Agfa DRYSTAR 5500: troubleshooting and error codes
The ServiceForge knowledge base holds 70 documented symptoms for the Agfa DRYSTAR 5500, with 162 ranked causes, 86 diagnostic checks with test points and expected values, and 137 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Horizontal Film Alignment (HAL) caution
Replacement and verification of front door assembly, feed axles, drive assembly, film sensors and related components for horizontal film alignment on DRYSTAR 5500/5503 imaging systems.
Ranked causes
- 1. Front door assembly wear or misalignment (5500) (unknown)
- 2. Front door assembly wear or misalignment (5503) (unknown)
- 3. Fixed feeding axle worn or damaged (4 wheels) (unknown)
- 4. Drive assembly worn or failed (VAL) (unknown)
- 5. Light barrier failed or misaligned (unknown)
- 6. Film transport belt worn or damaged (unknown)
- 7. Stepper motor failed or damaged (unknown)
- 8. Fixed feeding axle in horizontal alignment worn or damaged (unknown)
- 9. Film sensor (PU exit detect) worn or failed (unknown)
- 10. Film sensor (HAL detect) worn or failed (unknown)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MDM measurement failed
MDM measurement failed due to corrupt or unsuitable measurement input data
Ranked causes
- 1. Shock disturbed the measurement during MDM scan (common)
- 2. Image quality not OK due to faulty Thermal Head installation files (occasional)
- 3. Image quality not OK due to incorrect Drum position (occasional)
- 4. Thermal Head needs cleaning (occasional)
- 5. Thermal Head damaged or defective (occasional)
- 6. Film position incorrect during film scan (occasional)
- 7. Marker lines not detected during film scan (occasional)
- 8. MDM scan head not moving smoothly (occasional)
- 9. DSH board defective or malfunctioning (occasional)
- 10. Incorrect MDM calibration (occasional)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Vertical film alignment (VAL) caution
Adjustment of lift carrier assembly position and springs after replacement or maintenance to ensure correct vertical film positioning and print quality
Ranked causes
- 1. Lift carrier assembly misalignment after replacement (common)
- 2. Lift carrier assembly misalignment after replacement (DRYSTAR 5503) (common)
- 3. Spring replacement required (occasional)
- 4. Stepping motor replacement (DRYSTAR 5500) (occasional)
- 5. Stepping motor replacement (DRYSTAR 5503) (occasional)
- 6. Light barrier replacement required (occasional)
Print internal test image after lift carrier assembly replacement
- Test point
- SMPTE_lin.tif from operator menu 4-1-1
- Expected
- Proper vertical alignment and image quality
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: TH profile
Reduce vertical macro banding and density irregularities; measure film sensitometry for linear density reproduction
Ranked causes
- 1. Vertical macro banding visible (occasional)
- 2. Density reproduction check failed (occasional)
- 3. After MDM replacement (occasional)
- 4. After densitometer lamp exchange (occasional)
- 5. Line artifacts persist after cleaning (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drum module not moving
No signs of movement while monitoring the drum encoder and S8 during activation of the drum motor
Ranked causes
- 1. Connection problem (common)
- 2. Drum couldn't do a full turn within provided time (common)
- 3. Drum motor M1 malfunctioning (occasional)
- 4. Encoder S32 malfunctioning (occasional)
- 5. Drum motor or encoder failure on MEC board (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad status indicator LED interpretation
Status LED shows operational state: green constant (ready), green blinking (startup/receiving/calculating/printing/calibrating), red blinking (service mode/tray empty/film jam), red constant (error).
Ranked causes
- 1. Green constant LED (common)
- 2. Green blinking LED (common)
- 3. Red blinking LED with upper/lower tray indicator (common)
- 4. Red blinking LED with film jam indicator (occasional)
- 5. Red constant LED with error status (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print drum motor M1 not connected or malfunctioning
HC10903 error code; warning message 'Call service' displayed
Ranked causes
- 1. Flat drum belt problem (common)
- 2. Toothed drum-motor belt problem (common)
- 3. Connection problem (occasional)
- 4. Defect motor M1 (occasional)
- 5. Motor functionality failure on MEC board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lower pickup motor M5 not connected or malfunctioning
HC10917 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M5 (occasional)
- 3. Defect pickup mechanism (occasional)
- 4. Sensor functionality failure on MEC board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal printhead heatsink temperature too high
Heatsink-temperature measurement of thermal head too high. Engine is still allowed to print but image quality may decrease
Ranked causes
- 1. Cooling of thermal head not efficient due to blocked airflow (common)
- 2. Malfunctioning thermal printhead cooling fan (occasional)
- 3. One or more thermistors on thermal printhead defective (occasional)
- 4. Heatsink-temperature measurement by MEC board not correct (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Upper pickup motor M4 not connected or malfunctioning
HC10916 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M4 (occasional)
- 3. Defect pickup mechanism (occasional)
- 4. Sensor functionality failure on MEC board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Print head replacement adjustments
Series of calibrations and checks required after print head replacement to ensure image quality and proper alignment
Ranked causes
- 1. Print head replacement requires film calibration from all trays (common)
- 2. Image artifacts may appear after print head replacement requiring further calibration (occasional)
- 3. Uneven density at image edges indicates aged print drum (occasional)
Perform film calibration from all trays
- Test point
- Service menu 8-3-1
- Expected
- Calibration successful for upper and all other trays
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drum calibration out of specification
Drum encoder detected more than allowed deviation from the standard pulse count per revolution
Ranked causes
- 1. Belt slip due to worn belts (common)
- 2. Belt slip due to insufficient belt tensioning (common)
- 3. Encoder read-out failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drum positioning motor M3 not connected or malfunctioning
HC10910 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M3 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exit roller motor M10 not connected or malfunctioning
HC10908 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M10 (occasional)
- 3. Sensor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film magazine locking solenoid EM1 not connected or malfunctioning
HC10918 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect solenoid EM1 (occasional)
- 3. Sensor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firewire cable not connected or malfunctioning
Firewire cable not connected or malfunctioning
Ranked causes
- 1. Connection problem (common)
- 2. Firewire functionality failure on CPU PCB crius or Tyndall CPU (occasional)
- 3. Firewire functionality failure on Themis/Tyndall board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Horizontal alignment motor M7 not connected or malfunctioning
HC10902 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M7 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light intensity too low
Light intensity as measured by the light detector on the DSH board is too low
Ranked causes
- 1. Halogen lamp not producing enough light intensity, possibly due to aging (common)
- 2. Light detector circuitry on DSH board not functioning correctly (occasional)
- 3. Light detector not receiving all available light due to masking or soiling (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MDM not leaving home position
MDM has not been detected leaving the home position within the time-out period
Ranked causes
- 1. MDM scan head is physically jammed (common)
- 2. MDM scan head transport mechanism is broken (occasional)
- 3. S7 defect (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MDM not returning to home position
MDM has not been detected returning to home position after performing a scan
Ranked causes
- 1. MDM scan head is physically jammed (common)
- 2. MDM scan head transport mechanism is broken (occasional)
- 3. S7 defect (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pickup take over motor M6 not connected or malfunctioning
HC10906 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M6 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan motor M11 of MDM not connected or malfunctioning
HC10905 error code; warning message 'Calibration module not working' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M11 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sorter position motor M12 not connected or malfunctioning
HC10909 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M12 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sorter transport motor M2 not connected or malfunctioning
HC10907 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect motor M2 (occasional)
- 3. Motor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal printhead cooling fan F1 not connected or malfunctioning
HC10913 error code; warning message 'Printhead problem' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Defect fan F1 (occasional)
- 3. Sensor functionality failure on MEC board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal printhead heatsink temperature too low
Heatsink-temperature measurement of thermal printhead too low. Engine is still allowed to print but image quality may decrease
Ranked causes
- 1. Thermal printhead temperature too low due to low ambient temperature (common)
- 2. One or more thermistors on thermal printhead defective (occasional)
- 3. Heatsink-temperature measurement by MEC board not correct (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal printhead thermistors connector not connected or thermistors defective
HC10915 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Flat cable defective (occasional)
- 3. Thermal printhead defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifacts visible in diagnostic images
Artifacts appear on printed diagnostic images from host system when checked at lightbox during maintenance
Ranked causes
- 1. Host system malfunction or configuration issue (unknown)
- 2. Printer or imaging subsystem defect (unknown)
Print diagnostic image from each connected host system
- Test point
- Printer output
- Expected
- Image without artifacts
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film density measurement
Perform film calibration to establish correct Dmax (maximum optical density) for each film tray using the internal macro densitometer (MDM).
Ranked causes
- 1. Initial film calibration required for customer films or replacement films (common)
- 2. Repeat calibration for additional film trays (common)
Verify MDM measurement area is artifact-free
- Test point
- MDM measurement area at lightbox
- Expected
- No dust stripes or artifacts visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film sensitometry
Measure film sensitometry values for internal reference to ensure linear density reproduction behavior and adapt maximum density
Ranked causes
- 1. New film package in use (occasional)
- 2. Maximum density drift detected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error or unexpected fault condition
Online diagnosis system detects error or fault; message displayed with classification as accident, warning, or defect requiring user, operator, or service action.
Ranked causes
- 1. Component disconnected or malfunctioning (motor, sensor, switch) (common)
- 2. Online diagnosis cannot determine root cause (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal head not reaching home position
Thermal head is not reaching home position
Ranked causes
- 1. Obstruction at drum positioning unit (common)
- 2. Defect TH positioning motor M3 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal head not reaching print position
Thermal head is not reaching print position
Ranked causes
- 1. Obstruction at drum positioning unit (common)
- 2. Defect TH positioning motor M3 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal head power voltage not OK high voltage
Power voltage delivered to THIB for the Thermal Head is not OK
Ranked causes
- 1. Power supply module defective (occasional)
- 2. THIB voltage measurement functionality malfunctioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal head voltage actuator not connected or malfunctioning
HC10911 error code; warning message 'Call Service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Failure on MEC board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal printhead data or power connector not connected or printhead resistors defective
HC10914 error code; warning message 'Call service' displayed
Ranked causes
- 1. Connection problem (common)
- 2. Flat cable defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
White lines on printed image
Line artifacts visible on thermal head output
Ranked causes
- 1. Thermal head resistor drift or out-of-range values (common)
- 2. Thermal head resistor out-of-range in fine measurement area (common)
Rough resistor measurement of all resistor values
- Test point
- Thermal head resistors
- Expected
- All resistor values within normal range
- Tools
- Service Menu 7 Test & Diagnose – 6 Diagnose mechanics – 5 Print Head Unit – 1 Resistor Check
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bearing replacement verification
Confirm bearing assembly is functioning after replacement
Ranked causes
- 1. Bearing assembly replaced (common)
Print internal test image after bearing replacement
- Test point
- Printer output
- Expected
- SMPTE_lin.tif image prints correctly without mechanical issues
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Density reproduction
Optical density printed output differs from target densities; verification of MDM calibration accuracy using external densitometer
Ranked causes
- 1. Optical density out of calibration range (common)
Perform film calibration
- Test point
- MDM system
- Expected
- Calibration completes successfully
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film
Film calibration to compensate density variations when new film package used, at maintenance, or if maximum density drifts
Ranked causes
- 1. Film density variation or drift of maximum density (common)
Print test film and measure with internal macro densitometer
- Test point
- Printer internal densitometer
- Expected
- Test film printed and measured; Dmax value displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film calibration failure recovery
If film calibration fails, verify presence of solid black border lines on film edges before attempting top/side border adjustment
Ranked causes
- 1. Missing solid black border lines on film edges (occasional)
Verify presence of two solid black border lines on film edges
- Test point
- Both left and right edges of failed calibration test film
- Expected
- Solid black lines visible on both sides of film
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film density measurement - DT2 Mammo tray
Validate Dmax measurement for DT2 Mammo film tray against densitometer reference values.
Ranked causes
- 1. DT2 Mammo film tray Dmax calibration validation (common)
Compare measured Dmax to DT2 Mammo target for X-Rite 310
- Test point
- Densitometer reading (DT2 Mammo film type)
- Expected
- 3.80 O.D., ±2%
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film density measurement - DT2B tray
Validate Dmax measurement for DT2B film tray against X-Rite or equivalent densitometer reference values.
Ranked causes
- 1. DT2B film tray Dmax calibration validation (common)
Compare measured Dmax to DT2B target for X-Rite 341
- Test point
- Densitometer reading (DT2B film type)
- Expected
- 3.10 O.D., ±2%
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film density measurement - DT2C tray
Validate Dmax measurement for DT2C film tray against densitometer reference values.
Ranked causes
- 1. DT2C film tray Dmax calibration validation (common)
Compare measured Dmax to DT2C target for X-Rite 310
- Test point
- Densitometer reading (DT2C film type)
- Expected
- 3.00 O.D., ±2%
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film registration
Reduce mechanical miss-registration of the densitometer assembly carriage during film scan
Ranked causes
- 1. Registration misalignment (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flat belt of drum motor assembly adjustment caution
Adjustment of flat belt position to ensure it runs centered on the flat belt pulley during motor operation
Ranked causes
- 1. Flat belt position off-center or misaligned after replacement (common)
Open drum compartment and insert service key
- Test point
- Drum compartment safety switch
- Expected
- Door open and service key seated
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal/vertical image alignment
Wizard-guided procedure to adjust image position on film for proper horizontal and vertical registration
Ranked causes
- 1. Image position on film requires horizontal/vertical adjustment (common)
Access horizontal/vertical alignment wizard
- Test point
- Service menu navigation
- Expected
- Alignment wizard interface opens with first step
- Tools
- Service Menu interface
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lift carrier assembly adjustment
Adjustment of lift carrier tilt to ensure even image border spacing on left and right edges of printed film
Ranked causes
- 1. Lift carrier tilt misaligned causing uneven top border spacing (common)
Print SMPTE test image on 14x17 film
- Test point
- Upper image border on printed film
- Expected
- Image prints on 14x17 inch film
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Local keypad backlight and contrast
Adjust backlight intensity and contrast settings on the local keypad display
Ranked causes
- 1. Keypad requires adjustment after Key4 board replacement (common)
Access Service Menu
- Test point
- Local keypad
- Expected
- Service Menu accessible
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MDM (densitometer reference)
Determine new reference values for the densitometer assembly via external densitometer
Ranked causes
- 1. Densitometer adaptation needed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MDM (Macro Densitometer)
Internal MDM must be adapted to external densitometer for image quality control, or density reproduction check fails, or after MDM or densitometer lamp replacement
Ranked causes
- 1. MDM requires adaptation to external densitometer or replacement (common)
Print test sheet with 20 different density areas
- Test point
- Printer output
- Expected
- Test sheet printed with distinct density areas
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Print Head Profile caution
Reduce line artifacts in print direction caused by Print Head and drum; performed after Print Head or drum exchange, at maintenance, or when troubleshooting image faults
Ranked causes
- 1. Line artifacts from Print Head or drum irregularities (common)
Perform Print Head cleaning procedure
- Test point
- Print Head compartment
- Expected
- Print Head cleaned and reinstalled
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Print head resistor check
Electrical continuity and resistance check of print head resistors; identifies line artifacts caused by resistor drift or failure
Ranked causes
- 1. Print head resistor out of range (occasional)
Perform rough resistor measurement of all resistor values
- Test point
- Print Head resistor network
- Expected
- All resistor values within acceptable drift tolerance
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Thermal head resistor check
Diagnostic procedure to verify thermal head electrical integrity and detect line artifacts caused by resistor issues
Ranked causes
- 1. Thermal head requires resistor validation before replacement decision (common)
Perform rough resistor measurement
- Test point
- All thermal head resistor positions
- Expected
- Resistor values within normal range, no large drifts
- Tools
- Service Menu 7 Test & Diagnose – 6 Diagnose mechanics – 5 Print Head Unit – 1 Resistor Check
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Check for dust stripes on print head caution
Vertical lines or dust artifacts visible on printed film; diagnostic flowchart to determine if cleaning or replacement needed
Ranked causes
- 1. Print head contamination or defect (common)
Print STI2 and medical_image test images
- Test point
- Printer output
- Expected
- No vertical stripes visible in either image
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cleaning module kit replacement verification
Confirm cleaning module is functioning after replacement
Ranked causes
- 1. Cleaning module kit replaced (common)
Print internal test image after cleaning module replacement
- Test point
- Printer output
- Expected
- SMPTE_lin.tif image prints cleanly without artifacts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drum replacement verification
Confirm drum assembly is working after replacement
Ranked causes
- 1. Drum assembly replaced (common)
Print test images after drum replacement
- Test point
- Film output
- Expected
- No print stripes, good image quality on both STI2 and medical images
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrostatic brush replacement verification
Confirm electrostatic brush is functioning after replacement
Ranked causes
- 1. Electrostatic brush replaced (common)
Print internal test image after brush replacement
- Test point
- Printer output
- Expected
- SMPTE_lin.tif image prints correctly without quality issues
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film calibration disabled for RF tag missing
Automatic film calibration does not run when RF tag is absent from film pack
Ranked causes
- 1. RF tag missing from film pack (occasional)
Verify RF tag presence on film pack
- Test point
- Film pack RF tag
- Expected
- RF tag present and readable
- Tools
- Visual inspection, RF tag reader
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam after reboot with partial printing
First film after reboot causes data underrun, leading to film jam and only partly printed film
Ranked causes
- 1. Software bug in versions prior to 1.81_C11 (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Film jam condition
- Film jam reported despite no actual jam
- Incorrect total cycles displayed after thermal head removal
- MDM light detector detecting too much stray light
- Offline diagnosis system troubleshooting required
- Print stripes visible on film
- Spring replacement verification
- Thermal print head voltage out of range
- Top border inconsistency on first film after reboot
- Very low Dmax after film calibration
10 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Film jam in output section (false) | Erroneous film jam detection in output sorter section | Sorter exit sensor debounce insufficient; film covering sensor triggers toggle event | error |
HC10000 | Crius PCB not functioning correctly | Self-test detected Crius PCB failure | error |
HC10001 | Crius board battery RAM test failed | Self-test detected Battery RAM failure | error |
HC10002 | RAM test failed on Crius CPU or Tyndall CPU | Self-test detected CPU RAM failure | error |
HC10003 | CPU temperature too high on Crius or Tyndall | Temperature sensor detected high temperature | warning |
HC10050 | CPU PCB on MEC not functioning correctly | Self-test detected MEC CPU PCB failure | error |
HC10052 | CPU RAM on MEC test failed | Self-test detected MEC CPU RAM failure | error |
HC10053 | CPU temperature too high on Themis or Tyndall MEC | Temperature sensor detected high temperature on MEC | warning |
HC10100 | MEC PCB test failed | Self-test detected MEC PCB failure | error |
HC10200 | DSH test failed | Self-test detected DSH failure | error |
HC10300 | Power supply module test failed | Self-test detected power supply module failure | error |
HC10301 | Power supply module temperature too high | Temperature sensor detected high temperature | error |
HC10302 | Power heater voltage out of range | Power heater voltage from power supply module out of range | error |
HC10303 | Power logic voltage out of range | Power logic voltage from power supply module out of range | error |
HC10304 | Power motor voltage out of range | Power motor voltage from power supply module out of range | error |
HC10305 | Thermal printhead voltage from power supply module is out of range | Power supply module voltage measurement indicates value outside acceptable limits | error |
HC10400 | Thermal head resistance value calculated from RTD measurements is out of tolerated range | All thermal head resistors measured and average resistance value exceeds specification per D:\thsystem\default.thpar | warning |
HC10500 | Errors detected during hard disk diagnostic | Hard disk diagnostic routine fails | error |
HC10600 | Floppy drive problem detected during diagnostic | Floppy drive diagnostic routine fails | error |
HC10700 | Key-pad problem detected during diagnostic | Key-pad diagnostic routine fails | error |
HC10800 | Cover open switch SW1 is not connected or malfunctioning | Cover open sensor SW1 fails diagnostic test | error |
HC10801 | Cover open switch SW2 is not connected or malfunctioning | Cover open sensor SW2 fails diagnostic test | error |
HC10802 | Middle magazine empty sensor S12 is not connected or malfunctioning | Middle tray empty detection sensor S12 fails diagnostic test | error |
HC10803 | Upper magazine empty sensor S12 is not connected or malfunctioning (2-tray printer) | Upper tray empty detection sensor S12 fails diagnostic test | error |
HC10804 | Upper magazine empty sensor S33 is not connected or malfunctioning (3-tray printer) | Upper tray empty detection sensor S33 fails diagnostic test | error |
HC10805 | Upper pickup passed sensor S19 is not connected or malfunctioning (2-tray printer) | Upper pickup passed detection sensor S19 fails diagnostic test | error |
HC10806 | Upper pickup module home position sensor S15 is not connected or malfunctioning | Home position sensor S15 fails diagnostic test | error |
HC10807 | Middle pickup passed sensor S19 is not connected or malfunctioning | Middle pickup passed detection sensor S19 fails diagnostic test | error |
HC10808 | Lower magazine empty sensor S13 is not connected or malfunctioning | Lower tray empty detection sensor S13 fails diagnostic test | error |
HC10809 | Upper pickup module pickup position sensor S16 is not connected or malfunctioning | Pickup position sensor S16 fails diagnostic test | error |
HC10810 | Lower pickup passed sensor S20 is not connected or malfunctioning | Lower pickup passed detection sensor S20 fails diagnostic test | error |
HC10811 | Middle pickup module home position sensor S15 not connected or malfunctioning | S15 sensor failure or disconnection | error |
HC10812 | Lower pickup module home position sensor S17 not connected or malfunctioning | S17 sensor failure or disconnection | error |
HC10813 | Lower pickup module pickup position sensor S18 not connected or malfunctioning | S18 sensor failure or disconnection | error |
HC10814 | Drum passed sensor S24 not connected or malfunctioning | S24 sensor failure or disconnection | error |
HC10815 | Drum entry sensor S23 not connected or malfunctioning | S23 sensor failure or disconnection | error |
HC10816 | Printdrum reference position sensor S8 not connected or malfunctioning | S8 sensor failure or disconnection | error |
HC10817 | Ambient temperature measurement out of allowed window | Temperature outside specifications or THIB malfunction | warning |
HC10818 | Middle pickup module pickup position sensor S16 not connected or malfunctioning | S16 sensor failure or disconnection | error |
HC10819 | MDM light sensor fails | MDM light sensor failure | error |
HC10820 | MDM home position sensor S7 not connected or malfunctioning | S7 sensor failure or disconnection | error |
HC10821 | Pickup module magazine lock position sensor S9 not connected or malfunctioning | S9 sensor failure or disconnection | error |
HC10822 | Pickup module pickup exit sensor S21 not connected or malfunctioning | S21 sensor failure or disconnection | error |
HC10823 | Output module sorter exit sensor S27 not connected or malfunctioning | S27 sensor failure or disconnection | error |
HC10824 | Alignment module vertical home position sensor S5 not connected or malfunctioning | S5 sensor failure or disconnection | error |
HC10825 | Alignment module horizontal home position sensor S6 not connected or malfunctioning | S6 sensor failure or disconnection | error |
HC10826 | Alignment module alignment entry sensor S22 not connected or malfunctioning | S22 sensor failure or disconnection | error |
HC10827 | Drum module home position sensor S1 not connected or malfunctioning | S1 sensor failure or disconnection | error |
HC10828 | Drum module 14 inch print format position sensor S2 not connected or malfunctioning | Sensor S2 failure or disconnection | error |
HC10829 | Thermal head power voltage delivered by power supply and measured on THIB out of range | Voltage out of range on THIB | error |
HC10830 | Drum module 10 inch print format position sensor S3 not connected or malfunctioning | Sensor S3 failure or disconnection | error |
HC10831 | Output module sorter entry sensor S25 not connected or malfunctioning | Sensor S25 failure or disconnection | error |
HC10832 | Output module first output position sensor S31 not connected or malfunctioning | Sensor S31 failure or disconnection | error |
HC10833 | Output module second output tray position sensor S30 not connected or malfunctioning | Sensor S30 failure or disconnection | error |
HC10834 | Output module third output tray position sensor S29 not connected or malfunctioning | Sensor S29 failure or disconnection | error |
HC10835 | Output module fourth output tray position sensor S28 not connected or malfunctioning | Sensor S28 failure or disconnection | error |
HC10836 | Upper pickup passed sensor S36 not connected or malfunctioning | Sensor S36 failure or disconnection | error |
HC10837 | Upper pickup module home position sensor S34 not connected or malfunctioning | Sensor S34 failure or disconnection | error |
HC10838 | Upper pickup module pickup position sensor S35 not connected or malfunctioning | Sensor S35 failure or disconnection | error |
HC10839 | Jam clearance toggle switch SW5 not connected or malfunctioning | Switch SW5 failure or disconnection | error |
HC10900 | Vacuum pump motor M13 not connected or malfunctioning | Motor M13 failure or disconnection | error |
HC10901 | Vertical alignment motor M8 not connected or malfunctioning | Motor M8 failure or disconnection | error |
HC10902 | Horizontal alignment motor M7 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10903 | Print drum motor M1 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10904 | MDM lamp not connected or malfunctioning or light sensor malfunctioning | Lamp or sensor malfunction or connection failure | error |
HC10905 | Scan motor M11 of MDM not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10906 | Pickup module take over motor M6 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10907 | Output module sorter transport motor M2 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10908 | Output module exit roller motor M10 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10909 | Output module sorter position motor M12 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10910 | Drum module positioning motor M3 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10911 | Thermal head voltage actuator not connected or malfunctioning | Power connection failure or board malfunction | error |
HC10912 | Blemish heater foil failure (false alarm) | Accumulated-power check on thermal head preheating enabled | error |
HC10912 | Thermal head heater not connected or malfunctioning | Heater malfunction or connection failure | error |
HC10913 | Thermal printhead cooling fan F1 not connected or malfunctioning | Fan malfunction or connection failure | error |
HC10914 | Thermal printhead data connector or power connector not connected or all printhead resistors defective | Connector connection failure or resistor malfunction | error |
HC10915 | Thermal printhead thermistors connector not connected or all thermistors defective | Connector connection failure or thermistor malfunction | error |
HC10916 | Upper pickup motor M4 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10917 | Lower pickup motor M5 not connected or malfunctioning | Motor malfunction or connection failure | error |
HC10918 | Film magazine locking solenoid EM1 not connected or malfunctioning | Solenoid malfunction or connection failure | error |
HC10919 | 10 inch film format align solenoid EM2 not connected or malfunctioning | EM2 solenoid failure or disconnection | error |
HC10920 | MDM module output takeover motor M9 not connected or malfunctioning | M9 motor failure or disconnection | error |
HC10921 | Upper pickup module, upper magazine vacuum valve V1 not connected or malfunctioning | V1 vacuum valve failure or disconnection | error |
HC10922 | Lower pickup module, lower magazine vacuum valve V2 not connected or malfunctioning | V2 vacuum valve failure or disconnection | error |
HC10923 | Middle pickup module, upper magazine vacuum valve V1 not connected or malfunctioning | V1 vacuum valve failure or disconnection | error |
HC10924 | Middle pickup motor M4 not connected or malfunctioning | M4 motor failure or disconnection | error |
HC10927 | Upper pickup module, upper magazine vacuum valve V3 not connected or malfunctioning | V3 vacuum valve failure or disconnection | error |
HC10928 | Upper pickup motor M14 not connected or malfunctioning | M14 motor failure or disconnection | error |
HC10930 | Bsp module film movement sensor S42 is not connected or malfunctioning | S42 sensor failure or disconnection | error |
HC10931 | Bsp module black-sensor S43 is not connected or malfunctioning | S43 sensor failure or disconnection | error |
HC11000 | Firewire cable not connected or malfunctioning | Firewire connection failure | error |
HC11001 | 24 v power for the actuators is not available on the MEC board | Power supply failure or disconnection | error |
HC11002 | Power sensor failure on MEC board | Sensor functionality failure | error |
HC11003 | the voltages 24v, 12v, 5v are not ok on MEC board | Voltage regulation failure | error |
HC11004 | The KEY5 board of the display is not connected to the MEC board | KEY5 board disconnection | error |
HC11005 | The DSH board of the MDM module is not connected to the MEC board | DSH board disconnection or malfunction | warning |
HC11006 | Connector P21 not detected on MEC | P21 connector unplugged or loose on MEC board | error |
HC11007 | Connector P27 not detected on MEC | P27 connector unplugged or loose on MEC board | error |
HC11008 | Connector P28 not detected on MEC | P28 connector unplugged or loose on MEC board | error |
HC11009 | Power supply has no serial connection to MEC | Power supply serial connection lost or failed | error |
HC11010 | Connector P29 not detected on MEC | P29 connector unplugged or loose on MEC board | error |
HC11011 | Connector P30 not detected on MEC | P30 connector unplugged or loose on MEC board | error |
HC11012 | Connector P31 not detected on MEC | P31 connector unplugged or loose on MEC board | error |
HC11013 | Connector P34 not detected on MEC | P34 connector unplugged or loose on MEC board | error |
HC11014 | Connector P35 not detected on MEC | P35 connector unplugged or loose on MEC board | error |
HC11015 | Bad connectivity between thermal printhead P1 and MEC P22 | Cable or connector connection lost or degraded between TH P1 and MEC P22 | error |
HC11016 | Bad connectivity between thermal printhead P2 and MEC P23 | Cable or connector connection lost or degraded between TH P2 and MEC P23 | error |
HC11017 | No connection detected between MEC and THIB | Flat cable connection lost between MEC P38 and THIB P2 | error |
HC11018 | Upper tray tag reader not connected or malfunctioning (2 tray printer) | Tag reader connection lost or device failed | error |
HC11019 | Lower tray tag reader not connected or malfunctioning | Tag reader connection lost or device failed | error |
HC11020 | No power connection between THIB and thermal head | Power cable connection lost between THIB and thermal head | error |
HC11021 | Middle tray tag reader not connected or malfunctioning | Tag reader connection lost or device failed | error |
HC11022 | Upper tray tag reader not connected or malfunctioning (3 tray printer) | Tag reader connection lost or device failed | error |
HC11023 | No connection detected at safety switch | Safety switch connection lost or failed | error |
HC11100 | Drum module mechanical problem - no movement detected | No signs of movement while monitoring drum encoder and S8 during drum motor activation | error |
HC11101 | Drum calibration failed - excessive pulse count deviation | Drum encoder detected more than allowed deviation from standard pulse count per revolution | error |
HC11102 | Thermal head thermistor mean value bad | Thermal head mean value for thermistors out of specification | warning |
HC11103 | Thermal head not reaching print position | Thermal head unable to reach print position during operation | error |
HC11104 | Thermal head not reaching home position | Thermal head unable to reach home position during operation | error |
HC11105 | Thermal head power voltage out of specification | Power voltage delivered to THIB for thermal head is not OK | error |
HC11106 | Thermal head fan or cooling not OK | Cooling fan or thermal head cooling system malfunction detected | warning |
HC11107 | Thermal head temperature out of range | Thermal head temperature measurement outside acceptable limits | warning |
HC11108 | Heating thermal head temperature out of range | Thermal head heating element temperature out of specification | warning |
HC11109 | Firewire cable not connected or malfunctioning | Firewire interface not responding | error |
HC11110 | Take-over drive not working | Take-over mechanism drive system failure | error |
HC11111 | MDM light detection too low | Light intensity measured by DSH board is too low | warning |
HC11112 | MDM light detection too high from stray light | Light detection sensor detecting too much stray light | warning |
HC11113 | MDM not returning to home position after scan | MDM scan head has not been detected returning to home position after performing scan | warning |
HC11114 | MDM not leaving home position within timeout | MDM scan head has not been detected leaving home position within time-out period | warning |
HC11115 | MDM measurement failed - corrupt input data | MDM measurement failed due to corrupt or unsuitable measurement input data | error |
HC11116 | Decurl module heating insufficient | Decurl module temperature below specification | warning |
HC11117 | Decurl module temperature too high | Decurl module temperature above specification | error |
HC11118 | Decurl module mechanical failure | Decurl module mechanical system malfunction | error |
HC11119 | Thermal printhead heatsink temperature too low | Heatsink-temperature measurement of thermal printhead too low | warning |
HC11120 | Thermal printhead heatsink temperature too high | Heatsink-temperature measurement of thermal head too high | warning |
HC11121 | Bitgenerator test failed | Bitgenerator self-test failure detected during initialization or operation | error |
HC11211 | Remove input jam error | Frequent occurrences with 11x14" films | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.