Agilent 6010A: troubleshooting and error codes
The ServiceForge knowledge base holds 35 documented symptoms for the Agilent 6010A, with 54 ranked causes, 83 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power supply malfunction on startup high voltage
Power supply fails turn-on self-test or gives indication of malfunction
Ranked causes
- 1. Input power not available or power cord/circuit breaker issue (common)
- 2. Incorrect mode switch settings (common)
- 3. Loose connections or interrupted circuits (occasional)
- 4. Internal hardware defect (rare)
Verify input power available
- Test point
- AC power inlet; power cord connection
- Expected
- Input AC power available; power cord connected securely to AC outlet
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails self-test or indicates malfunction
Power supply fails turn-on self-test or shows other malfunction indication
Ranked causes
- 1. Input power not available or incorrect (common)
- 2. Mode switch A2S1 incorrectly configured (common)
- 3. Loose connections or interrupted circuits (common)
- 4. Ground terminal not properly connected to earth ground (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV regulation poor or transient response slow high voltage
Output voltage regulation is inadequate or response to load changes is sluggish
Ranked causes
- 1. Low AC mains voltage (common)
- 2. CV Circuit malfunction (common)
- 3. Incorrect sensing configuration (occasional)
Measure AC mains input voltage
- Test point
- AC input
- Expected
- Within specification for unit
- Tools
- Voltmeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maximum voltage less than 200Vdc high voltage
Output voltage cannot reach specified 200Vdc maximum
Ranked causes
- 1. CV Circuit fault (common)
- 2. Output rectifier diodes defective (occasional)
- 3. Mains voltage select jumper incorrect (occasional)
Set output voltage to maximum and measure + OUT
- Test point
- + OUT
- Expected
- 200Vdc nominal
- Tools
- DMM, variable external supply
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit produces no output high voltage
Power supply fails to generate output voltage at rear terminals
Ranked causes
- 1. PFET driver control signal absent or faulty (PWM-ON and PWM-OFF both low) (common)
- 2. PFET failed (PWM signals present but no output) (common)
- 3. Rectifier or input rail voltage missing (occasional)
Verify bias voltages at control board test connector
- Test point
- A2J7
- Expected
- Bias voltages per Table 3-1
- Tools
- Voltmeter, control board test connector P/N 5060-2865
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC regulation poor high voltage
Output current regulation inadequate or limit does not function properly
Ranked causes
- 1. Low AC mains voltage (common)
- 2. CC circuit component failure (common)
Measure AC mains voltage
- Test point
- AC input
- Expected
- Within specification
- Tools
- Voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current oscillation in CC mode
Constant current output exhibits unwanted oscillations or instability; regulated current rippling or unstable
Ranked causes
- 1. Differentiator circuit A2U4A fault (common)
- 2. CC Error Amp or summing junction fault (occasional)
Measure I-MON output
- Test point
- A2U1-out (I-MON), referenced to A2J7-4 (GND)
- Expected
- I-MON in volts ≈ 1/3 x (output current in amperes) or 1/24 for 6011A V, Within expected ratio
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output noise greater than 1 kHz high voltage
High-frequency noise present on output exceeding 1 kHz
Ranked causes
- 1. Transformer or filter defect (common)
- 2. Snubber capacitors or resistors defective (occasional)
Measure output noise with oscilloscope
- Test point
- + OUT, – OUT
- Expected
- Minimal noise > 1 kHz
- Tools
- Oscilloscope
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage oscillation in CV mode
Constant voltage output exhibits unwanted oscillations or instability; regulated voltage unstable or rippling
Ranked causes
- 1. CV Error Amp circuit fault (common)
- 2. CV loop compensation or feedback circuit fault (occasional)
Measure CV monitor output V-MON
- Test point
- A2U2-out (V-MON), measured against A2J7-4 (GND)
- Expected
- V-MON = 1/40 x (voltage between +S and -S) in volts, Within expected ratio
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply output voltage not to specification caution
Supply produces output but measurements do not meet specified values or ranges
Ranked causes
- 1. Front panel control potentiometer or switch failure (common)
- 2. DC-to-DC converter circuit malfunction (occasional)
Verify front panel voltage outputs
- Test point
- A3 test connector per Table 3-3
- Expected
- CV VOLTAGE 0-5V, CC VOLTAGE 0-5V
- Tools
- Test connector, voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unexplained OVP shutdowns high voltage
Output shutdowns triggered by overvoltage protection circuit without clear cause
Ranked causes
- 1. OVP Circuit malfunction (common)
- 2. CV Circuit fault (common)
Measure voltage at test point A2J7-2
- Test point
- A2J7-2
- Expected
- 1.25Vdc
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit cycles on and off randomly high voltage
Power supply cycles between on and off states without operator command
Ranked causes
- 1. AC surge and dropout detector malfunction (common)
- 2. Mains voltage select switch incorrect (occasional)
Verify stable AC mains input voltage
- Test point
- AC input to isolation transformer
- Expected
- Stable mains voltage
- Tools
- Voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant current circuit
Calibrate constant current circuit on Control Board A2 after component replacement
Ranked causes
- 1. Control Board A2 CC circuit requires I-MON zero and full scale calibration after repair (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant voltage circuit
Calibrate constant voltage circuit on Control Board A2 except current source components
Ranked causes
- 1. Control Board A2 CV circuit (non-current-source) requires zero and common-mode calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Control board test points voltage verification high voltage
Verify voltage levels at control board test connector A2J7 test points for proper circuit operation
Ranked causes
- 1. Reference or bias voltage out of specification (unknown)
Measure +5V bias voltage at pin 24
- Test point
- A2J7 pin 24
- Expected
- 5.0 Vdc
- Tools
- Voltmeter
18 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Diode board power limit
Calibrate power limit after replacement of CR4 or CR5/CR1 on Diode Board A5
Ranked causes
- 1. Diode Board A5 requires power limit calibration after CR4 or CR5/CR1 replacement (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: I-MON zero and full scale
Calibrate current monitoring zero and full scale on Main Board A1 after component replacement
Ranked causes
- 1. Main Board A1 requires I-MON calibration after repair of R11/R13 (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Meter full scale
Calibrate meter full scale on Control Board A2 after repair of U7, R84, R85, R24
Ranked causes
- 1. Control Board A2 U7, R84, R85, R24 requires meter full scale calibration after repair (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: OVP circuit overvoltage protection threshold
Verify and calibrate overvoltage protection (OVP) circuit response to ensure unit limits output above set threshold
Ranked causes
- 1. OVP circuit requires calibration or verification (unknown)
Measure OVP Bias Supply A2U7-2
- Test point
- A2U7-2, referenced to A2J7-4 (GND)
- Expected
- 2.5Vdc
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power limit comparator
Calibrate power limit comparator on Control Board A2 after component replacement
Ranked causes
- 1. Control Board A2 Power Limit Comparator requires calibration after repair (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Resistance programming full scale
Calibrate resistance programming full scale for CV current source on Control Board A2
Ranked causes
- 1. Control Board A2 CV current source requires resistance programming full scale calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Voltage monitoring zero
Verify zero calibration of voltage monitoring circuit on Main Board (A1)
Ranked causes
- 1. Main Board A1 requires V-MON zero calibration after repair of R11/R13 (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC circuit functional verification high voltage
Test constant current loop response with external load configured in constant current mode; verify differentiator and error amp functions
Ranked causes
- 1. CC control circuit requires verification (unknown)
Measure I-MON with IP=5V (6015A model)
- Test point
- A2U4-8, referenced to A2J7-4
- Expected
- 0.125Vdc (6015A) or 0.88Vdc for other models
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC oscillates with inductive loads high voltage
Output current or voltage oscillates when driving inductive loads
Ranked causes
- 1. CC feedback compensation defect (common)
Connect inductive load and observe output for oscillations
- Test point
- + OUT, – OUT
- Expected
- Stable output without oscillation
- Tools
- Oscilloscope, inductive load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV circuit functional verification high voltage
Test constant voltage control loop response and buffer amplifier gain with CV loop open and power mesh shutdown
Ranked causes
- 1. CV control circuit requires verification (unknown)
Measure VM (monitor output)
- Test point
- A2U5-1, referenced to A2J7-4
- Expected
- 3.75Vdc
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV load effect test
Verify constant voltage load regulation by measuring output voltage change from open-circuit to full load
Ranked causes
- 1. Output voltage drift under load indicates CV regulation error (unknown)
Set output voltage to specified value at full current
- Test point
- + S and - S terminals
- Expected
- 60Vdc (6010A), 7.0Vdc (6011A), 20.0Vdc (6012B), 200Vdc (6015A)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV load transient recovery time test
Verify output voltage recovery time following step load change between 10% and 100% of rated current
Ranked causes
- 1. Slow output voltage recovery on load transient indicates regulation circuit issue (unknown)
Set output voltage to specified value
- Test point
- + S and - S terminals
- Expected
- 60Vdc (6010A), 7Vdc (6011A), 20Vdc (6012B), 200Vdc (6015A)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV oscillates with capacitive loads high voltage
Output voltage oscillates or becomes unstable when driving capacitive loads
Ranked causes
- 1. CV feedback compensation defect (common)
Connect capacitive load and observe output for oscillations
- Test point
- + OUT, – OUT
- Expected
- Stable output without oscillation
- Tools
- Oscilloscope, capacitive load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV overshoots beyond setpoint high voltage
Output voltage exceeds setpoint and does not stabilize
Ranked causes
- 1. CV control loop component failure (common)
Set output to 20Vdc and observe step response
- Test point
- + OUT
- Expected
- Voltage settles to 20Vdc without overshoot
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV PARD peak measurement test
Verify constant voltage ripple and noise peak-to-peak from 20kHz to 20MHz across full load range
Ranked causes
- 1. Output peak-to-peak noise exceeds specification indicates ripple/noise problem (unknown)
Connect oscilloscope with blocking capacitors and differential measurement setup
- Test point
- + OUT and - OUT terminals
- Expected
- Proper differential two-channel setup per Figure 2-5
- Tools
- Oscilloscope, 0.01µF capacitors, 50Ω coaxial cables
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV PARD RMS measurement test
Verify constant voltage ripple and noise (PARD) in RMS voltage from 20Hz to 20MHz across full load range
Ranked causes
- 1. Output RMS noise exceeds specification indicates ripple/noise problem (unknown)
Connect test equipment with common-mode choke per Figure 2-4
- Test point
- + S and - S terminals
- Expected
- Proper ground connection through common-mode choke
- Tools
- 50Ω coaxial cable, common-mode choke, true RMS voltmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV source effect (line regulation) test high voltage
Verify constant voltage line regulation by measuring output voltage change from minimum to maximum input AC
Ranked causes
- 1. Output voltage drift with input AC variation indicates CV line regulation error (unknown)
Connect unit through variable autotransformer at low line voltage
- Test point
- AC power input
- Expected
- 104Vac for 120Vac nominal
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel display or control inoperative caution
Front panel switches, potentiometers, or display not responding or showing incorrect values
Ranked causes
- 1. Front panel component failure or solder joint cold (common)
Measure +7.5V bias supply on front panel
- Test point
- A3 test connector pin 1
- Expected
- 7.5V, ±5%
- Tools
- Voltmeter, test connector
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No current limit high voltage
Output current does not limit to specified maximum
Ranked causes
- 1. CC Circuit failure (common)
Measure output current at maximum setting
- Test point
- + OUT, – OUT
- Expected
- ≥ 17Adc typical
- Tools
- DMM, external load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overall troubleshooting setup and power supply verification high voltage
Verify basic power supply outputs and circuit node voltages during bench diagnostics with main troubleshooting setup configuration
Ranked causes
- 1. Power supply outputs out of specification (unknown)
Measure +15V supply on A5C3 with power ON
- Test point
- A5C3 (+) node
- Expected
- 10Vdc
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI