Agilent 6023A: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Agilent 6023A, with 32 ranked causes, 67 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Front-panel display error or control failure
Display digits missing, display settings error, volts or amps reading incorrect, decimal point wrong, or potentiometer controls ineffective
Ranked causes
- 1. Display LED segment failure (common)
- 2. Display input buffer or decoder IC failure (common)
- 3. Decimal point driver failure (occasional)
- 4. Front-panel potentiometer failure (occasional)
Visually inspect display LEDs
- Test point
- A3DS1 through A3DS7
- Expected
- All segments light during power-on self-test
- Tools
- Visual inspection
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC turn-on relay not closing high voltage
Relay A1K1 fails to close 1.0 seconds after power applied; DROPOUT signal does not go high at 1.1 seconds; prevents PWM enable
Ranked causes
- 1. Bias voltage detector failure (common)
- 2. 1-second delay circuit malfunction (occasional)
- 3. Relay driver or relay A1K1 failure (occasional)
Verify 5V unreg bias supply present
- Test point
- A2J3-1
- Expected
- 5.0Vdc after 2s wait
- Tools
- DMM
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supply voltage incorrect or absent high voltage
Measured bias supply voltages (+5V, +15V, -15V, +8.9V, +2.5V) deviate from specification or are absent at test points on A2 control board
Ranked causes
- 1. +5V supply circuit failure (common)
- 2. +15V regulator or filter failure (common)
- 3. -15V regulator failure (common)
Measure +5V input voltage
- Test point
- A2J3-22
- Expected
- ≈ 20Vdc
- Tools
- DMM
8 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC-to-DC converter not switching high voltage
Power MOSFET gate signals not present or incorrect waveforms; output power not being generated from 40V input
Ranked causes
- 1. PWM-ON or PWM-OFF drive signals missing or incorrect (common)
- 2. PFET gate driver circuit failure (common)
- 3. Pulse transformer drive coupling failure (occasional)
Verify 40V supply to converter
- Test point
- A4Q2-D to A4Q4-S
- Expected
- 39Vdc, ±1 V
- Tools
- DMM
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current regulation failure high voltage
Output current does not reach specified maximum; constant current control not functioning properly; may oscillate with inductive loads
Ranked causes
- 1. CC monitor amplifier failure (common)
- 2. CC Error Amplifier or feedback network fault (common)
- 3. CC loop differentiator circuit instability (occasional)
Measure I-MON at 3.0A input current
- Test point
- A2U2B-7
- Expected
- 0.50Vdc, ±0.05 V
- Tools
- DMM
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage oscillation in CV regulation high voltage
Output voltage oscillates when under constant voltage control; system unstable particularly with capacitive loads
Ranked causes
- 1. CV Error Amplifier circuit oscillation (common)
- 2. CV loop compensation network failure (occasional)
- 3. Capacitive load interaction with CV circuit compensation (occasional)
Measure CV monitor output voltage
- Test point
- A2U10C-8 (V-MON)
- Expected
- 3.75Vdc, ±0.25 V
- Tools
- DMM
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PWM or clock circuit malfunction high voltage
Pulse width modulator does not generate correct output pulses or 20KHz clock not functioning; prevents power stage operation
Ranked causes
- 1. PWM inhibit gate inputs stuck low (common)
- 2. Clock generator failure (common)
- 3. Power limit circuit preventing PWM (occasional)
Check inhibit gate input 1
- Test point
- A2U19-1
- Expected
- Hi (≥ 5Vdc TTL)
- Tools
- DMM
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails turn-on self-test or shows malfunction indication high voltage
Power supply gives malfunction indication, fails self-test, or does not operate as expected
Ranked causes
- 1. Input power unavailable or circuit breaker tripped (common)
- 2. Incorrect mode switch settings (common)
- 3. Loose or interrupted connections to power supply (common)
Verify input power availability
- Test point
- Power cord inlet and rear-panel circuit breaker
- Expected
- AC power present at power cord inlet; circuit breaker in ON position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output from power supply high voltage
Unit produces no output voltage or current; load receives no power
Ranked causes
- 1. A4Q3 and A4Q4 PFET failure or fuses A4F1/A4F2 open (common)
- 2. Absence of 320Vdc rail voltage (common)
Verify bias voltages and PWM control signals present
- Test point
- A2J3 control board test connector fingers
- Expected
- PWM-ON and PWM-OFF per Table 3-1; bias voltages stable
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance failure: output present but out of specification high voltage
Power supply produces output but measurements do not meet specifications; voltage or current regulation poor
Ranked causes
- 1. Front-panel board A3 defective (common)
- 2. Functional circuit defect in control or converter stages (occasional)
Verify measurements at A2J7 control board test connector
- Test point
- A2J7 test connector
- Expected
- All signals and voltages per Table 3-1 (±5% unless range given)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or missing output voltage at test points high voltage
Voltage reading at control board test connector A2J7 does not match expected value
Ranked causes
- 1. Regulator or reference circuit fault (common)
Measure +5V at test connector A2J7 pin 1
- Test point
- A2J7 pin 1
- Expected
- 5.0 V, +/-0.25 V
- Tools
- DMM
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current oscillation
Output current oscillates during normal operation
Ranked causes
- 1. Differentiator circuit at fault (common)
Verify oscillations stop after R16 removal
- Test point
- Output
- Expected
- Oscillations stop
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status signal incorrect at control board test points
Status signal (CV, CC, OV, DROPOUT, OT) indicates wrong operational state at test connector
Ranked causes
- 1. Comparator or status logic circuit failure (occasional)
Check CV (constant voltage) status at pin 17
- Test point
- A2J7 pin 17
- Expected
- TTL Lo if in CV operation
- Tools
- DMM or logic probe
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.