ADAC Vertex Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 60 documented symptoms for the ADAC Vertex Plus, with 70 ranked causes, 96 diagnostic checks with test points and expected values, and 101 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Encoder position discrepancy detected
System detects mechanical slip or linkage problems indicated by difference between encoder and potentiometer position readings.
Ranked causes
- 1. Drive chain fallen off sprocket (occasional)
- 2. Gear slipping relative to potentiometer (occasional)
- 3. Potentiometer slipping relative to encoder (occasional)
- 4. Encoder slipping relative to mechanism (occasional)
- 5. Mechanism slipping relative to potentiometer (occasional)
Monitor encoder and potentiometer position agreement during motion
- Test point
- Display Panel error code status
- Expected
- No U3 error code
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector radius patient safety
Detectors may collide or hit table if radius calibration is incorrect, especially when selecting Cardiac 90 & Lock preprogram option or performing ECT studies with heads in 90° position.
Ranked causes
- 1. Radius calibration out of specification (common)
- 2. Radius Sensor locations not stored accurately (common)
Position detector #1 at first calibration point
- Test point
- Brass Leadscrew Support Nut on right side of detector
- Expected
- 2.9 cm (11/8 inch) above black Thrust Support Block
- Tools
- Handcontrol, level
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry rotation
Encoder and potentiometer positions must be synchronized for rotation mechanism. Calibration establishes the relationship between potentiometer voltage and encoder pulses.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
- 2. Calibration performed at incorrect angle range (occasional)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor Controller Gain and Speed
Gantry motions (Rotate, Translate, Radius, Table Z/X) require bidirectional speed balancing and max speed adjustment via potentiometers on Motor Controller boards to prevent stalling, Maytagging (overshooting), jerky movement, hunting, or motion errors.
Ranked causes
- 1. IR Comp pot out of adjustment (common)
- 2. Max Speed pot out of adjustment (common)
Enter Cal Mode and access MC Gain option
- Test point
- Display Panel
- Expected
- MC Gain option appears with motion selection (G ROT, RAD 1, RAD 2, etc.)
17 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor Controller Initial Potentiometer Setup
Initial pot and switch positions on Motor Controller boards must be set to predetermined limits before adjustment of Max Speed and IR Comp pots. Five pots (DB, RevTQ, FwdTQ, SloSpd, AccDec) and SW 501 switches require one-time setup.
Ranked causes
- 1. Pots and switches not set to initial positions (common)
- 2. SW 501 switches not set to initial positions (occasional)
Verify if installing new Motor Controller board
- Test point
- Hardware installation status
- Expected
- New controller board mounted on Motor Controller Panel Assembly
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor controller speed and stability caution
Motor controller boards on the Motor Controller Panel Assembly require calibration of Max Speed and IR Comp potentiometers to ensure proper gantry motion control without stalling or oscillation.
Ranked causes
- 1. Incorrect IR Comp setting causing directional speed mismatch (common)
- 2. Incorrect Max Speed setting causing stalling or oscillation (common)
Verify motor speeds in opposite directions are equal
- Test point
- Motor motion in CW and CCW directions
- Expected
- Same speed in both directions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Potentiometer replacement procedure
Steps to replace a potentiometer on any gantry motion mechanism while ensuring correct centering and calibration.
Ranked causes
- 1. Potentiometer failure or drift (occasional)
- 2. Incorrect wiper position during installation (common)
Center potentiometer wiper before installation
- Test point
- Potentiometer wiper position
- Expected
- Fully to one end, then 5 turns in opposite direction
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrate gantry motions
Perform voltage, speed, motor control gain and software limit tests for gantry translate, rotate, and other motions
Ranked causes
- 1. Routine calibration required (common)
Navigate to Calibrate Motions menu
- Test point
- Gantry Display Panel
- Expected
- Calibrate Motions option displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector 1 radius
Encoder and potentiometer synchronization for detector 1 radial motion (zoom in/out). Applies to all models except Cardio.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector 2 radius
Encoder and potentiometer synchronization for detector 2 radial motion (zoom in/out). Applies to all models except Cardio.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector Latch Sensors patient safety
Adjustment of latch sensors on detector #1 to ensure collimator is properly detected as latched. Incorrect latch sensor setting may allow system to continue exchange when collimator is not securely latched.
Ranked causes
- 1. Latch sensor position not detecting magnet halfway through lock notch (occasional)
Verify latch sensor tone at bottom of lock notch
- Test point
- Foot-end, front corner of Detector #1 lock plate
- Expected
- Tone should be on when latch pin is at bottom
- Tools
- Audible sensor feedback
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector Leveling caution
Leveling detector heads to ensure perpendicular orientation and prevent rocking, critical for accurate ECT reconstructions
Ranked causes
- 1. Detectors not level (common)
Verify Radius Base Plates vertical before leveling detector
- Test point
- Both detector Radius Base Plates
- Expected
- Vertical
- Tools
- Small level
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector Leveling initial setup
Initial setup for detector leveling procedures, including gantry rotation and detector positioning
Ranked causes
- 1. Detector leveling prerequisites not met (common)
Confirm Gantry Rotate calibration complete
- Test point
- Calibration status display
- Expected
- Gantry Rotate cal points stored
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector Lock Plate Sensors caution
Adjustment of lock plate sensors on detectors #1 and #2 to ensure proper magnet position detection during collimator locking. Sensors must detect magnet locations accurately.
Ranked causes
- 1. Lock plate magnet off-center (occasional)
Test Detector #1 lock plate sensor unlock position
- Test point
- Foot-end and head-end lock plates on Detector #1
- Expected
- Tone should sound when lock plates move to unlock notch
- Tools
- Audible sensor feedback
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Foot vertical table motion
Encoder and potentiometer synchronization for vertical (Z-axis) motion of patient table foot end.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry calibration and limit data backup
Back up gantry calibration and limit data from RAM to PC files and floppy disc for disaster recovery.
Ranked causes
- 1. Routine maintenance and disaster recovery (common)
Access DOS prompt from Acquisition Console
- Test point
- Main Menu
- Expected
- DOS prompt accessible via EXIT selection
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry motion
System requires recalibration after potentiometer replacement or U-3 error to set scale factor relating voltage to mechanism position
Ranked causes
- 1. Potentiometer replaced or U-3 encoder mismatch error (common)
Enable Cal Mode on MPU board
- Test point
- Cal Mode Rocker Switch (3rd from left on MPU PCB)
- Expected
- Switch activated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry Rotate caution
Gantry rotation angle calibration using potentiometer feedback at -180° and +180° positions, verified with level on Radius Base Plate
Ranked causes
- 1. Gantry 0° and 180° positions not level (common)
Verify Radius Base Plate is vertical at -180° gantry position
- Test point
- Left detector #1 Radius Base Plate
- Expected
- Level (bubble centered)
- Tools
- Small level
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry Rotate at initial setup
Initial gantry rotation setup ensuring detectors are properly positioned and not at hardware limits
Ranked causes
- 1. Gantry not properly initialized before calibration (common)
Confirm detectors are 180° apart
- Test point
- Detector angular position display
- Expected
- 180° separation
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry translate
Encoder and potentiometer synchronization for gantry translation (longitudinal movement along acquisition axis).
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head vertical table motion
Encoder and potentiometer synchronization for vertical (Z-axis) motion of patient table head end.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lateral table motion
Encoder and potentiometer synchronization for lateral (X-axis) motion of patient table. Does not apply to Solus model.
Ranked causes
- 1. Potentiometer wiper not centered on new installation (common)
Verify potentiometer wiper centering after installation
- Test point
- Potentiometer mid-range
- Expected
- 2000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Radius check patient safety
Independent verification that radius calibration has been performed correctly and detectors are properly positioned to prevent collisions during Cardiac 90 & Lock operations or ECT studies.
Ranked causes
- 1. Radius calibration dimensions out of tolerance (common)
Measure distance between T-square center and detector #1 bucket edge
- Test point
- Detector #1 bucket edge (distance X)
- Expected
- Equal to detector #2 measurement within ±1/16 inch inches, ±1/16 inch
- Tools
- Carpenter's T-square
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Radius Sensors Setting caution
Set radius sensor magnet locations in system RAM. Radius sensors detect magnet positions on radius base plate for accurate detector positioning during automatic collimator exchanges.
Ranked causes
- 1. Radius sensor magnet locations not stored in RAM (common)
Set Radius #1 sensor magnet location
- Test point
- Detector #1 radius drive
- Expected
- System finds and stores magnet location in RAM, reprints RADIUS 1 POS SENS message
- Tools
- Display panel, Cal rocker switch
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Carriage park sensor not as expected
Park sensor state is incorrect, preventing normal gantry operation
Ranked causes
- 1. Carriage park sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
E-Stop error with red emergency lights remaining on caution
E-Stop error occurs that doesn't involve a collision, causes the red emergency lights on the Display Panel to remain on, and cannot be cleared by pressing the manual override button.
Ranked causes
- 1. Out-of-range power supply voltages (common)
Measure +9.85V Reference voltage
- Test point
- MPU board ADC Mux value 0
- Expected
- 9.87 V, 9.82 to 9.92 V
- Tools
- Digital voltmeter, oscilloscope
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Encoder handler timeout
Bit slice code not responding, indicating MPU board failure
Ranked causes
- 1. Bad MPU board (bit slice code not responding) (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry motor belt broken
Motor belt broken, detected by broken belt sensor
Ranked causes
- 1. Motor belt broken (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry moved opposite to requested direction
Motor moved in wrong direction, indicating wiring issue
Ranked causes
- 1. Incorrect motor wiring (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry MPU diagnostic tests fail for unexplained reasons
Diagnostic tests on the Gantry MPU board fail without a clear cause.
Ranked causes
- 1. Out-of-specification power supply voltages (common)
Perform complete voltage sweep on MPU board
- Test point
- All ADC Mux values 0, 1, 4, 5, 6, 7
- Expected
- See Table 4-1 for each Mux value
- Tools
- Digital voltmeter, oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry MPU replacement with calibration restoration
Restore calibration after replacing Gantry MPU by reloading GANTCAL.DAT and rechecking all calibrations.
Ranked causes
- 1. Gantry MPU replaced (rare)
Verify GANTCAL.DAT file is available from backup
- Test point
- Floppy disc archive
- Expected
- GANTCAL.DAT file readable and intact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry Rotate 0° sensor not as expected
Rotation sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Gantry Rotate 1, 0° sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry Rotate 180° sensor not as expected
Rotation sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Gantry Rotate 2, 180° sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry Rotate Clutch Error, Motion Halted
Clutch disconnected or slipped, causing gantry rotation position change >2°
Ranked causes
- 1. Gantry Rotate Clutch disconnected or slipped (position change >2°) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry Rotate floating
Ring lock sensor unreadable, causing floating gantry rotation without lock
Ranked causes
- 1. Ring lock sensor unreadable (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry will not move under Handcontroller commands caution
Gantry does not respond to movement commands sent via Handcontroller.
Ranked causes
- 1. Out-of-range power supply voltages affecting gantry motor control (common)
Verify +5V supply voltage is within range
- Test point
- MPU board ADC Mux value 5
- Expected
- 5.0 V, 4.8 to 5.2 V
- Tools
- Digital voltmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hardware limit reached caution
Gantry mechanism hit physical limit due to negligence in calibration mode or bad calibration/limit sensor
Ranked causes
- 1. Negligence in calibration mode or bad calibration/limit sensor (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head #1 collimator lock sensor not as expected
Collimator lock sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Head #1 collimator lock sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head #1 target sensor not as expected
Target sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Head #1 target sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head #2 collimator lock sensor not as expected
Collimator lock sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Head #2 collimator lock sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head #2 target sensor not as expected
Target sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Head #2 target sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Limit checksum failed
RAM checksum failure indicating corrupted limit data requiring reinitialization
Ranked causes
- 1. MPU board RAM corrupted (limit checksum failed) (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NUAC line source fault condition patient safety
NUAC controller detects fault and is inoperable; gantry may reset and study will be interrupted
Ranked causes
- 1. NUAC controller fault; motor not connected, wiring fault, or bad bias resistor (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Potentiometer encoder mismatch
Pot and encoder values do not match, causing calibration error
Ranked causes
- 1. Bad potentiometer, bad encoder, or bad calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Potentiometer reading out of range
Pot voltage outside valid range, preventing proper position measurement
Ranked causes
- 1. Potentiometer outside valid range 350-3900 (0-4000 for Table Z) (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Potentiometer slipping
Potentiometer calibration values too close together, indicating slippage or bad calibration
Ranked causes
- 1. Bad potentiometer or bad calibration (max/min values too close) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preprogrammed motion internal error
Internal programming error preventing preprogrammed motion execution
Ranked causes
- 1. Programming error in preprogrammed motion routine (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radius Drive 1 sensor not as expected
Radius drive sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Radius Drive 1 sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radius Drive 2 sensor not as expected
Radius drive sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Radius Drive 2 sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensors not present
Serial sensor or SAD board is not detected or has failed
Ranked causes
- 1. Bad serial sensor or SAD board (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software limit hit
Gantry motion exceeds configured software limits, indicating calibration error
Ranked causes
- 1. Out of calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spectrum problems occur during spectrum changes
Spectrum acquisition issues occur, particularly when making spectrum changes; need to monitor PM tube high voltages.
Ranked causes
- 1. Unstable or out-of-range power supply voltages during spectrum transitions (occasional)
Monitor +15V supply voltage stability during spectrum change
- Test point
- MPU board ADC Mux value 7
- Expected
- 15.0 V stable, 14.5 to 15.5 V
- Tools
- Digital voltmeter or Display Panel monitoring
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System oscillates before reaching target position caution
Motor oscillates or 'Maytags' around the target position rather than smoothly reaching it.
Ranked causes
- 1. Max Speed or IR Comp set too high (common)
Observe motor behavior for oscillation around target
- Test point
- Motor motion near target position
- Expected
- Motor smoothly approaches and reaches target without oscillation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System stalls before reaching target position caution
Motor fails to reach the target position and stops moving before arrival.
Ranked causes
- 1. Max Speed or IR Comp set too low (common)
Observe motor behavior during motion to target
- Test point
- Motor motion toward target position
- Expected
- Motor reaches target position or creeps steadily toward it
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test gantry rotation position sensors
Verify GROT 1 and GROT 2 position sensors functioning at 0° and 180° positions
Ranked causes
- 1. Position sensor verification required (common)
Verify GROT 1 position sensor at 0°
- Test point
- GROT 1 Position 0 Sensor
- Expected
- Sensor active at 0° gantry rotation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Translate sensor not as expected
Translate mechanism sensor state is incorrect, preventing normal operation
Ranked causes
- 1. Translate sensor out of correct state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to lock ring
Ring lock device is not in correct position and is slipping, preventing proper locking
Ranked causes
- 1. Ring lock device out of position and slipping (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unrequested motion occurred patient safety
Gantry moved without handcontrol or software command due to loose cables, user extraction, or motor controller failure
Ranked causes
- 1. Cables loose, user manually extracted patient, or motor controller failed (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unwanted spurious interrupt
System detects unexpected hardware interrupt or internal software bug
Ranked causes
- 1. Hardware noise or internal software bug (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify collimator identification
Read electrical identification number of mounted collimators and verify stability during gantry rotation
Ranked causes
- 1. Collimator identification check required (common)
Display collimator identification numbers
- Test point
- Service Mode menu
- Expected
- Identification numbers of mounted collimators visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
F-25 | Limit checksum failed in MPU board | RAM checksum validation fails on MPU board | critical |
F-81 | Unrequested motion | Gantry motion occurred without handcontrol or software command | critical |
F-92 | Gantry moved opposite to requested direction | Motor moved in wrong direction | critical |
F-94 | Gantry motor belt broken | Broken belt sensor detects fault | critical |
F-99 | Hardware limit reached | Gantry mechanism hit physical limit | critical |
R-1 | Communication error | DUART parity, framing or overrun error | warning |
R-12 | Detector head has improper collimator type and parity | Bad collimator ID tag or improper collimator setup | error |
R-16 | Detector head collision | Physical collision or wiring issue detected | error |
R-18 | Manual override active | Override exists in pre-program or computer mode | info |
R-19 | Axis out of range | PC commanded gantry motion but did not exit motion command | error |
R-2 | Emergency stop | E-stop button pressed | error |
R-21 | Cannot update calibration RAM | Cal switch is off or Cal RAM write attempted without enable | error |
R-22 | Parameter error | PC received command with incorrect parameter | error |
R-23 | Bad checksum in Cal RAM | Cal RAM checksum validation failed | error |
R-24 | Gantry is not paused | Resume sent by PC when gantry was not paused | error |
R-27 | Cal enable switch is off | Attempt was made to write to Cal RAM with Cal switch off | error |
R-29 | Suspend not allowed | Pause from PC not allowed in current operating mode | info |
R-30 | Motion already active | PC attempted to move a motion that was already moving | warning |
R-32 | NUAC shutter not opened | Shutter on specified line source did not open when commanded | error |
R-33 | NUAC: Line source at invalid position | PC tried to move Gantry Rotate #2 relative to #1 when NUAC line sources at incorrect position | warning |
R-38 | NUAC: Line source error - no move setup done | Acquisition station or diagnostic issued move command to NUAC device without preceding setup command | error |
R-39 | Unexpected position sensor state in preprogrammed mode | Sensor malfunction during preprogrammed operation | warning |
R-4 | Exchanger trays not parked | Automatic or manual exchanger out of position; bridge up sensor not set or bad carriage park sensor | warning |
R-40 | Radial map undefined | NCR motion attempted before defining radial map | warning |
R-43 | Radial map bounds error | Radial map caused motion to move outside software limits | error |
R-44 | NUAC: not enabled | NUAC device installed but enable jumper not set | warning |
R-45 | NUAC: Line source error - line source not mounted | NUAC acquisition or preprogrammed command executed when line sources not mounted | error |
R-46 | NUAC: Line source error - line source not parked | Command or preprogrammed option detected line sources not parked, possibly from premature stop | error |
R-47 | NUAC: Line source error - line source shutter not closed | Line source detected open when not commanded to open | error |
R-48 | NUAC: Line source error - wrong line source mounted | System detected wrong line source mounted on at least one head | error |
R-49 | NUAC: Line source error - line source not communicating | Power not getting to NUAC device or communication failure | error |
R-52 | Preprogrammed motion not allowed in Override | Override mode active when preprogrammed motion attempted | warning |
R-53 | Head #1 Rotate out of range for preprogrammed motion | Head Rotate in outer room position during preprogrammed motion | warning |
R-54 | Preprogrammed motion not allowed in current state | System in collision or E-stop state | error |
R-56 | Preprogrammed motion is aborted | User stopped PC-initiated preprogrammed sequence | info |
R-57 | Head #2 Rotate out of range for preprogrammed motion | Head Rotate in outer room position during preprogrammed motion | warning |
R-58 | Head #2 not in position | Head not in position to exit outer room | warning |
R-5A | Motion not allowed | On Cardio system, gantry motion when detectors at relative 180° | warning |
R-5B | Direction of movement not allowed | Gantry moved in wrong direction with rings unlocked in diagnostic mode | warning |
R-5F | Slave motion incomplete | Motion did not reach final position | error |
R-6 | Relative motion error | Slave motion not keeping up with master during simultaneous motion | warning |
R-60 | Motion incomplete | Motion did not reach final position | error |
R-61 | Active motion type change | Software tried to change type of active motion | critical |
R-62 | Slave startup error | Slave and master position mismatch at motion startup | error |
R-63 | Overheat watchdog (Manual Exchanger) | User failed to unlatch collimator within 2 minutes | critical |
R-64 | Table motion not allowed | Fast Table Z attempted when carriage not parked | warning |
R-76 | Collimator locking mechanism not latched | Collimator latch sensor off or collimator latch open | error |
R-77 | Locking plate error | Locking plate sensor not unlocked | error |
R-78 | Detector Head Lock Plate Lock sensor not locked | Lock Pin sensor off or collimator on detector | error |
R-79 | Position sensor not calibrated | During sensor calibration, software could not find sensor | error |
R-8 | Motor power off | Motor power is off for any reason that activates motion halt lights | warning |
R-83 | Head #1 is unlocked | Detector unlocked state | warning |
R-84 | Head #2 rotate is unlocked | Detector unlocked state | warning |
R-85 | Gantry position error | Gantry not at correct position for lock/unlock | error |
R-88 | Motion not allowed | User attempted motion with Cal or limit switch on | warning |
R-91 | Command length error | Gantry received command with unexpected length | error |
R-96 | Mechanical calibration does not agree with sensor | Target sensor not seeing cassette or sensor not working or motion not calibrated | error |
R-a1 | Collimator exchanger motion failure | Collimator exchanger motion did not reach destination | error |
R-a2 | Collimator exchanger latch sensor error | Latch sensor in unexpected state | error |
R-a3 | Collimator exchanger lock sensor error | Lock sensor in unexpected state | error |
R-a4 | Collimator exchanger unlock sensor error | Unlock sensor in unexpected state | error |
R-a5 | Collimator exchanger locking plate sensor not as expected | Locking Plate Lock and Unlock Sensors both off or mismatched state | error |
R-a6 | No collimator ID | ID cannot be read from tag after collimator mounted | error |
R-a7 | Collimator exchanger map error | Collimator map incorrect or sensor not working | error |
R-a8 | Collimator exchanger communication error | Probable noise in communication or not connected | error |
R-a9 | Collimator exchanger is not ready | Elevator position out of range or 2nd pot cannot be read | error |
R-aa | Collimator exchanger not safe to move | Position or sensor indicates not safe for motion | error |
R-ab | Collimator exchanger carriage position failure | Carriage position by pot #1 differs from pot #2 | error |
R-ac | Collimator exchanger cannot be parked | Motion not at exchange position for preprogrammed park command | error |
R-ad | Preprogrammed internal script/sequence error | Internal software error in programmer or sequence | critical |
R-ae | Preprogrammed time-out | Waiting for display to set override | error |
R-af | Collimator Exchanger not allowed with brain pallet | Collimator exchange attempted with brain pallet installed | warning |
R-b0 | Sensor not as expected - Head #1 locked in | Head #1 Lock sensor not in correct state | warning |
R-b1 | Sensor not as expected - Head #1 locked out | Head #1 Lock sensor not in correct state | warning |
R-b2 | Sensor not as expected - Head #2 locked in | Head #2 Lock sensor not in correct state | warning |
R-b3 | Sensor not as expected - Head #2 locked out | Head #2 Lock sensor not in correct state | warning |
R-b4 | Gantry rotate 90 lock sensor not as expected | Gantry rotate 90 lock sensor not in correct state | warning |
R-b5 | Gantry rotate 2 180 lock sensor not as expected | Gantry rotate 2 180 lock sensor not in correct state | warning |
R-b6 | Carriage park sensor not as expected | Park sensor not in correct state | warning |
R-b7 | Head #1 collimator lock sensor not as expected | Collimator lock sensor not in correct state | warning |
R-b8 | Head #2 collimator lock sensor not as expected | Collimator lock sensor not in correct state | warning |
R-b9 | Head #1 target sensor not as expected | Target sensor not in correct state | warning |
R-ba | Head #2 target sensor not as expected | Target sensor not in correct state | warning |
R-bb | Gantry Rotate 2, 180° sensor not as expected | Rotation sensor not in correct state | warning |
R-bc | Gantry Rotate 1, 0° sensor not as expected | Rotation sensor not in correct state | warning |
R-bd | Translate sensor not as expected | Translate mechanism sensor not in correct state | warning |
R-be | Radius Drive 1 sensor not as expected | Radius drive sensor not in correct state | warning |
R-bf | Radius Drive 2 sensor not as expected | Radius drive sensor not in correct state | warning |
U-10 | Unwanted spurious interrupt | Unexpected hardware interrupt or internal software fault | warning |
U-14 | Software limit hit | Gantry motion exceeds configured limits | warning |
U-20 | Potentiometer slipping | Max and min calibration values too close together | warning |
U3 | Encoder and potentiometer position out of tolerance | Detected difference between encoder-determined and potentiometer-determined positions exceeds acceptable range | error |
U-3 | Potentiometer encoder mismatch | Pot and encoder values do not match | error |
U-34 | NUAC line source fault | NUAC controller detects fault condition | critical |
U-5 | Potentiometer reading out of range | Pot voltage outside 350-3900 range (0-4000 for Table Z) | error |
U-59 | Gantry Rotate floating | Ring lock sensor unreadable | error |
U-7 | Encoder handler timeout | Bit slice code not responding | error |
U-89 | Gantry Rotate Clutch Error, Motion Halted | Clutch disconnected or slipped (position change >2°) | error |
U-98 | Unable to lock ring | Ring lock device not in correct position | error |
U-a0 | Preprogrammed motion internal error | Internal programming error in preprogrammed motion routine | error |
U-f | Sensors not present | Serial sensor or SAD board not detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI