVM from M to A2J7 pin 3 (V-MON buffered)
- Test point
- M rear panel to A2J7 pin 3
- Expected
- DVM connected per Figure 2-1
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Voltage Monitor Zero high voltage
Set voltage monitor zero calibration per procedure on A2 control board
Ranked causes
- 1. Calibration adjustment required (common)
Set control commands
- Test point
- Via GPIB
- Expected
- VSET 0; ISET 0; OUT OFF
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel address latch and decoder signatures
Signature verification at address latches U13-U17 and decoders that enable shift registers
Ranked causes
- 1. Address latch or decoder fault (unknown)
Measure address control inputs
- Test point
- U14-26, U14-35, U14-38, U15-9, U17-3,4,5,6
- Expected
- U14-26 toggling, U14-35 37F8, U14-38 1ABC, latches transition Lo-Hi or Hi-Lo at power cycle
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel LED display driver signatures
Signature measurements at 7-segment display driver registers for current and voltage displays
Ranked causes
- 1. Incorrect display signature (unknown)
Verify display driver inputs at reset
- Test point
- U1 to U10 pin 1
- Expected
- 6H15 after power cycle, Lo to Hi transition at ~160 ms
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel RPG switch and latch signatures
Signature measurements for rotary pulse generator latch U11, flip-flop U12 and gate U18 for status reading
Ranked causes
- 1. RPG latch or input port fault (unknown)
Measure RPG latch inputs at rest
- Test point
- U11-1, U11-2, U12-4, U12-10, U16-10
- Expected
- U11-1 1ABC, U11-2 toggling unstable, U12-4,10 toggling unstable, U16-10 9H84
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current oscillation on CC loop caution
CC output oscillates or ripples excessively instead of remaining stable at set point.
Ranked causes
- 1. CC differentiator circuit instability (occasional)
Measure I-MON current monitor voltage at A2U4-8
- Test point
- A2U4-8
- Expected
- 0.88 Vdc (6030A), 0.125 Vdc (6031A), 0.30 Vdc (6032A), 1.00 Vdc (6035A)
- Tools
- Multimeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage oscillation on CV loop caution
CV output oscillates or ripples excessively instead of remaining stable.
Ranked causes
- 1. CV Error Amp feedback or loop instability (occasional)
Measure V-MON voltage at A2U5-1
- Test point
- A2U5-1
- Expected
- 3.75 Vdc (with VP=0)
- Tools
- Multimeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary processor signature analysis
Signature measurements at processor address, data and control lines for revision A.00.00 through A.00.04 ROM
Ranked causes
- 1. Incorrect address line signature (unknown)
Measure signature at address line A0
- Test point
- U16-12 or U6-12 or U8-12 or U12-11 depending on revision
- Expected
- A46A for rev A.00.04, A46A for A.00.02, A46A for A.00.01, A46A for A.00.00
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Readback DAC circuits not functioning or waveforms incorrect
Readback DAC output and multiplexer waveforms do not match expected oscilloscope traces; voltage readback measurements incorrect during self-test and front panel operation.
Ranked causes
- 1. Readback DAC defective or multiplexer not routing signals correctly (unknown)
Obtain readback DAC waveform during power-on self-test
- Test point
- U24-7 (turn-on self-test output)
- Expected
- Waveform as shown in Figure 3-4
- Tools
- Oscilloscope
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary processor signature analysis setup
Procedure to place secondary circuit in SA mode by jumper connection and measure circuit signatures
Ranked causes
- 1. Secondary circuit not entering SA mode (rare)
Configure signature analyzer for secondary circuit
- Test point
- SA clock, start, stop, ground connections
- Expected
- Clock at U4 pin 23, start at U4 pin 22, stop at U4 pin 22, ground at U4 pin 20
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshooting Option 001 blank front panel unit
Power supply has blank front panel with only LINE switch, OVP ADJUST control, and pilot light; front panel unavailable for problem isolation
Ranked causes
- 1. Unknown defect; front panel unavailable for isolation (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshooting Option 001 unit isolation difficulty
Unable to isolate problem to specific section in Option 001 power supply due to blank front panel with no controls or indicators except LINE switch, OVP ADJUST, and pilot light
Ranked causes
- 1. Blank front panel lacks diagnostic indicators (common)
Use isolation flowchart to identify problem area
- Test point
- Flowchart decision tree
- Expected
- Problem isolated to specific supply section
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ERROR 10 | Internal ROM Test failure | Self-test failure in internal ROM memory | error |
ERROR 12 | ADC Zero Too High | ADC zero calibration point exceeds acceptable range | error |
ERROR 13 | Voltage DAC Full Scale Low | Voltage DAC full scale output below acceptable range | error |
ERROR 14 | Voltage DAC Full Scale High | Voltage DAC full scale output above acceptable range | error |
ERROR 15 | Voltage DAC Zero Low | Voltage DAC zero point output below acceptable range | error |
ERROR 16 | Voltage DAC Zero High | Voltage DAC zero point output above acceptable range | error |
ERROR 17 | Current DAC Full Scale Low | Current DAC full scale output below acceptable range | error |
ERROR 18 | Current DAC Full Scale High | Current DAC full scale output above acceptable range | error |
ERROR 19 | Current DAC Zero Low | Current DAC zero point output below acceptable range | error |
ERROR 20 | Current DAC Zero High | Current DAC zero point output above acceptable range | error |
ERROR 4 | External RAM Test failure | Self-test failure in external RAM memory | error |
ERROR 5 | Internal RAM Test failure | Self-test failure in internal RAM memory | error |
ERROR 6 | External ROM Test failure | Self-test failure in external ROM memory | error |
ERROR 7 | GPIB Test failure | Self-test failure in GPIB interface | error |
ERROR 8 | GPIB address set to 31 | GPIB address configuration error | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.