FR FieldReady Academy
Infusion pumps

Abbott GemStar: troubleshooting and error codes

The ServiceForge knowledge base holds 30 documented symptoms for the Abbott GemStar, with 84 ranked causes, 71 diagnostic checks with test points and expected values, and 188 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Infuser fails self test at power-on patient safety

Device alarms during initialization mode; one or more internal tests (RAM, ROM, motor, keys, watchdog, CONFIG, battery, power loss, CRC, calibration) does not complete successfully

Ranked causes

  1. 1. RAM memory defect (rare)
  2. 2. ROM checksum error or corrupted firmware (rare)
  3. 3. Motor control malfunction (occasional)
  4. 4. Stuck key detection during self test (occasional)
  5. 5. PIC watchdog timer failure (rare)
  6. 6. CONFIG register corruption (rare)
  7. 7. Battery voltage out of range (occasional)
  8. 8. Power loss completion circuit malfunction (rare)
  9. 9. EEPROM calibration data corruption (occasional)
First check

Observe initialization mode message and alarm indication

Test point
Infuser display
Expected
Message indicating self tests being performed; alarm if test fails within 30 seconds

1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Docking station does not function high voltage

Infuser does not lock in docking station, LED does not illuminate, device does not power on when docked, or continuity test fails

Ranked causes

  1. 1. Dovetail mechanism damaged (occasional)
  2. 2. Detent release mechanism stuck or broken (occasional)
  3. 3. 3 VDC/printer/bolus port connector defective (occasional)
  4. 4. Bolus/printer pass-through connector defective (occasional)
  5. 5. Pole clamp loose or damaged (occasional)
  6. 6. Physical damage to docking station housing (occasional)
  7. 7. AC mains power cord or plug defective (occasional)
  8. 8. Docking station LED or power circuit failure (rare)
First check

Inspect dovetail mechanism

Test point
Dovetail rail on docking station
Expected
In good condition, no bending or damage

10 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rechargeable battery pack does not function

Battery pack does not slide into place, does not lock firmly, green LED does not flash, latch mechanism does not release, or charger LED does not illuminate

Ranked causes

  1. 1. 3 VDC/printer/bolus port connector defective or misaligned (occasional)
  2. 2. Pass-through port connector defective or misaligned (occasional)
  3. 3. Battery pack mechanical locking mechanism failure (occasional)
  4. 4. Green LED or LED driver circuit failure (rare)
  5. 5. Battery pack charger LED failure (rare)
First check

Inspect 3 VDC/printer/bolus port connector contacts

Test point
Connector pins on battery pack and infuser
Expected
Contacts present and in good condition

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote bolus cord and switch malfunction

Bolus cord frayed or torn, switch body damaged or shows fluid ingress, connector pins missing or corroded, continuity test fails, or cord does not lock when connected

Ranked causes

  1. 1. Bolus cord insulation damaged (occasional)
  2. 2. Switch body damaged or fluid ingress (occasional)
  3. 3. Bolus connector pins defective (occasional)
  4. 4. Internal wire break in bolus cord (occasional)
  5. 5. Infuser bolus port connector loose or damaged (occasional)
First check

Inspect bolus cord insulation

Test point
Cord jacket entire length
Expected
Free from fraying and torn insulation

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery door degradation

O-ring missing or deteriorated, door latch/plunger hinges fractured, spring loose or collapsed, wiper contact contaminated or corroded

Ranked causes

  1. 1. O-ring wear or degradation (common)
  2. 2. Door latch or hinge fracture (occasional)
  3. 3. Spring mechanism failure (occasional)
  4. 4. Wiper contact contamination or corrosion (common)
First check

Inspect o-ring presence and condition

Test point
Battery door o-ring
Expected
Present and undamaged

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Cassette alarm with pressure reading above 150 patient safety

High upstream pressure (above 150) from syringe injection, user pressure during delivery, or elevated container height.

Ranked causes

  1. 1. Fluid injected via upstream Y-site with backcheck valve (common)
  2. 2. User pressed syringe during delivery (common)
  3. 3. Container elevated more than 2 feet with high pump rate (common)
  4. 4. Sticky buildup in cassette pressure sensing hole (occasional)
First check

Read upstream pressure value

Test point
Pressure sensor output
Expected
above 150 ADC units, >150

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distal Occlusion alarm patient safety

Outlet pressure too high, restricting delivery flow from clamped line, manifold valve closure, patient line restriction, or low occlusion setting.

Ranked causes

  1. 1. Outlet tube clamped or manifold valve closed (common)
  2. 2. Patient line restricted by kinks or clots (common)
  3. 3. Distal occlusion setting too low for fluid or equipment (occasional)
  4. 4. Pumping into elevated patient (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC adaptor LED does not illuminate on power high voltage

Green power LED fails to light when AC adaptor connected to 3-volt connector

Ranked causes

  1. 1. AC adaptor not delivering power (common)
  2. 2. Defective LED or LED driver circuit (occasional)
  3. 3. Loose or damaged AC adaptor plug connection (occasional)
First check

Measure AC adaptor output voltage

Test point
AC adaptor connector pins
Expected
3 VDC ± 5%, ±5%
Tools
Multimeter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-Line alarm with readings 1–5 patient safety

Very low sensor readings indicate air in line or sensor damage; zero reading indicates air or broken sensor.

Ranked causes

  1. 1. Air in the infusion line (common)
  2. 2. Distal tube pulled excessively hard (occasional)
  3. 3. Broken air sensor (occasional)
First check

Observe air sensor reading

Test point
Air sensor output
Expected
1–5 range or zero ADC units, 1–5 or 0

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-Line alarm with readings 50–80 patient safety

Air sensor reading between 50 and 80 indicates bubbles near cassette outlet, partially pulled distal tube, or wet tubing with air.

Ranked causes

  1. 1. Buildup of fine bubbles near cassette outlet (common)
  2. 2. Distal tube partially pulled from air sensor pincers (occasional)
  3. 3. Air in tube with wet exterior (occasional)
First check

Observe air sensor reading value

Test point
Air sensor output
Expected
50–80 range ADC units, 50–80

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power loss alarm not functioning patient safety

Red LED does not illuminate, beeper does not sound, POWER LOSS does not display when battery door removed during battery operation

Ranked causes

  1. 1. Red LED failure (occasional)
  2. 2. Beeper/speaker failure (occasional)
  3. 3. Power loss detection circuit failure (rare)
First check

Install batteries, power on, then remove battery door

Test point
Red LED and beeper during door removal
Expected
Red LED illuminates and beeper sounds simultaneously

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal Occlusion alarm patient safety

Inlet pressure too low to allow delivery, preventing infusion flow.

Ranked causes

  1. 1. Inlet tubing clamped (common)
  2. 2. Ball check or burette set valve closed (common)
  3. 3. Syringe stuck in line (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on

Infuser fails to power on with fresh AA batteries, UNIT SELF-TEST does not display, audible alarm does not sound

Ranked causes

  1. 1. Dead or missing batteries (common)
  2. 2. Faulty on/off switch (rare)
  3. 3. Power supply circuit failure (rare)
First check

Install fresh disposable AA batteries

Test point
Battery compartment
Expected
Two fresh AA batteries installed

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC adaptor connection problem high voltage

AC adaptor plug not properly engaged or retained at infuser 3-volt connector, or infuser does not power on when connected

Ranked causes

  1. 1. AC adaptor plug connector loose or misaligned (common)
  2. 2. Infuser 3-volt connector damaged or corroded (occasional)
First check

Connect AC adaptor and verify engagement

Test point
AC adaptor plug at infuser 3-volt connector
Expected
Plug properly engaged and retained

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC adaptor output voltage out of specification high voltage

AC adaptor does not provide 3 VDC ± 5% output voltage

Ranked causes

  1. 1. Transformer or voltage regulation circuit failure (common)
  2. 2. Damaged power cord affecting internal circuit (occasional)
First check

Measure AC adaptor output voltage with multimeter

Test point
AC adaptor connector output pins
Expected
3 VDC ± 5%, ±5%
Tools
Multimeter

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-line test fails

Air-in-line alarm does not occur within one minute when proximal end is removed from reservoir, indicating air detection malfunction.

Ranked causes

  1. 1. Air detection sensor contaminated or blocked (common)
  2. 2. Air detection sensor circuit malfunction (occasional)
First check

Release any clamps from previous tests

Test point
Infuser set
Expected
Set is unobstructed

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery compartment degradation

Circuit board wiper contact contaminated or corroded, battery cap missing or degraded

Ranked causes

  1. 1. Wiper contact contamination or corrosion (common)
  2. 2. Battery cap missing or degraded (occasional)
First check

Inspect circuit board wiper contact

Test point
Circuit board wiper contact
Expected
Present and free of contamination, corrosion, and deformation

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Cassette A alarm with reading above 180 caution

High air sensor reading (above 180) from flattened tubing due to age/temperature or wet tubing exterior.

Ranked causes

  1. 1. Tubing flattened by long use or high temperature (common)
  2. 2. Wet tubing exterior affecting sensor reading (occasional)
First check

Read air sensor value

Test point
Air sensor output
Expected
above 180 ADC units, >180

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distal occlusion test fails

Occlusion alarm does not occur within 30 seconds when set is clamped distally, indicating occlusion detection failure.

Ranked causes

  1. 1. Distal occlusion sensor contaminated or blocked (common)
  2. 2. Distal occlusion sensor circuit malfunction (occasional)
First check

Clamp infuser set distally

Test point
Infuser set 10-12 inches below cassette
Expected
Set clamped, no flow possible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED test fails

Green power LED or red alarm LED does not illuminate, indicating LED or circuit malfunction.

Ranked causes

  1. 1. Power LED not illuminating (occasional)
  2. 2. Alarm LED not illuminating (occasional)
First check

Power LED test: Verify green power LED illuminates

Test point
Green power LED indicator
Expected
Green power LED illuminates

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Loss alarm

Infuser has experienced a power loss event initiated by hardware or software.

Ranked causes

  1. 1. Hardware-initiated power loss (H) (unknown)
  2. 2. Software-initiated power loss (S) (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal occlusion test fails

Occlusion alarm does not occur within 30 seconds when set is clamped above cassette, indicating proximal occlusion sensor malfunction.

Ranked causes

  1. 1. Proximal occlusion sensor contaminated or blocked (common)
  2. 2. Proximal occlusion sensor circuit malfunction (occasional)
First check

Clamp infuser set above cassette

Test point
Infuser set proximal to cassette
Expected
Set clamped, no flow possible above cassette

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Volume accuracy test fails

Infuser does not deliver between 19 and 21 mL within three minutes, indicating delivery mechanism malfunction.

Ranked causes

  1. 1. Infuser delivers less than 19 mL (occasional)
  2. 2. Infuser delivers more than 21 mL (occasional)
First check

Load primed infuser set and position in water reservoir

Test point
Infuser set proximal end in 50 mL water reservoir
Expected
Set properly loaded and submerged in water

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifacts on ECG, EMG, or EEG monitoring patient safety

Electrical noise and artifacts appear on voltage-sensing monitoring equipment such as ECG, EMG, or EEG machines during fluid infusion. Artifacts vary at a rate associated with the infusion rate.

Ranked causes

  1. 1. Electronic noise generated by the infuser (common)
First check

Stop infusion and observe monitoring display

Test point
ECG, EMG, or EEG display
Expected
Artifact disappears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Cassette alarm with pressure reading below 90 caution

Low upstream pressure (below 90) indicates cassette not fully latched at inlet end.

Ranked causes

  1. 1. Cassette fully or partly unlatched at inlet (common)
First check

Read upstream pressure value

Test point
Pressure sensor output
Expected
below 90 ADC units, <90

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Cassette D alarm caution

Distal pressure sensor detects cassette not properly latched at outlet end.

Ranked causes

  1. 1. Cassette not fully snapped in at outlet end (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test fails

Display does not fill with solid boxes during display test, indicating LCD malfunction.

Ranked causes

  1. 1. Display malfunction or LCD failure (rare)
First check

Display test: Verify display fills with solid boxes

Test point
LCD display screen
Expected
Display fills with solid boxes for ten seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low Batteries alarm

Battery voltage is below acceptable level; continued operation is limited.

Ranked causes

  1. 1. Battery discharge below safe operating level (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operation test procedure required patient safety

Scheduled maintenance test to verify infuser operation. Performed in STOP mode, approximately 15 minutes, must be unlocked. Minimum once every 12 months per Hospira recommendation.

Ranked causes

  1. 1. Routine maintenance and verification (common)
First check

Power test: Verify power sources

Test point
Display, batteries, external power
Expected
Display indicates both external power and batteries connected

8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Switch to Batteries alarm

Main power has been lost; infuser is operating on battery power.

Ranked causes

  1. 1. AC mains power failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01/000CONFIG register contents are not what is expectedCONFIG register tested at power onerror
02/000Stack errorStack corruption check during power on or continuouslyerror
03/000Invalid interruptUnsupported ISR callederror
03/001Interrupt overlapInterrupt service routine called before completionerror
03/002Invalid IRQ interruptIRQ interrupt called with no interrupts processederror
04/000RAM Test errorRead/write test patterns 0XAA and 0X55 failerror
05/000EEPROM write error - data read back does not match data writtenerror
05/000 - EEPROM_WR_ERRORCPU EEPROM write verification failedData written to EEPROM does not match data read backerror
05/001EEPROM write error - invalid EEPROM addresserror
05/001 - EEPROM_WR_ERRORInvalid EEPROM address errorEEPROM write function receives invalid addresserror
06/000Pump configuration CRC errorerror
06/000 - CRC_CHECKSUM_ERROR (PUMP_CONFIG_CRC)Pump configuration data CRC mismatchConfiguration CRC verification fails before operationserror
06/001Infusion data CRC errorerror
06/001 - CRC_CHECKSUM_ERROR (INFUSION_DATA_CRC)Infusion data integrity CRC mismatchInfusion data CRC verification fails at startup or runtimeerror
06/002Pump data CRC errorerror
06/002 - CRC_CHECKSUM_ERROR (PUMP_DATA_CRC)Infuser data integrity CRC mismatchInfuser data CRC verification fails at startup or runtimeerror
06/003Program data CRC errorerror
06/003 - CRC_CHECKSUM_ERROR (PROGRAM_CRC)Infuser program data CRC mismatchProgram CRC verification fails during mode transitionserror
06/004Speed protocol CRC errorerror
06/004 - CRC_CHECKSUM_ERROR (SPEED_PROTOCOL_CRC)Speed protocol data CRC mismatchSpeed protocol CRC fails during protocol operationserror
06/005ROM CRC errorerror
06/005 - CRC_CHECKSUM_ERROR (ROM_CRC)Flash RAM firmware integrity verification failedFlash RAM checksum verification fails at power-on or hourlycritical
06/006Protected variable errorerror
06/006 - CRC_CHECKSUM_ERROR (PROTECTED_VAR)Critical air/pressure sensing variable corruptionMirrored RAM data does not match original during verificationcritical
06/007Dose data CRC errorerror
06/007 - CRC_CHECKSUM_ERROR (DOSE_DATA_CRC)Dose data integrity CRC mismatchDose data CRC fails at startup or during dose operationerror
06/008Air calibration CRC errorerror
06/008 - CRC_CHECKSUM_ERROR (AIR_CAL_CRC)Air calibration data CRC mismatchAir calibration CRC fails at power-on or background checkerror
06/009Pressure calibration CRC errorerror
06/009 - CRC_CHECKSUM_ERROR (PRESS_CAL_CRC)Pressure calibration data CRC mismatchPressure calibration CRC fails at power-on or background checkerror
06/010Motor calibration CRC errorerror
06/010 - CRC_CHECKSUM_ERROR (MOTOR_CAL_CRC)Motor calibration data CRC mismatchMotor calibration CRC fails at power-on or delivery starterror
06/011Settings CRC errorerror
06/011 - CRC_CHECKSUM_ERROR (SETTINGS_CRC)Infuser settings data corruptionSettings CRC fails when entering function or mode changeerror
07/000High air sensor valueAir detection sensor reading higherror
07/000 - AIR_SENSOR_ERROR (HIGH_VALUE)Air sensor pre-trigger value exceeds calibration baselineMeasured air sensor value greater than calibrated levelerror
07/001Negative volume samplederror
07/001 - AIR_SENSOR_ERROR (AIR_NEGATIVE_VOLUME)Negative volume detected between delivery checkpointsCalculated volume between checkpoints is negativeerror
07/002Bad air sensor stateerror
07/002 - AIR_SENSOR_ERROR (AIR_BAD_STATE)Air sensor state variable invalid or unrecognizedState variable is not a recognized air sensor stateerror
07/003Bad air sensor eventerror
07/003 - AIR_SENSOR_ERROR (AIR_BAD_EVENT)Air sensor received unrecognized event codeEvent code is not in recognized air sensor event listerror
07/004Air sensor not calibratederror
07/004 - AIR_SENSOR_ERROR (AIR_NOT_CALIBRATED)Air sensor calibration values missing or invalidAIR MAX PRE LEVEL is zero at power-onerror
07/005Excessive volume samplederror
07/005 - AIR_SENSOR_ERROR (AIR_EXCESSIVE_VOLUME)Air volume between checkpoints exceeds 100 microlitersCalculated volume between checkpoints exceeds thresholderror
08/005Bad pressure sensor eventerror
08/005 - PRESSURE_SENSOR_ERROR (PRESS_BAD_EVENT)Pressure sensor received invalid event codeReceived event not in valid pressure sensor event listerror
08/006Bad pressure sensor stateerror
08/006 - PRESSURE_SENSOR_ERROR (PRESS_BAD_STATE)Pressure sensor state variable invalidState value not defined in pressure sensor processingerror
08/007Distal pressure is out of rangeerror
08/007 - PRESSURE_SENSOR_ERROR (PRESS_DISTAL_RANGE)Distal occlusion threshold out of valid 0-255 rangeRecomputed distal threshold after plunger home is out of rangeerror
08/008Proximal pressure is out of rangeerror
08/008 - PRESSURE_SENSOR_ERROR (PRESS_PROX_RANGE)Proximal occlusion threshold out of valid 0-255 rangeRecomputed proximal threshold during delivery start is out of rangeerror
09/000Short term overdeliveryerror
09/000/MOTOR_IPRF_ENC_OS_SYNCMotor encoder position synchronization error during IPRF deliveryMicroprocessor misses motor encoder counts during IPRF stepping, causing plunger overadvanceerror
09/001Backward motor movementerror
09/001/MOTOR_BACKWARD_MOVEMENTExcessive backward motor movement detectedType 1: Single rollback >40 counts. Type 2: Accumulated rollbacks >16 counts exceed 40 within 16-stroke windowerror
09/002Motor not calibratederror
09/002/MOTOR_NOT_CALIBRATEDMotor calibration values out of specificationAt power-on, calibration values tested; PWM value for new rate falls below minimum set during calibrationerror
09/006Power on motor testerror
09/006/NO_MOTOR_CONTROLPower-on motor control test failedAt power-on: motor does not turn when switch FET off, or motor does not turn when FET onerror
09/007Motor not turning when it should be turningerror
09/007/MOTOR_NOT_PUMPINGNo motor movement when expected during operationPWM value reaches 200 with no encoder counts detected during ramp-up, homing, stepping, or continuous deliveryerror
09/008Bad motor stateerror
09/008/MOTOR_BAD_STATEMotor state variable out of valid rangeMotor state variable checked during operation; value outside acceptable rangeerror
09/009Bad motor eventerror
09/009Motor error during initial calculation, normal operation, or rate adjustmentPWM_LOW_COUNTER value exceeds 10 during testingerror
09/009/MOTOR_BAD_EVENTInvalid motor event received for current stateEvent processing not allowed within current motor stateerror
09/010Bad motor rateerror
09/010/MOTOR_BAD_RATEInvalid infusion rate parameter receivedWhen start command received, rate verified; value outside acceptable range (e.g., IPRF <0.1 mL/hr, continuous >1000 mL/hr, invalid taper parameters)error
09/011Motor stack errorerror
09/011/MOTOR_STACK_ERRORMotor control stack processing errorStack full when pushing state; stack empty when popping; invalid stack pointer during push or pop operationerror
09/012Motor step overlaperror
09/012/MOTOR_STEP_OVERLAPMotor step did not complete before next step requestedTimer expires during IPRF homing (state 15) or stepping (state 35) without step completionerror
09/013Motor rate control errorInvalid motor control during rate adjustmenterror
09/014Motor performing invalid step numberStep number out of expected range for delivery rate at end of IPRF steperror
09/015Motor overshoot of stopping positionTwo consecutive motor steps overshoot stopping position by 2+ pointserror
09/016Motor encoder overflow not clearedForward motor counts exceed 432 + 80 + 8*N per output shaft revolutionerror
09/017Motor received invalid restart commandMotor ISR receives restart command when current rate is zeroerror
09/018Motor over-delivery detectedOutput shaft turns exceed expected by more than 25% every 5 secondserror
09/019Motor under-delivery detectedOutput shaft turns fall more than 25% below expected every 5 secondserror
09/020Motor takeoff detected during stopped deliveryOutput shaft turns exceed limit (current turns + 3) while delivery not in progresserror
09/020Motor runawayerror
10/000Beeper errorerror
10/000Beeper not functioning or beeper check circuit malfunctioningIncorrect digital signal at power on or during non-infusion safety taskwarning
11/0005V line overvoltage detected5V line voltage measured greater than expected at power on or once per seconderror
11/000More than 5.5 volts measured on the 5 V lineerror
11/0015V line undervoltage detected5V line voltage measured less than expected at power on or once per seconderror
11/001Less than 4.5 volts measured on the 5 V lineerror
11/002Battery voltage input overvoltage detectedAA battery input voltage exceeds expected at power on or once per secondwarning
11/002More than 3.2 volts measured on the AA battery voltage inputerror
11/003External power input overvoltage detectedExternal power input voltage exceeds expected at power on or once per secondwarning
11/003More than 3.6 volts measured on the external voltage inputerror
11/004Less than 2.0 volts measured on the lithium battery inputerror
11/004Lithium battery discharged or shortedLithium battery input voltage less than expected at power on or once per secondwarning
12/000Stuck keypad key detectedKey remains active for more than one minute at power on or during operationwarning
13/000IRQ test of oscillator - timing errorerror
13-000TIME_BASE_ERROR: Number of RTIs counted is not within expected rangeTiming integrity check failure once per seconderror
13/001RTI test of one second interrupt - timing errorerror
13-001TIME_BASE_ERROR: Number of RTIs exceeds upper limit since last interruptRTI count exceeds upper limit valueerror
14-001WATCHDOG_ERROR: Watchdog PIC circuit not functioningMotor moves more than 20 counts during PIC test at power onerror
14-003WATCHDOG_ERROR: Periodic task not updating counterCounter unchanged for 20 consecutive checkserror
15/000Power down errorerror
15-N/APOWER_DOWN_ERROR: Abnormal system state at power onSystem state indicates incomplete power-down processingerror
16/001Air-in-line was active when it should not be activeerror
16-001-INVALID_STATESOFTWARE_ERROR: Air-in-line alarm active during start commandAir-in-line alarm status already active when start/restart receivederror
16/002Key event timeouterror
16-002-KEY_EVENT_TIMEOUTSOFTWARE_ERROR: Callback functions failed to acknowledge key eventNo acknowledgement within 6 seconds of key eventerror
16/003Invalid alarm semaphoreerror
16-003-INVALID_ALARM_SEMAPHORESOFTWARE_ERROR: Invalid alarm semaphore value receivedAlarm task receives undefined semaphore valueerror
16/004Invalid alarm messageerror
16-004-INVALID_ALARM_MESSAGESOFTWARE_ERROR: Invalid alarm action requestedAlarm task receives undefined action value in messageerror
16/005Invalid alarm typeerror
16-005-INVALID_ALARM_TYPESOFTWARE_ERROR: Invalid alarm type receivedAlarm task receives invalid alarm type in message or history requesterror
16/006Invalid alarm callbackerror
16-006-INVALID_ALARM_CALLBACKSOFTWARE_ERROR: Invalid alarm callback IDAlarm task receives invalid callback ID during key presserror
16/007Invalid sound typeerror
16-007-INVALID_SOUND_TYPESOFTWARE_ERROR: Invalid sound type requestedAudible alarm functions receive undefined sound type valueerror
16/008IQUEUE fullerror
16-008-IQUEUE_FULLSOFTWARE_ERROR: Queue full during add requestQueue processing receives add data request when queue is fullerror
16/009Infusion safety task received an invalid message typeerror
16-009-ISA_INVALID_MSGSOFTWARE_ERROR: Infusion safety task received unrecognizable message codeInfusion safety task receives invalid message typeerror
16/010Infusion safety task received a null message typeerror
16-010-ISA_NULL_MSGSOFTWARE_ERROR: Infusion safety task received NULL messageRTXC passes NULL message to infusion safety taskerror
16/011Infusion safety task received an invalid pressure messageerror
16-011-ISA_INVALID_PRESSURESOFTWARE_ERROR: Infusion safety task received invalid pressure messagePressure message type code unrecognizederror
16/012Infusion safety task did not receive the expected queue informationerror
16-012-ISA_EMPTY_IQUEUESOFTWARE_ERROR: Expected queue information not received by infusion safety taskAir and pressure queues empty or check cassette message received without type valueerror
16/013Infusion safety task received an invalid semaphore valueerror
16-013-ISA_BAD_SEMAPHORESOFTWARE_ERROR: Infusion safety task received invalid semaphore valueSemaphore value undefined for infusion safety taskerror
16/014Infusion safety task received an invalid air messageerror
16-014-ISA_INVALID_AIRSOFTWARE_ERROR: Infusion safety task received invalid air messageAir message data value undefinederror
16/015Infusion safety task received an invalid check cassette messageerror
16-015-ISA_BAD_CK_CASS_CODESOFTWARE_ERROR: Infusion safety task received invalid check cassette messageCheck cassette message data undefinederror
16/016Infusion safety task received a resume without an initial starterror
16-016-ISA_BAD_RESUMESOFTWARE_ERROR: RESUME message received without prior STARTResume message without preceding start messageerror
16/017Infusion safety task detects a rate mismatcherror
16-017-ISA_RATE_MISMATCHSOFTWARE_ERROR: Infusion rate mismatch between tasksRate mismatch between infusion task and safety task for 8 consecutive sampleserror
16/018Infusion safety task detects a mode mismatcherror
16-018-ISA_MODE_MISMATCHSOFTWARE_ERROR: Main mode and infusing flag inconsistentGLOBAL_MAIN_MODE and PUMP_INFUSING variables not consistent for 3 successive sampleserror
16/019IED has message conflicterror
16/020Remote queue out fullerror
16-020-REMOTEQ_OUT_FULLSOFTWARE_ERROR: Remote output message queue fullNeed to store message at end of full outgoing queueerror
16/021Remote queue out emptyerror
16-021-REMOTEQ_OUT_EMPTYSOFTWARE_ERROR: Remote output message queue empty when access attemptedSCI_OUTPUT task attempts send or free from empty queueerror
16/022Remote queue out bad stateerror
16-022-REMOTEQ_OUT_BAD_STATESOFTWARE_ERROR: Remote output message queue in bad stateNeed to store message when queue is in bad stateerror
16/023ISA bad rateerror
16-023-ISA_BAD_RATESOFTWARE_ERROR: Infusion rate exceeds safe limitTapered rate > 450 mL/hr or non-tapered rate > 1000 mL/hrerror
16-024-INVALID_TIME_IN_SECONDSSOFTWARE_ERROR: Infuser clock rollover year 2004 to 2005Infuser time rolls over from 2004 to 2005error
18/000History pointer error while inserting a new record into the history logerror
18-000-WRITE_HLOG_ERRORHISTORY_PTR_ERROR: History pointers out of accepted limitsHistory pointers fail validation when record writtenerror
18/001History pointer error while traversing to bottom of historyerror
18-001-TRAVERSING_TO_BOTTOM_ERRORHISTORY_PTR_ERROR: History traversal pointer error going to bottomTop and bottom pointers out of range during GET_PREV_HRECORD callerror
18/002History pointer error while traversing to top of historyerror
18-002-TRAVERSING_TO_TOP_ERRORHISTORY_PTR_ERROR: History traversal pointer error going to topTop and bottom pointers out of range during GET_NEXT_HRECORD callerror
21/000Remote communication input buffer errorerror
21-000-000-MAX_OVERRUNS_EXCEEDEDCOMM_ERROR: SCI buffer overrun exceeded maximumSCI interrupt attempts to add byte to full data buffererror
21-000-001-REMOTE_INPUT_QUEUE_EMPTYCOMM_ERROR: SCI input task accessing empty message queueSCI input task attempts extract from empty incoming message queueerror
21/001Keypad queue errorerror
21/002Remote communication input message errorerror
21-002-004-BAD_KEY_MSGCOMM_ERROR: SCI input task received unknown key eventSCI input task receives unrecognized key eventerror
21-002-005-BAD_MSG_TYPECOMM_ERROR: SCI input task received unknown message typeSCI input task receives unrecognized message typeerror
21-002-006-BAD_SESS_RQSTCOMM_ERROR: SCI input task received bad session requestSCI input task receives invalid session request messageerror
Air-in-LineAir detected in infusion line by air sensorSensor reading 1–5, 50–80, or indicates air presencewarning
ALMOST EMPTYContainer approaching emptyContainer fluid level drops below thresholdinfo
CHANGE BATTERIESBattery replacement required soonBattery voltage drops below usable thresholdwarning
CHECK CASSETTECassette malfunction or incompatibilityCassette not detected, not compatible, or mechanical faulterror
Check Cassette AHigh air sensor reading above 180Air sensor reading exceeds 180 ADC unitswarning
Check Cassette DCassette not properly latched at outlet endDistal pressure sensor detects improper latchwarning
Check Cassette PHigh upstream pressure above 150Pressure reading exceeds 150 ADC unitswarning
CHECK PRINTERPrinter malfunction or unavailablePrinter not responding or communication errorwarning
DISTAL OCCLUSIONInfusion line occluded downstream of pumpPressure exceeds distal occlusion thresholdwarning
Distal OcclusionOutlet pressure too high, blocking deliveryOutlet pressure exceeds safe operating levelerror
EMPTY CONTAINERContainer is emptyContainer fluid level reaches empty thresholdwarning
FLASHING DISPLAYLCD display malfunctionDisplay not rendering normallywarning
Low BatteriesBattery voltage below safe operating levelBattery charge falls below acceptable thresholdcritical
POWER LOSSAC power disconnected or failedAC power supply disconnected or no AC voltage detectedwarning
Power Loss HHardware-initiated power loss eventHardware power control circuits detected power losserror
Power Loss SSoftware-initiated power loss eventSoftware power management logic initiated shutdownerror
PROGRAM INCOMPLETEInfusion program not fully enteredUser attempts to start infusion with incomplete program parameterswarning
PROXIMAL OCCLUSIONInfusion line occluded upstream of pumpPressure exceeds proximal occlusion thresholdwarning
Proximal OcclusionInlet pressure insufficient for deliveryInlet pressure too low to allow infusion flowerror
Self test failure alarmOne or more of the nine self test routines did not complete successfullyInitiated automatically at power-on during approximately 30-second initialization modecritical
STARTGeneric alert messageDevice state or condition requires operator attention at startupinfo
Switch to BatteriesMain AC power lost; infuser operating on batteryLoss of AC mains powerwarning
USING BATTERIESOperating on battery powerAC power disconnected, device switched to or operating on batteryinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.