Advanced Sterilization Products AER: troubleshooting and error codes
The ServiceForge knowledge base holds 75 documented symptoms for the Advanced Sterilization Products AER, with 189 ranked causes, 177 diagnostic checks with test points and expected values, and 11 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No soap injection into basin
Soap does not flow into basin during wash cycle despite soap supply being available
Ranked causes
- 1. Soap bottle empty or disconnected (common)
- 2. Soap bottle hose connection loose (common)
- 3. Soap injector inlet/outlet hose kinked or loose (common)
- 4. Soap injector port disconnected (common)
- 5. Basin soap injector port clogged (occasional)
- 6. Lid sensor switch malfunction (occasional)
- 7. Soap injector air lock (occasional)
- 8. Soap injector power or electrical fault (occasional)
- 9. Soap injector harness disconnected (occasional)
Inspect soap bottle level
- Test point
- Soap bottle
- Expected
- Soap bottle filled with soap
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E06 - Insufficient fresh water flow to basin
Fresh water not flowing into basin in adequate amounts; backup timer shut off before basin float switch activated.
Ranked causes
- 1. Kinked or blocked water supply hose (common)
- 2. Clogged or blocked basin float switch or canister (common)
- 3. Water flow inadequate or hose kinked (common)
- 4. Basin level float sensor malfunction (occasional)
- 5. Tap water supply valve obstruction (occasional)
- 6. Low water supply pressure (occasional)
- 7. Water filtration system blocked or clogged (occasional)
- 8. Water inlet valve blockage or failure (occasional)
Check water supply hose for kinks or blockage
- Test point
- Water supply hose full length
- Expected
- Hose straight and clear
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No air circulation
No air flow detected during operation; compressor or air delivery not functioning
Ranked causes
- 1. Compressor malfunction (common)
- 2. Blocked endoscope channels (common)
- 3. Tubing connection failure (common)
- 4. Control panel not responding (occasional)
- 5. Compressor capacitor failure (occasional)
- 6. Relief valve adjustment incorrect (occasional)
- 7. Channel valve malfunction (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No circulation of liquid through endoscope channels
Liquid delivery to endoscope channels not functioning
Ranked causes
- 1. Tubing connection failure at adapters (common)
- 2. Quick release coupling disconnection (common)
- 3. Tubing kinked or blocked (common)
- 4. Tubing connected to wrong endoscope (occasional)
- 5. Main pump malfunction (occasional)
- 6. Main pump valve failure (occasional)
- 7. Channel valve malfunction (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cycle does not start when START is pressed
Reprocessor cycle fails to initiate when START button pressed
Ranked causes
- 1. Lid not properly closed or interlock disengaged (common)
- 2. Disinfectant reservoir level insufficient (common)
- 3. Reset switch position incorrect (occasional)
- 4. Control panel harness or connector issue (occasional)
- 5. P1/P2 or controller connector problem (occasional)
- 6. Control panel membrane defective (rare)
Verify lid closure
- Expected
- Lid fully closed and interlock engaged
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low disinfectant level alert (Code E05) patient safety
Display shows fewer than 3.5 or 4.5 gallons remaining in reservoir depending on model
Ranked causes
- 1. Waste valve not closing completely (common)
- 2. Float switch defective (top of reservoir) (occasional)
- 3. Float switch defective (bottom of reservoir) (occasional)
- 4. V1 valve obstruction or malfunction (occasional)
- 5. Main pump obstruction or failure (occasional)
- 6. V6 waste valve seal failure (occasional)
Check input light 0012 status
- Test point
- Input light 0012
- Expected
- Not lit (normal); lit indicates float switch defect
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin does not fill with water caution
Endoscope reprocessor basin fails to fill with water during cycle
Ranked causes
- 1. Water source not connected or turned off (common)
- 2. Lid sensor switch malfunction (occasional)
- 3. Basin level switch malfunction (occasional)
- 4. Tap water valve malfunction or blocked (occasional)
- 5. Water filtration system clogged (occasional)
Verify water source is on and supply line connected
- Test point
- water inlet connection and source valve
- Expected
- water on, hose connected and not kinked
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E01 displayed
High level warning switch activated; disinfectant reservoir may be too full or disinfectant mixing with water from tap water valve or reservoir valve
Ranked causes
- 1. Disinfectant reservoir too full (unknown)
- 2. Disinfectant diluted with tap water (unknown)
- 3. High level float switch stuck (unknown)
- 4. V4 tap water valve defective (unknown)
- 5. V7 fill tank/disinfectant reservoir valve defective (unknown)
Customer test strips failing
- Test point
- Test strips
- Expected
- Pass
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E03 disinfectant reservoir empty
Disinfectant reservoir is empty; only FILL DISINFECTANT cycle function available
Ranked causes
- 1. Waste valve stuck open (common)
- 2. Obstruction in basket strainer or drain (common)
- 3. Main pump or main pump valve failure (occasional)
- 4. Disinfectant valve V7 closed or defective (occasional)
- 5. Float switch defective or clogged (occasional)
Verify E03 error code display
- Test point
- Reprocessor display
- Expected
- E03 code shown
5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant Not Draining From Basin
Disinfectant fluid remains in basin at end of cycle or manually
Ranked causes
- 1. Basin strainer or drain blocked with debris (common)
- 2. Lid sensor switch faulty (occasional)
- 3. Waste valve faulty (occasional)
- 4. Main pump or pump valve obstructed (occasional)
- 5. Drain line kinked (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Machine stops mid-cycle
Endoscope reprocessor stops unexpectedly during a processing cycle
Ranked causes
- 1. Power supply failure or missing mains power (common)
- 2. Main power fuse is blown (common)
- 3. Lid not properly closed (occasional)
- 4. START button not illuminated or STOP button illuminated (occasional)
- 5. Control system requires reset (occasional)
Verify 115 VAC mains power at equipment
- Test point
- Power outlet and equipment input
- Expected
- 115 VAC present
- Tools
- Multimeter
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual air function not working
The manual air dry function on the endoscope reprocessor is not operating
Ranked causes
- 1. Lid sensor switch fault (common)
- 2. Air compressor or air valve not functioning (common)
- 3. Tubing set connection loose at quick release couplings (occasional)
- 4. Tubing set kinked or pinched (occasional)
- 5. Blockage in tubing or moisture trap filter (occasional)
Test lid sensor switch operation
- Test point
- Lid sensor switch contact
- Expected
- Switch properly detecting lid open/closed
- Tools
- Visual inspection, multimeter
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No spray action in basin caution
Spray tower not dispensing liquid or not spinning during operation
Ranked causes
- 1. Spray tower blockage (common)
- 2. Basin drain or strainer blockage (common)
- 3. Spray tower not spinning (occasional)
- 4. Main pump assembly failure (occasional)
- 5. Spray tower valve failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot set disinfect time patient safety
System does not allow disinfect time to be entered or set
Ranked causes
- 1. Disinfect time not entered before cycle start (common)
- 2. Disinfectant reservoir empty (common)
- 3. Defective membrane switch or loose connector (rare)
- 4. Defective input light or control panel circuit (rare)
Verify DISINFECT TIME button accessible
- Test point
- Control panel DISINFECT TIME button
- Expected
- Button responsive and input light 0004 illuminated
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channels not air purged
Air purge cycle does not deliver air to endoscope channels
Ranked causes
- 1. Tubing connection or routing problem (common)
- 2. Air compressor not running (common)
- 3. Relief valve pressure or air valve issue (occasional)
- 4. Controller/microprocessor output problem (occasional)
Run manual air cycle with tubing connected
- Expected
- Air flows to channels
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant Heater Light Not Lit
Heater indicator light is not illuminated on control panel
Ranked causes
- 1. Heater disconnected or turned off (common)
- 2. Power off or unit not functioning (occasional)
- 3. Temperature sensor connection faulty (occasional)
- 4. Disinfectant reservoir not full (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant Not Entering Basin
Disinfectant not flowing into basin during cycle
Ranked causes
- 1. Reservoir empty (common)
- 2. Disinfectant pump or pump valve not working (occasional)
- 3. Waste valve stuck open (occasional)
- 4. Lid sensor switch faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display microprocessor not illuminating
Microprocessor indicator lights do not light on control panel
Ranked causes
- 1. Display harness connection loose or damaged (common)
- 2. Control panel connector P1 or P2 loose or damaged (common)
- 3. Hex display module defective (occasional)
- 4. Control panel assembly defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error E01: High disinfectant level warning caution
Disinfectant reservoir filled beyond acceptable level or disinfectant diluted with water; high-level float switch activated
Ranked causes
- 1. Disinfectant reservoir too full (common)
- 2. Disinfectant diluted with tap water from V4 valve leak (occasional)
- 3. Disinfectant diluted with tap water from V7 fill/reservoir valve leak (occasional)
- 4. High-level float switch stuck or defective (occasional)
Verify E01 error reappears
- Test point
- Error display
- Expected
- E01 appears again after cycle
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display on control panel
Control panel display is blank or non-responsive
Ranked causes
- 1. No power to unit (common)
- 2. Main power switch off (common)
- 3. Fuse F1 blown (occasional)
- 4. Display harness disconnected (occasional)
Verify power is connected to unit
- Expected
- 115 VAC available at outlet; power cord secured
- Tools
- Voltmeter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No printer output
Printer does not produce paper output when cycle completes
Ranked causes
- 1. Paper depleted in printer (common)
- 2. Printer ribbon depleted (common)
- 3. Printer not connected (common)
- 4. Software fault (occasional)
Press paper advance button
- Test point
- Printer paper advance mechanism
- Expected
- Paper advances through printer
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump or solenoid valve not operating
Component does not activate during cycle; indicator light on controller/microprocessor may or may not be illuminated
Ranked causes
- 1. Solenoid or pump component is faulty (occasional)
- 2. Problem in harness, power distribution board, or microprocessor board connections (occasional)
- 3. Blown fuse in pump circuit (occasional)
- 4. Sensor warning signal preventing component activation (occasional)
Verify indicator light on controller/microprocessor is illuminated for the component
- Test point
- Controller/microprocessor display
- Expected
- Indicator light illuminated during cycle using that component
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Box operation creates electrical shock or disinfectant contact hazard high voltage
Liquid transferred to inappropriate places due to Service Box overriding safety switches during troubleshooting
Ranked causes
- 1. Service Box switches left in ON position before connection or after use (common)
- 2. Lid open during Service Box operation allowing spray tower activation (common)
- 3. Basin overflow from bypassed float switch (common)
- 4. Pump run dry from empty reservoir (occasional)
Verify all Service Box switches are in OFF position
- Test point
- Service Box switches
- Expected
- All switches in down (OFF) position
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waste water not draining from basin caution
Basin water not draining or draining improperly
Ranked causes
- 1. Waste hose kinked or blocked (common)
- 2. Basin strainer or drain blockage (common)
- 3. Waste valve failure (occasional)
- 4. Lid sensor switch malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin float switch not operating correctly
Float unable to move freely in canister, rubbing on inside, or installed incorrectly causing water flow blockage or false level indication
Ranked causes
- 1. Float stuck or rubbing on canister interior (common)
- 2. Float clogged or obstructed (occasional)
- 3. Float installed upside-down (occasional)
Visually inspect float movement in canister
- Test point
- Float in canister, left rear corner of basin
- Expected
- Float moves up and down freely without rubbing
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E04 displayed
E04 error code shown; disinfectant reservoir contains fluid preventing FILL DISINFECTANT cycle
Ranked causes
- 1. Liquid remains in disinfectant reservoir (common)
- 2. Disinfectant reservoir empty float stuck (occasional)
- 3. Float switch defective (rare)
Check input light status on C40K board
- Test point
- C40K board indicator lights
- Expected
- 0011 lit; 0012 or 0014 not lit
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E08 displayed
E08 error appears on 387P units with printer; system indicates wash time not entered
Ranked causes
- 1. Wash time not entered into system (common)
- 2. Disinfectant reservoir empty (common)
- 3. Wash time input control malfunction (occasional)
Verify WASH TIME has been entered
- Test point
- system display or input field
- Expected
- WASH TIME value is set
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Machine does not stop when the lid is lifted
Endoscope reprocessor continues operation when lid is opened instead of stopping immediately
Ranked causes
- 1. Lid switch light not on when lid is open (common)
- 2. Magnet on the lid is missing or displaced (occasional)
- 3. Lid switch magnet connection loose or disconnected (occasional)
Verify lid switch light indicator illuminates when lid is open
- Test point
- Light indicator (0009) on C40K or microprocessor board
- Expected
- Light on when lid open
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water filtration system lights not functioning
Filtration system indicator lights not illuminating
Ranked causes
- 1. Burned out light bulbs (common)
- 2. Blown fuses on control unit (common)
- 3. Power supply issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin fill to correct level fails
Basin does not fill to appropriate operating level when tap water supply activated
Ranked causes
- 1. Basin float switch malfunction (occasional)
- 2. Tap water valve stuck or blocked (occasional)
Observe basin filling process
- Test point
- Basin water level during fill cycle
- Expected
- Basin fills to appropriate level before float actuates
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E07 - Power failure during cycle patient safety
Power loss occurred during processing; cycle may be incomplete with residual fluid in chamber.
Ranked causes
- 1. Power outage during cycle execution (common)
- 2. Residual fluid in chamber from interrupted cycle (occasional)
Verify utility power is restored
- Test point
- Power inlet and circuit breaker
- Expected
- Power present, breaker in ON position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cycle interrupted by lid opening
Active disinfection or wash cycle halts when reprocessor lid is lifted during operation
Ranked causes
- 1. Operator opens lid during cycle (common)
- 2. Lid switch malfunction (occasional)
Verify lid is fully closed
- Test point
- Lid position and magnet alignment
- Expected
- Lid fully seated with magnet aligned to switch
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant contact hazard from improper reservoir management patient safety
Technician error in handling disinfectant reservoir during Service Box troubleshooting
Ranked causes
- 1. Disinfectant not drained from reservoir before troubleshooting (common)
- 2. Water left in reservoir at end of certification without customer notification (occasional)
Verify reservoir is drained before troubleshooting
- Test point
- Disinfectant reservoir
- Expected
- Empty or containing only water
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Harness damaged or disconnected caution
Harnesses show visible damage, are crimped, cut, or connectors are pulled apart
Ranked causes
- 1. Physical damage to harness from equipment operation or maintenance (occasional)
- 2. Connector not fully seated or intentionally disconnected (common)
Visually inspect harness for cuts or crimps
- Test point
- Entire length of harness
- Expected
- No visible damage, intact insulation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump or solenoid valve indicator light not illuminated
Controller/microprocessor indicator light for pump or solenoid valve does not light when component should be active
Ranked causes
- 1. Sensor warning signal present (e.g., lid open) (common)
- 2. Problem with controller/microprocessor (occasional)
Verify lid is fully closed
- Test point
- Equipment lid/door
- Expected
- Closed and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Associated component does not operate despite indicator on and power present caution
Output indicator illuminates and correct voltage present at connector, but device (valve, pump, etc.) remains inactive
Ranked causes
- 1. Component internal failure or wiring harness disconnected/damaged (common)
Verify indicator is illuminated
- Test point
- Channel 1, 3, or 4 indicator light
- Expected
- Indicator LED lit when component should operate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back-up battery failure
Time and cycle settings lost after power off; back-up battery in Omron controller no longer maintains memory
Ranked causes
- 1. Back-up battery depleted or failed (typical life 5 years) (occasional)
Measure voltage between controller/microprocessor terminal and COMMON
- Test point
- controller/microprocessor terminal and COMMON
- Expected
- no voltage indicates battery is good; 115 VAC indicates battery should be replaced, 0 V (pass); 115 VAC (fail)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back-up battery failure indicated by high voltage at controller terminal high voltage
Back-up battery no longer maintains time and cycle settings; 115 VAC measured across controller/microprocessor terminal and COMMON indicates battery depletion.
Ranked causes
- 1. Back-up battery end-of-life (5-year life expectancy) (occasional)
Measure voltage between controller/microprocessor terminal and COMMON
- Test point
- Controller/microprocessor terminal and COMMON
- Expected
- No voltage (0 VAC)
- Tools
- Voltmeter or DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin overfilling caution
Endoscope reprocessor basin continues to fill beyond normal level
Ranked causes
- 1. Tap water valve stuck in open position (common)
Test tap water valve closure
- Test point
- tap water valve outlet
- Expected
- valve stops water flow when commanded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Product Certification
System performance certification procedure following PM, installation, or component replacement
Ranked causes
- 1. Routine certification required after maintenance or installation (common)
Check for leaks
- Test point
- System connections and seals
- Expected
- No visible leaks
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Certification adjustment or replacement indicated
System requires adjustment or component replacement if certification procedures reveal performance issues
Ranked causes
- 1. Defective component or misalignment detected during certification (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E02 maximum disinfect cycle count reached
Number of disinfect cycles equals maximum programmed count; only EMPTY DISINFECTANT cycle function available
Ranked causes
- 1. Disinfect cycle count limit reached (common)
Verify E02 error code display
- Test point
- Reprocessor display
- Expected
- E02 code shown
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code E04 with proper water drainage needed
E04 displayed when verifying float switch operation after water dislodging procedure
Ranked causes
- 1. Float switch stuck or defective after dislodging procedure (occasional)
Verify light status after EMPTY cycle
- Test point
- C40K board lights 0011, 0012, 0013
- Expected
- 0011 lit; 0012 and 0013 not lit
Cycle time changes not communicated to customer
Service Box troubleshooting may alter cycle times; customer not informed of changes
Ranked causes
- 1. Technician changes cycle times without documenting or communicating to customer (occasional)
Verify cycle times match original configuration or document changes
- Test point
- Reprocessor cycle time settings
- Expected
- Original settings or documented changes recorded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant approaching replacement
System nearing need for disinfectant replacement; fluid levels low
Ranked causes
- 1. Disinfectant concentration depleted (common)
Access Service Box and verify fluid drain capability
- Test point
- Service Box drain port
- Expected
- Service Box positioned to drain fluids
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant reservoir empty patient safety
Disinfectant level depleted; system cannot execute FILL DISINFECTANT cycle
Ranked causes
- 1. Normal disinfectant consumption during cycles (common)
Verify disinfectant reservoir level
- Test point
- Disinfectant reservoir visual indicator
- Expected
- Reservoir above minimum level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant reservoir leaks caution
Fluid leaks from replacement reservoir during installation or operation; water or disinfectant pooling around reservoir
Ranked causes
- 1. Defective replacement reservoir (occasional)
Inspect replacement reservoir for leaks before installation
- Test point
- Replacement reservoir exterior surface
- Expected
- No water leakage when reservoir filled with water test full reservoir
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disinfectant system operation high voltage
Service procedures for disinfectant pump, valves, and associated liquid handling systems
Ranked causes
- 1. Service Box power and valve control setup for disinfectant circulation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drain disinfectant from reservoir caution
Procedure to evacuate liquid from disinfectant reservoir through service connector
Ranked causes
- 1. Service maintenance or disinfectant reservoir emptying required (common)
Verify all three required switches are set to ON
- Test point
- Disinfectant pump, Main pump, Waste Valve switches on Service Box
- Expected
- All three switches in ON position; disinfectant reservoir valve remains CLOSED
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid pathway operation during troubleshooting high voltage
Procedure to transfer liquid between points by setting appropriate switches and consulting flow diagram.
Ranked causes
- 1. Diagnostic testing of fluid handling system (common)
Select fluid pathway
- Test point
- Flow diagram and Service Box switches
- Expected
- Correct pathway identified from diagram; all associated switches available
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Harness or connector damaged caution
Visible cuts, crimps, or pulled-apart connectors on pump/valve harnesses
Ranked causes
- 1. Harness physically damaged or loose (common)
Visually inspect all harnesses for cuts or crimps
- Test point
- Harness lengths and routing path
- Expected
- No visible cuts or crimped insulation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater element malfunction risk caution
Disinfectant level insufficient to keep heater submerged; heater disable switch activated
Ranked causes
- 1. Disinfectant level drops below heater element (occasional)
Check disinfectant level relative to heater element
- Test point
- Disinfectant level in reservoir
- Expected
- Level above heater element
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Install show chip for demonstration mode caution
ASP AER systems require a show chip for mock display of WASH, RINSE, DISINFECT, and AIR cycles without water and waste supply connection
Ranked causes
- 1. System configured for demonstration purposes (unknown)
Disconnect ASP AER from power source
- Test point
- Power supply
- Expected
- System powered down
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient air flow during air purging caution
Air compressor output pressure is insufficient; reduced air flow during endoscope channel purging cycle
Ranked causes
- 1. Air compressor output pressure set too low or relief valve requires adjustment (common)
Remove all tubing sets from endoscope channel ports
- Test point
- Endoscope channel ports
- Expected
- All tubing sets disconnected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lid switch not detecting closure
Lid switch fails to sense when lid is properly closed; indicator does not turn off
Ranked causes
- 1. Magnet on lid not fully seated (common)
Verify magnet seating when lid closed
- Test point
- lid magnet
- Expected
- magnet fully seated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid may be forced back into air compressor caution
When passing air through the reprocessor, liquid can back-flow into the air compressor
Ranked causes
- 1. Air valve opened before line pressurization (common)
Ensure all Service Box switches are OFF
- Test point
- Service Box switches
- Expected
- All OFF
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main pump does not operate or flow is inadequate caution
Pump fails to deliver disinfectant fluid; inadequate flow rate or no flow observed when pump should be running
Ranked causes
- 1. Main pump failure or inadequate performance (occasional)
Verify pump power and fluid supply availability
- Test point
- Pump motor power connection; disinfectant reservoir level
- Expected
- Power applied to pump; fluid available in reservoir
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microprocessor board replacement required caution
Procedure to remove and replace microprocessor board; device is static sensitive
Ranked causes
- 1. Microprocessor board hardware failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Omron controller replacement required caution
Procedure to remove and replace failed Omron controller on ASP AER endoscope reprocessor
Ranked causes
- 1. Omron controller hardware failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output indicator does not illuminate
Controller output indicator LED stays off despite command or during expected operation
Ranked causes
- 1. Controller circuit fault or missing input signal activation (common)
Verify power supply to controller
- Test point
- Channel 1, 3, or 4 connector terminal
- Expected
- 115 VAC on channel 1, 24 VDC on channels 3 and 4
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Power line voltage drops below minimum
- Printer does not print or prints faded
- Printer output shows fading
- Printer paper depleted
- Pump running empty causes equipment damage
- Relays not functioning for pump control
- Replace disinfectant reservoir
- Replace microprocessor board
- Replace Omron Controller
- Reservoir leaks during WASH AND DISINFECT cycle
- Service Box connection and operation
- Solenoid valve does not operate
- System not responding to input commands
- Unit does not operate properly
- Unit does not operate properly when power line voltage drops
15 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E01 | High disinfectant level warning; possible water dilution | High-level float switch activated or test strips fail (indicating dilution) | warning |
E01 | High level warning; disinfectant reservoir too full or disinfectant diluted with water | High level warning switch activated | warning |
E02 | Maximum disinfect cycle count reached | Number of completed disinfect cycles equals programmed maximum limit | warning |
E03 | Disinfectant reservoir empty | Disinfectant level drops below minimum during operation or cycle | error |
E03 | End of cycle status code | Received at completion of certification cycle | info |
E04 | Disinfectant reservoir contains fluid | Liquid present in disinfectant reservoir preventing FILL DISINFECTANT cycle | warning |
E05 | Low disinfectant level in reservoir (fewer than 3.5 or 4.5 gallons remaining) | Disinfectant drops below threshold level depending on model variant | warning |
E06 | Fresh water insufficient flow to basin | Basin float switch not activated before backup timer expiration | error |
E07 | Power failure during cycle | Mains power lost during active processing cycle | warning |
E08 | Wash time not entered or missing disinfectant | System detects WASH TIME parameter is absent or disinfectant reservoir is empty | warning |
E09 | Disinfect time not entered | Disinfect time field left empty before pressing START to begin a cycle | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.