How Radiation Harms: Ionization, Dose Quantities and Units, Occupational and Public Limits
Unit 2: Radiation, Magnetic, and Facility Safety. Lesson 6 of 60 in CRES Academy.
Every protective feature you maintain exists because x-rays ionize tissue. This lesson is the vocabulary and the numbers behind all of it, and it is where the unit-conversion questions come from.
Ionization and what it damages
An x-ray photon with enough energy knocks an electron off an atom, leaving an ion pair. That is ionization, and it is the difference between x-rays and the radio waves a wireless access point puts out. X-rays and gamma rays are electromagnetic radiation; alpha particles, beta particles, and neutrons are particulate radiation, which carries mass, gives up its energy over a very short distance, and is what a nuclear medicine source emits inside the patient. Damage happens two ways: the direct effect is a photon hitting a critical molecule such as DNA, and the indirect effect is a photon splitting water into reactive fragments that then attack the molecule. Because the body is mostly water, the indirect effect does most of the harm.
- Somatic damage is to the exposed person's own tissues; genetic damage is to the reproductive cells and so to descendants.
- Dividing cells are the most sensitive, which is why bone marrow, the gonads, and children matter most.
- Dose rate matters: the same total dose delivered slowly does less harm than the same dose all at once, because repair has time to work.
Two kinds of effect sit behind every limit in this course. Stochastic effects, such as cancer and genetic change, have no threshold: any dose carries some probability of them, and the probability rises with dose while the severity does not. Deterministic effects, such as skin reddening, hair loss, and cataract, need a threshold dose before they appear at all, and above it they get worse as dose rises. Occupational and public limits exist to keep stochastic risk small; the dose alerts on a long fluoroscopic case exist to keep deterministic injury from happening at all.
Quantities and units
Exposure, air kerma roentgen (R); 1 R = 2.58 x 10 to the minus 4 C/kg; air kerma in gray is the SI form Absorbed dose gray (Gy) or rad; 1 Gy = 100 rad Equivalent dose sievert (Sv) or rem; 1 Sv = 100 rem Effective dose sievert, weighted for radiation type and tissue sensitivity Prefixes milli is a thousandth, micro a millionth; 1 mSv = 1000 microsieverts
Work the conversions both ways. 5 rem is 50 mSv. 0.2 mSv is 20 mrem. 2 Gy is 200 rad.
Where background dose comes from
Natural background is about 3 mSv per year, radon being the largest part, and the largest artificial source for the population is medical imaging. One chest image is about 0.02 mSv, so a staff member's annual dose is best understood against background rather than against zero.
The limits
Occupational effective dose 50 mSv per year, cumulative 10 mSv x age in years Lens of the eye 150 mSv per year Thyroid, skin, hands, and feet 500 mSv per year Public 1 mSv per year frequent, 5 mSv infrequent Embryo or fetus 5 mSv over the gestation, 0.5 mSv in any month once declared Education and training under 18 1 mSv per year
A worked example: a badge reporting 0.4 mSv in a quarter projects to about 1.6 mSv a year, which is a few percent of the occupational limit and is a normal result for a radiographer; the same reading on a badge that never leaves the booth says something is wrong with the badge or the room.
Personnel monitoring
Badges are thermoluminescent or optically stimulated detectors read by a service, usually quarterly, and worn at the collar outside the apron so they read what reaches the unprotected neck and head. A thermoluminescent badge stores the energy it absorbs in a crystal and gives it back as light when the reader heats it, which is thermoluminescence; the properties of the crystal are the reason it is used, because lithium fluoride absorbs radiation much the way tissue does. In pregnancy a second badge goes at the waist under the apron. The reports go to the radiation safety officer and the committee, who look for trends, for a badge left in the room, and for anyone approaching an investigation level. If you move a badge rack or leave your own badge on a cabinet, you have corrupted the record.
How Radiation Harms: Ionization, Dose Quantities and Units, Occupational and Public Limits, explained slowly
Ionization is the whole reason the rules exist
A radio transmitter and an x-ray tube both put out electromagnetic energy. The difference is that a single x-ray photon carries enough energy to strip an electron from an atom, leaving an ion pair and a chemically reactive mess behind. That is ionization, and it is why there are dose limits for one and not the other.
- The direct effect is a photon striking a critical molecule, above all DNA.
- The indirect effect is a photon splitting water into reactive fragments that then attack the molecule. Because the body is mostly water, this route does most of the damage.
- Somatic damage is to the exposed person. Genetic damage is to reproductive cells, and so to descendants.
- Dividing cells are most sensitive, which ranks bone marrow, the gonads, and children at the top of the risk list.
- Dose rate matters: the same total dose given slowly does less harm than the same dose at once, because repair mechanisms have time to work. That is the biological reason pulsed fluoroscopy is more than a dose arithmetic trick.
Two kinds of effect, and which limit answers which
Every number in the limit table exists to hold down one of two very different risks, and the exam expects the two words.
Stochastic effects cancer and genetic change; no threshold, so any dose carries some
probability; more dose raises the chance, not the severity
Deterministic effects skin reddening, hair loss, cataract, reduced fertility; a threshold
must be passed first, and above it severity rises with doseOccupational and public limits are written against stochastic risk, which is why they are low, annual, and cumulative. The alerts and trigger levels on a long interventional case are written against deterministic injury, which is why they are about one patch of skin during one procedure rather than a yearly total. A technician who keeps those apart can answer the common exam trap: a dose limit protects a worker over a career, while a skin dose trigger protects one patient during one case.
The quantities, in the order a question walks them
Quantity Asks Units and conversions
Exposure, how much ionization in air roentgen (R); 1 R = 2.58 x 10 to the
air kerma minus 4 C/kg; air kerma in gray is the SI form
Absorbed dose how much energy per kg of tissue gray (Gy), or rad; 1 Gy = 100 rad
Equivalent dose absorbed dose weighted for how sievert (Sv), or rem; 1 Sv = 100 rem
damaging that radiation type is
Effective dose equivalent dose weighted for which sievert
tissues were hit
Prefixes milli a thousandth, micro a 1 mSv = 1000 microsieverts = 100 mrem
millionthPractice the conversions in both directions until they are reflexes: 5 rem is 50 mSv; 50 mSv is 5 rem; 0.2 mSv is 20 mrem; 2 Gy is 200 rad; 300 mrem is 3 mSv. In diagnostic imaging the weighting factor for x-rays is one, so an absorbed dose in milligray and an equivalent dose in millisievert come out numerically the same, which is why the two units get used loosely in conversation and precisely on exams.
Background, and the scale of the numbers
Natural background is about 3 mSv per year, with radon the largest single contributor, and the largest artificial contribution to population dose is medical imaging. One chest image is about 0.02 mSv. Those two numbers together are the only honest way to explain a staff dose report: a badge reading a fraction of a millisievert in a quarter is a small addition to what the person received anyway from the ground they stand on.
The limits
Who or what Limit Occupational, effective dose 50 mSv per year Occupational, cumulative 10 mSv x age in years Lens of the eye 150 mSv per year Thyroid, skin, hands, and feet 500 mSv per year Public 1 mSv per year frequent exposure, 5 mSv infrequent Embryo or fetus 5 mSv over the gestation; 0.5 mSv in any month once declared Education and training, under 18 1 mSv per year
Worked example on the cumulative limit: a 40-year-old technologist's lifetime occupational limit is 10 x 40 = 400 mSv. Worked example on a badge: a quarterly report of 0.4 mSv projects to about 1.6 mSv a year, roughly 3 percent of the annual limit and a normal figure for someone doing fluoroscopy properly. The same badge reading 15 mSv in a quarter projects to 60 mSv, over the limit, and is an investigation: was the badge left in a room, worn under the apron, or is somebody standing in the wrong place during runs.
Personnel monitoring, as a process
Badges are TLD or OSL detectors, read by an outside service, normally on a quarterly cycle. They are worn at the collar outside the apron, because that reads what reaches the unprotected neck, eyes, and thyroid, which is where the limits that bite are. A declared pregnant worker wears a second badge at the waist under the apron. Reports go to the RSO and the radiation safety committee, who look for trends, outliers, and the signature of a badge that never went home. For you there are two rules: your badge stays on you and never on a cabinet in the room, and if you move a badge rack during a renovation, tell the officer where it went.
What to memorize
- Ionization, direct and indirect effect, somatic and genetic, dividing cells most sensitive, dose rate matters.
- 1 Gy = 100 rad, 1 Sv = 100 rem, 1 mSv = 1000 microsieverts.
- Occupational 50 mSv per year and 10 mSv times age cumulative; lens 150; thyroid, skin, hands, feet 500; public 1 and 5; declared pregnancy 0.5 per month; under 18, 1 mSv.
- Background about 3 mSv a year; one chest image about 0.02 mSv.
- Badge at the collar outside the apron, read quarterly, reviewed by the officer and the committee.