ACIST Medical Systems CVi: troubleshooting and error codes
The ServiceForge knowledge base holds 51 documented symptoms for the ACIST Medical Systems CVi, with 106 ranked causes, 171 diagnostic checks with test points and expected values, and 23 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power supply does not power up injector high voltage
Injector head receives no power from the power supply unit; no 24 VDC output or system does not power on.
Ranked causes
- 1. Bad connection or faulty cable between power supply and injector head (common)
- 2. Blown fuse in power supply (common)
- 3. Voltage selection plug positioned incorrectly (common)
- 4. Incoming power outside system specifications (occasional)
- 5. Power supply is faulty (occasional)
- 6. Injector head is faulty (rare)
Inspect cable connection between power supply and injector head
- Test point
- Power connector between power supply and injector head
- Expected
- Cable fully seated and no visible damage
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contrast empty / no contrast alert patient safety
Control panel displays 'CONTRAST EMPTY / NO CONTRAST' alert message
Ranked causes
- 1. No contrast detected by universal contrast sensor (common)
- 2. Bubbles in tubing (common)
- 3. Tubing not placed properly in sensor (occasional)
- 4. Contrast sensor dirty (occasional)
- 5. Universal contrast sensor faulty (rare)
Verify contrast bottle status and tubing flow path
- Test point
- Contrast bottle and side clamp
- Expected
- Bottle not empty, side clamp open, air vent open
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hand controller calibration failed or requires calibration
Control panel displays 'HAND CONTROLLER CALIBRATION FAILED' or 'HAND CONTROLLER REQUIRES CALIBRATION' message
Ranked causes
- 1. Contrast button pressed too early or too late (common)
- 2. Wrong button pressed during calibration (occasional)
- 3. Hand controller not calibrated at setup (occasional)
- 4. Luer connections not tight (occasional)
- 5. Control panel PCBA faulty (rare)
Verify calibration message displayed before pressing contrast button
- Test point
- Control panel screen message
- Expected
- Calibration start message visible before button press
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold valve closed alert patient safety
Control panel displays 'MANIFOLD VALVE CLOSED' alert message
Ranked causes
- 1. Manifold not properly placed in holder (common)
- 2. Manifold stuck in open position due to syringe ram pressure (occasional)
- 3. Contrast tubing slide clamp closed (occasional)
- 4. Manifold sensor out of alignment (rare)
- 5. Manifold sensor faulty (rare)
Verify manifold is properly seated in holder
- Test point
- Manifold position in system holder
- Expected
- Manifold fully inserted and secured
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure limit exceeded alert patient safety
Control panel displays 'PRESSURE LIMIT EXCEEDED (1)' or 'PRESSURE LIMIT EXCEEDED (2)' alert
Ranked causes
- 1. Stopcock not open (common)
- 2. Catheter kinked (common)
- 3. Preset flow pressure too low (occasional)
- 4. Preset flow rate too high (occasional)
- 5. Rise time too short (occasional)
Verify stopcock is fully open
- Test point
- Stopcock valve at patient connection
- Expected
- Stopcock handle aligned with open flow path
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air column detected alert patient safety
Control panel displays 'AIR COLUMN DETECTED!' alert message
Ranked causes
- 1. Air in high pressure tubing (common)
- 2. Air column detector door not fully closed (occasional)
- 3. High pressure tubing not secured by strain relief (occasional)
- 4. Air column detector sensor faulty (rare)
Verify stopcock is off and inspect tubing for air bubbles
- Test point
- High pressure tubing visual inspection
- Expected
- No air bubbles visible in tubing
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby alert active
Control panel displays 'STANDBY' alert or system enters standby mode unexpectedly
Ranked causes
- 1. Standby button accidentally pressed (common)
- 2. Control panel to injector head cable connections loose (occasional)
- 3. Control panel to injector head cable faulty (occasional)
- 4. Standby switch faulty (rare)
Verify Standby button is not pressed
- Test point
- Standby button on right of control panel
- Expected
- Button released and not depressed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe valve open or closed alert
Control panel displays 'SYRINGE VALVE OPEN/CLOSED' alert message
Ranked causes
- 1. Syringe valve sensor disconnected or pins damaged (common)
- 2. Syringe valve sensor dirty (occasional)
- 3. Syringe valve sensor not operating correctly (occasional)
- 4. Syringe loaded before software prompt (occasional)
Inspect gray syringe valve cable connector
- Test point
- Gray syringe valve cable connector and pins
- Expected
- Connector clean, no damaged pins, red dots aligned
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air column detect messages not displayed when expected patient safety
Air visible in fluid pathway but system does not display air detection alert
Ranked causes
- 1. Air column detect sensor is dirty (common)
- 2. Air column detect sensor housing or tubing is wet (common)
- 3. Air column detect sensor is faulty (occasional)
Inspect air column detect sensor for dirt/contamination
- Test point
- Air column detect sensor
- Expected
- Sensor clean and clear
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Contrast sensor trigger point voltage
Adjustment procedure for contrast sensor detection threshold on injector pump electronics using multimeter measurements at test points T76, T59, and T77
Ranked causes
- 1. Contrast sensor trigger voltage requires calibration during pump electronics replacement (common)
- 2. Contrast sensor output voltage verification required after calibration (common)
- 3. Contrast sensor detection circuit verification with test fixture (common)
Measure voltage at test point T76 with digital multimeter set to millivolts DC
- Test point
- T76
- Expected
- Record measured voltage in mV
- Tools
- Digital multimeter
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contrast empty caution
Contrast sensor is not detecting the presence of contrast in the spike tubing.
Ranked causes
- 1. Contrast spike tubing clamp closed (common)
- 2. Contrast sensor door not snapped shut (common)
- 3. Contrast sensor dirty or obstructed (occasional)
Verify contrast spike tubing clamp is open
- Test point
- Contrast spike tubing clamp
- Expected
- Clamp in open position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contrast/saline button on hand controller not working patient safety
Hand controller contrast or saline selection button does not respond to presses
Ranked causes
- 1. Bad connection between hand controller and control panel (common)
- 2. Hand controller calibration not performed or performed incorrectly (common)
- 3. Hand controller is faulty (occasional)
Inspect and reconnect hand controller leur fittings
- Test point
- Hand controller connector and control panel bottom
- Expected
- Fittings fully seated and secure; no air gaps
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
General system functional errors
System exhibits functional errors not associated with specific alert messages.
Ranked causes
- 1. Power supply disconnected or voltage incorrect (common)
- 2. Damaged or loose cables/tubing (occasional)
- 3. System cleanliness issue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hand controller calibration failed
HC calibration was unsuccessful due to user timing error, button depression insufficient, loose connection, or defective unit.
Ranked causes
- 1. HC button not depressed within 5 second interval (common)
- 2. Contrast button on HC not depressed far enough (occasional)
- 3. HC connection to touch screen loose (occasional)
Attempt calibration with attention to 5 second interval
- Test point
- HC button
- Expected
- button depressed within 5 seconds of calibration start
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Medex pressure transducer reading
Syringe pressure reading displayed on control panel is lower than expected
Ranked causes
- 1. Faulty transducer or backplate (occasional)
- 2. Improper or inconsistent zeroing techniques (occasional)
- 3. Transducer cartridge not seated properly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peristaltic pump over torque caution
Peristaltic (saline) pump pressure required to dispense fluid exceeds internally set safety limit.
Ranked causes
- 1. Saline pump tubing improperly seated or latched (common)
- 2. Saline tubing blockage or kink (common)
- 3. Stopcock closed (occasional)
Verify saline pump tubing is properly seated and latched
- Test point
- Saline pump tubing connector
- Expected
- Tubing seated and latched securely, guides down and locked
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Contrast empty detection
Verification that contrast empty sensor correctly detects presence and absence of contrast fixture
Ranked causes
- 1. Contrast sensor requires testing for fixture presence and absence detection (common)
- 2. No Contrast warning system requires verification (common)
Wiggle contrast sensor fixture in holder
- Test point
- Contrast sensor test fixture position
- Expected
- Fixture displaced by wiggling
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Hand Controller and manifold fixture setup
Installation of hand controller test fixture and patient manifold valve test fixture with air column detection sensor
Ranked causes
- 1. Hand controller and manifold require test fixture calibration (common)
- 2. Air column detection sensor requires water and air column setup (common)
Insert Patient Manifold Valve Test Fixture
- Test point
- Manifold clip
- Expected
- Test fixture metal end in front of sensor
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Injection parameters and motor function patient safety
Verification of injection parameters, motor movement, actuator smoothness, and error-free operation
Ranked causes
- 1. Injection system requires parametric calibration and motor verification (common)
- 2. Motor and actuator require smooth operation verification (common)
Enter injection parameters
- Test point
- Control panel parameter fields
- Expected
- Mode: Pigtail, Flow Rate: 10 ml/sec, Volume: 20 ml, Pressure: 1200 PSI, Rise Time: 1.0 sec
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check syringe valve caution
Syringe valve sensor has detected an incorrect position of the syringe valve.
Ranked causes
- 1. Syringe valve in incorrect position (common)
- 2. Syringe loaded before 'Load Syringe' prompt (occasional)
Confirm syringe purge completed
- Test point
- System display
- Expected
- Purge operation successful
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostics download data corruption or transfer failure
Loss or corruption of diagnostics log data during retrieval from ACIST NVRam; incomplete or no transfer occurs
Ranked causes
- 1. Data in temporary directory overwritten by prior download (common)
- 2. Handheld device software interference (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure limit exceeded caution
System determined that on-the-fly flow rate limiting was not possible and halted injection. Total volume was not completed.
Ranked causes
- 1. Stopcock closed or catheter kinked (common)
- 2. Inadequate PSI setting or excessive flow rate (common)
Verify stopcock position
- Test point
- Stopcock valve
- Expected
- Open position
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby button depressed
Standby button has been depressed or control panel injector interface cable has become disconnected.
Ranked causes
- 1. Standby button pressed by user (common)
- 2. Control panel cable disconnected (occasional)
Verify standby button status
- Test point
- Standby button
- Expected
- Button not depressed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe valve sensor not functioning
Sensor cable disconnected, connector not mating, or sensor not properly installed in pressure sleeve assembly
Ranked causes
- 1. Syringe valve sensor cable disconnected or damaged (common)
- 2. Syringe valve sensor improperly installed or torqued (occasional)
Verify sensor cable connector is fully seated
- Test point
- Syringe valve sensor mating connector
- Expected
- Connector fully mated, no gap visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tubing moving forward in saline pump or constricted patient safety
Saline pump tubing advances through pump during operation or tubing is pinched/blocked
Ranked causes
- 1. Tubing guides need adjustment (common)
- 2. Damaged or broken teeth on guides (occasional)
Verify tubing guide position and tubing centering
- Test point
- Saline pump tubing guides
- Expected
- Both black guides at top; tubing centered with equal length each side; teeth over tubing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air column detected patient safety
Air column detect sensor has detected air in the tubing during operation.
Ranked causes
- 1. Air present in tubing (common)
Visual inspection of tubing for air bubbles
- Test point
- All tubing
- Expected
- No air visible
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration points are out of range
Touchscreen calibration points have drifted or are invalid
Ranked causes
- 1. Improper calibration technique or dried contrast on screen (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air column detection during purge
Verification of air column detection sensor operation and warning system with water and air in tubing
Ranked causes
- 1. Air column sensor requires water filling and air presence detection (common)
Clear air from air column detect sensor tubing
- Test point
- Air column detect sensor tube
- Expected
- Tubing contains water with at least one air column
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Chamber door safety interlock patient safety
Verification that chamber door open state is detected and blocks injection
Ranked causes
- 1. Chamber door sensor requires testing for open detection and injection blocking (common)
Open syringe mounting chamber
- Test point
- Syringe mounting chamber revolver
- Expected
- Chamber opened until it contacts plunger
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Contrast valve purge operation
Verification of contrast valve operation during syringe purge with sensor simulation
Ranked causes
- 1. Contrast valve requires syringe valve sensor simulation and purge completion verification (common)
Simulate open syringe valve
- Test point
- Syringe valve sensor
- Expected
- Finger placed in sensor
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Control Panel touch screen
Touch screen calibration procedure after system boot, ensuring accurate touchscreen response from normal use position
Ranked causes
- 1. Touchscreen requires calibration after boot (common)
Touch screen when ACIST display appears
- Test point
- Display screen
- Expected
- ACIST Medical Systems display screen appears
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: End case shutdown sequence
Verification of proper system shutdown and plunger retraction at end of service testing
Ranked causes
- 1. System shutdown requires proper end case procedure and plunger retraction (common)
Initiate end case procedure
- Test point
- Control panel END CASE button
- Expected
- END CASE button pressed
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient manifold valve detection
Verification of patient manifold valve sensor detecting open and closed manifold states
Ranked causes
- 1. Manifold valve sensor requires testing for open and closed state detection (common)
Remove Patient Manifold Test Fixture
- Test point
- Manifold fixture holder
- Expected
- Fixture removed from holder
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Saline flush operation
Verification of saline flush execution, peristaltic pump motor performance, and smooth quiet operation
Ranked causes
- 1. Saline flush system requires mode selection and motor verification (common)
Select Saline mode
- Test point
- Control panel mode selection
- Expected
- Saline mode selected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Standby function during injection patient safety
Verification that standby button halts injection and displays warning message
Ranked causes
- 1. Standby function requires testing during active injection (common)
Perform arbitrary injection with same parameters as before
- Test point
- Control panel injection setup
- Expected
- Previous injection parameters available
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe and contrast system setup
Setup and initialization of syringe valve sensor, contrast sensor, and saline purging system with simulated components
Ranked causes
- 1. Syringe valve sensor requires simulation and testing (common)
Select language and mode
- Test point
- Control panel display
- Expected
- English language selected, Cardiac mode for CVi
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe refill function
Verification that syringe refills correctly from 98-100 ml with contrast
Ranked causes
- 1. Refill system requires syringe valve sensor simulation and volume verification (common)
Simulate open syringe valve
- Test point
- Syringe valve sensor
- Expected
- Finger placed in sensor
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Checking the universal contrast sensor
Verify contrast sensor operation by measuring voltage at test point T76
Ranked causes
- 1. Contrast sensor malfunction (unknown)
Disconnect injector control panel cable
- Test point
- Injector head
- Expected
- Disconnected
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Download diagnostics log from ACIST injector system
Procedure to retrieve event data, error information and system use data from NVRam memory via serial connection to technician PC
Ranked causes
- 1. Routine diagnostic data retrieval for service analysis (common)
Prepare laptop and clear temporary directory
- Test point
- C:\ACIST\TEMP folder
- Expected
- Directory empty and accessible
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
End case error message
System continues to display 'To continue, disconnect and discard: 1.Saline and High Pressure Tubing 2.Manifold assembly' during End Case function after removal
Ranked causes
- 1. Air column detect sensor not functioning properly (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold sensor misalignment or not detecting
Sensor light does not illuminate when test fixture is positioned, or illuminates prematurely before proper positioning
Ranked causes
- 1. Sensor not properly aligned with manifold test gauge (common)
Position test fixture fully to the right side of slot
- Test point
- Manifold sensor slot
- Expected
- Sensor lights (red) inside barrel under potting
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold valve failed to close multiple times patient safety
Control panel displays 'MANIFOLD VALVE FAILED TO CLOSE X TIMES USING CURRENT SYRINGE. CONTRAST MAY BE CONTAMINATED.'
Ranked causes
- 1. Manifold valve stuck or sluggish (common)
Determine if syringe and contrast should be replaced
- Test point
- Alert message and procedure history
- Expected
- Operator decision: Yes to replace or No to continue
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold valve open caution
Manifold valve is in an improper position and not functioning correctly.
Ranked causes
- 1. Manifold valve repositioning needed (common)
Visual inspection of manifold valve
- Test point
- Manifold valve
- Expected
- Valve in closed position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peristaltic pump replacement caution
Complete removal and installation of peristaltic pump assembly including motor, adapter, and faceplate with wire reconnection
Ranked causes
- 1. Peristaltic pump failure or wear requiring complete assembly replacement (unknown)
Disconnect injector control panel cable before starting work
- Test point
- Control panel cable connector
- Expected
- Cable disconnected
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power loss or voltage abnormality causes system malfunction patient safety
System exhibits abnormal operation following power blackouts, brownouts, or voltage surges. Operator should immediately perform emergency shutdown to ensure patient safety.
Ranked causes
- 1. Power supply instability from external source (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power on and software initialization
Verification of power supply startup and successful software boot without self-test errors
Ranked causes
- 1. System requires initial power on and boot verification (common)
Turn on power switch
- Test point
- Power supply power switch
- Expected
- Power switch activated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage check failure high voltage
One or more DC output voltages at power supply connector are outside specified ranges.
Ranked causes
- 1. Power supply internal failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe loaded before prompt
Syringe was loaded prior to Load Syringe prompt appearing.
Ranked causes
- 1. Premature syringe loading (common)
Remove syringe and wait for prompt
- Test point
- System prompt display
- Expected
- Load Syringe prompt appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe valve interface cable damaged or not passing through housing
Cable connector contacts damaged, snap ring loose, cable jammed or not routed correctly through mounting chamber
Ranked causes
- 1. Syringe valve interface cable worn, corroded, or contact pins damaged (occasional)
Inspect connector for debris or oxidation
- Test point
- Syringe valve interface cable connector contacts
- Expected
- Clean, shiny metal contacts, no green oxidation or contrast buildup
- Tools
- Small probe or pick
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System alert message displayed
Control panel status window displays alert message indicating functional error.
Ranked causes
- 1. Functional error requiring diagnostic action (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Universal joint is stuck
Universal joint is very difficult to disengage due to dried contrast media in the locking knob area
Ranked causes
- 1. Dried contrast media in locking knob area (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AIR COLUMN DETECTED! | Air detected in high pressure tubing by air column detector | Air presence in high pressure tubing detected by sensor | warning |
Air Column Detected | Air column detect sensor has detected air in the tubing system. | Sensor detects air in patient tubing or contrast lines. | warning |
Check Syringe Valve | Syringe valve sensor has detected incorrect syringe valve position. | Sensor detects valve not in proper position or syringe loaded before Load Syringe prompt. | warning |
Contrast Empty | Contrast sensor is not detecting presence of contrast in spike tubing. | Sensor fails to detect contrast when expected or becomes undetected during operation. | warning |
CONTRAST EMPTY / NO CONTRAST | No contrast detected by universal contrast sensor | Contrast bottle empty or not detected by sensor | warning |
END CASE ERROR | End Case function unable to complete; message continues after kit removal | Air column detector malfunction or sensing error | error |
HAND CONTROLLER CALIBRATION FAILED | Hand controller calibration attempt failed | Incorrect button timing or connection issues during calibration | error |
Hand Controller Calibration Failed | HC calibration did not complete successfully. | Calibration procedure failed to accept button inputs. | error |
Hand Controller (HC) Calibration Failed | HC calibration did not complete successfully. | Calibration procedure interrupted or failed due to user action, button pressure, connection, or controller defect. | error |
HAND CONTROLLER REQUIRES CALIBRATION | Hand controller must be calibrated before use | System startup or hand controller changed/disconnected | info |
MANIFOLD VALVE CLOSED | Manifold valve stuck in closed position or not detected as open | Valve mechanical failure or sensor malfunction | error |
MANIFOLD VALVE FAILED TO CLOSE X TIMES | Manifold valve failed to close specified number of times during procedure | Valve sticking or sluggish performance during injection | warning |
Manifold Valve Open | Manifold valve is in an improper position. | Valve fails to seat properly or becomes unseated during operation. | warning |
Peristaltic Pump Over Torque | Saline pump pressure to dispense fluid exceeds internally set safety limit. | Pump motor encounters excessive back-pressure during saline dispensing. | error |
Pressure Limit Exceeded | System halted injection due to pressure exceeding set limits; flow rate limiting was not possible. | Injection pressure exceeded system limits during operation. | warning |
PRESSURE LIMIT EXCEEDED (1) | System pressure reached or exceeded limit setting during injection attempt | Resistance to flow exceeds pressure limit setting | warning |
Pressure Limit Exceeded (1) or (2) | System halted injection because on-the-fly flow rate limiting was not possible. Total injection volume not completed. | Measured injection pressure exceeded system safety limit despite flow rate limiting attempt. | critical |
Pressure Limit Exceeded (1) or (2) | Injection pressure exceeded system limits and flow rate limiting failed | System unable to limit flow rate on-the-fly; injection halted before completion | critical |
PRESSURE LIMIT EXCEEDED (2) | System pressure reached or exceeded limit setting on secondary channel/measurement | Resistance to flow exceeds pressure limit setting | warning |
STANDBY | System in standby mode; operation suspended | Standby button pressed or communication loss with injector head | info |
Standby | Standby button depressed or control panel/injector interface cable disconnected. | Standby button pressed or cable connection lost. | info |
Standby | Standby function activated | Standby button pressed or control panel cable disconnected | info |
SYRINGE VALVE OPEN/CLOSED | Syringe valve sensor malfunction or disconnection detected | Syringe valve sensor unable to detect valve position | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.