Agfa Drystar 4500 M: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Agfa Drystar 4500 M, with 23 ranked causes, 39 diagnostic checks with test points and expected values, and 10 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bad quality of printed images
Printing occurs but output quality is degraded; images are faint, have banding, scratches, or white dots/lines
Ranked causes
- 1. Consumables wrongly inserted (common)
- 2. Film contamination or transport obstruction causing dots/lines/scratches (common)
- 3. Low frequency banding in output (occasional)
- 4. Image quality degraded from normal operation (occasional)
Verify consumables installation
- Test point
- All consumables (film, ink, developer)
- Expected
- All consumables correctly inserted per specifications
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image quality degraded or defective
Visible defects on printed film including white dots, lines, density irregularities, or scratches
Ranked causes
- 1. Print head contaminated with dust (common)
- 2. Print head out of profile calibration (occasional)
- 3. Film sensitometry or system density incorrect (occasional)
- 4. Film sheet placed upside-down in input tray (rare)
Inspect printed film for white dots or lines in transport direction
- Test point
- Film surface
- Expected
- No visible white dots or lines
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not print
Drystar 4500 M is not printing a submitted job; may display error status or print queue issues
Ranked causes
- 1. Error condition present (common)
- 2. Power or network connection issue (common)
- 3. Print queue issue (occasional)
- 4. Floppy disk error (rare)
Check for error messages on display
- Test point
- Front panel display
- Expected
- Either Clear ready screen or specific error message displayed
- Tools
- Visual inspection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam in input tray caution
Film is stuck in the feed section of upper or lower input tray; display shows jam location indicator
Ranked causes
- 1. Film loaded upside down (common)
- 2. Improper film stack positioning (occasional)
Press Stop button to initiate stop sequence
- Test point
- Stop button
- Expected
- Printer finishes current jobs and displays 'OPEN THE INPUT TRAY' message
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifacts visible on mammography test image
Significant disturbing artifacts appear on the QC mammography test film when performing weekly quality control.
Ranked causes
- 1. Image quality degradation requiring maintenance (unknown)
Print QC mammography test image
- Test point
- Test film output
- Expected
- Image with visible test patterns without significant disturbing artifacts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drystar 4500 M does not print
Printer does not produce any output; no film advances or imaging occurs
Ranked causes
- 1. Status LED indicates error or service mode (common)
Observe status indicator LED color and blink pattern
- Test point
- Front panel LED indicator
- Expected
- Green constant (ready) or blinking (normal operation); not red
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam from protective sheet caution
Protective sheet from film pack is jammed in front transport section; occurs when film loaded upside down
Ranked causes
- 1. Protective sheet facing upward (film pack loaded upside down) (common)
Remove jammed protective sheet from front section
- Test point
- Front transport section
- Expected
- Protective sheet removed without damage
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam from unauthorized cover opening patient safety
Film jam caused by opening printer covers or input trays while film is being printed
Ranked causes
- 1. Operator opened printer covers or input trays during active printing without pressing Stop button (occasional)
Determine jam location from display indicator
- Test point
- Display screen
- Expected
- Pictogram shows jam location (upper tray, lower tray, front, or top section)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam in front transport section caution
Film is stuck in front section of film transport system; display shows jam indicator at front
Ranked causes
- 1. Unauthorized printer cover opening during printing (common)
Pull front door open
- Test point
- Front door
- Expected
- Door opens to reveal interior
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam in top transport section caution
Film is stuck in upper section of film transport system; display shows jam indicator at top
Ranked causes
- 1. Unauthorized printer cover opening during printing (common)
Press button on right side to open top cover
- Test point
- Top cover button [1]
- Expected
- Top cover [2] opens to reveal interior
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Geometric consistency variance on annual mammography QC
Measured image dimensions or aspect ratio deviate from reference values on annual geometric consistency test.
Ranked causes
- 1. Image size or distortion drift outside specification (unknown)
Measure distance A on geometric square
- Test point
- Left and right edges of geometric square on QC test image
- Expected
- Measured within 1.0% of reference value A_ref, <=1.0%
- Tools
- 30 cm machinist scale with 0.5 mm divisions
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low contrast visibility failure on mammography test
Cannot see the circle in the square at density scale extremes (high 100/95% and low 0/5%) on the QC test film.
Ranked causes
- 1. Insufficient contrast rendering or image quality degradation (unknown)
Verify visibility of circle in square at high density end (100/95%)
- Test point
- QC mammography test image high contrast area
- Expected
- Circle in square visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spatial resolution loss on mammography test
Five-dot clusters in the three ovals on the QC test film are not all visible when examined with a magnifying glass.
Ranked causes
- 1. Poor viewing conditions or image degradation (unknown)
Examine three ovals each containing 5-dot clusters with magnifying glass
- Test point
- QC mammography test image ovals
- Expected
- All five dots of each group visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Error status (red constant) | Critical error condition preventing operation | Hardware or software malfunction detected | critical |
Failed (Film calibration) | Film calibration process did not complete successfully | Film calibration procedure did not achieve required Dmax or density measurement targets | error |
Failed (Print head profile calibration) | Print head profile calibration did not complete successfully | Print head profile or registration measurement failed to meet calibration standards | error |
Film jam | Film is jammed in transport or input tray | Film fed incorrectly or obstruction in transport path | error |
Input tray empty | Film supply depleted in input tray | Input tray (upper or lower) has no film remaining | warning |
Key selftest failed | Key selftest failed or display failure in firmware | Self test performed during startup | error |
Key selftest failed due to no communication between Key and | Key and CPU communication error during startup | Self test performed during startup | error |
Service mode (red blinking) | Printer in service/maintenance mode; not available for normal printing | Service mode activated (requires technical access) | error |
Software selftest failed | Software selftest failure during startup | Self test performed during startup | error |
Warning status | Non-critical warning condition detected | Various operational warnings | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community